Peter Huber Kältemaschinenbau Minichiller 600w OLÉ Kompakter
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Minichiller 600w OLÉ Kompakter

Peter Huber Kältemaschinenbau Minichiller 600w OLÉ Kompakter — Recirculating chillers.

The Minichiller 600w OLÉ Kompakter is a compact recirculating chiller with a pressure/suction pump and OLÉ controller, covering -20 to 40 °C. It is distinguished within the series by its water-cooled condenser, requiring 3–6 bar differential pressure and 5–20 °C cooling water. It serves as a precise temperature control source for laboratory and industrial applications.

Peter Huber Kältemaschinenbau Minichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Minichiller 600w OLÉ Kompakter Current —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000w-H Kompakter —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Minichiller 1000w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000w Kompakter —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Minichiller 1200-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1200 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1200w-H Kompakter —Pilot ONEIII / FL-20–100——
Minichiller 1200w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 280 OLÉ Kompakter 53OLÉI / NFL-5–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300 OLÉ Kompakter 53OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300w-H OLÉ Kompakter 51OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300w OLÉ Kompakter 53OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 600-H OLÉ Kompakter 56OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 600 Kompakter —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D110-120V 1~ 60Hz
Minichiller 600 OLÉ Kompakter 56OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 60Hz
Minichiller 800 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 800w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz

Full specifications

Core Performance & Functional Specifications
Maximum delivery flow rate 24 l/min (pressure pump; 0, 7 bar)
Pump connection M16x1 AG male
Pump speed adjustable no
Pump type Pressure- and suction pump
Temperature range -20...40 °C
Dimensions, Weight & Mechanical Design
Weight 36 kg
Electrical, Power & Drive Specifications
Power supply 208-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller OLÉ
Display resolution 0, 1 K
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 24 l/h (at 15°C, flow -20°C)
at 15°C, flow -10°C: 30 l/h · at 15°C, flow 0°C: 36 l/h · at 15°C, flow 15°C: 39 l/h
Cooling water differential pressure 3 bar (min)
Cooling Water Pressure 6 bar (max)
Maximum return flow temperature 80 °C
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and Series Positioning

The Minichiller 600w OLÉ Kompakter is a water-cooled recirculating chiller from the Minichiller series. Unlike sibling models such as the Minichiller 1000 series, this unit uses a water-cooled condenser, which requires a cooling water supply with a differential pressure of 3–6 bar and an inlet temperature of 5–20 °C. The 'w' in the model code denotes water cooling. The OLÉ controller provides an OLED display with 0.1 K resolution and interfaces including USB (Device) and RS232. The pump is a pressure- and suction pump with a maximum delivery flow rate of 24 l/min at 0.7 bar, and it is not speed-adjustable. The unit has a protection rating of IP20 and safety class I/NFL.

Behavior Under Load and Media

Operating temperature range is -20 to 40 °C, with a maximum return line temperature of 80 °C. Cooling water consumption varies with flow temperature: 36 l/h at 0 °C, 30 l/h at -10 °C, 24 l/h at -20 °C, and 39 l/h at 15 °C, all at a cooling water temperature of 15 °C. The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used. For freeze protection below 0 °C, a mixture of water with glysantine or ethylene glycol is recommended. The refrigerant is R-290 (propane, A3 flammable), requiring assessment of the installation area for flammable refrigerant handling.

Installation, Control and Operation

The unit must be installed indoors in a dry location (IP20) with ambient temperature between 5 and 40 °C. A drip tray for condensation water and thermal fluid must be installed below the unit. Electrical lines directly below the unit must be protected against liquid ingress. The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM; only drinking water or filtered river/sea water may be used (sea water is prohibited). Anti-corrosion agents are mandatory if salts such as chlorides or bromide are added. The unit is classified as EN55011 Class A Group 1, intended for industrial electromagnetic environments. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but not simultaneously in opposite directions. Do not install valves or quick-release couplings in the feed/discharge lines to the application.

Integration & Operating Notes

Mounting
Install drip tray and protect cables below unit

Condensation can form at the cooling water connections and thermal fluid may escape. A drip tray must be placed below the unit, and any electrical lines running directly underneath must be protected against liquid ingress to prevent shock hazard.

Minichiller Operation Manual, pages 15 and 28
Process Limits
Cooling water must be 5–20 °C and 3–6 bar differential

The water-cooled condenser requires a cooling water supply within these limits. Lower differential pressure reduces cooling capacity; higher pressure may damage the circuit. Verify building water system before installation.

Minichiller 600w OLÉ Datasheet, page 1
Electrical
Supply tolerance ±5% voltage, ±2% frequency, not opposing

The unit accepts 208–240 V 1~/2~ 50/60 Hz. Voltage and frequency tolerances may not be applied simultaneously in opposite directions. Class A Group 1 equipment may cause radio interference in residential areas.

Minichiller 600w OLÉ Datasheet, page 2
Maintenance
Use anti-corrosion agents if salts are added to water

If chlorides or bromide are added to the cooling water circuit, anti-corrosion agents are mandatory to prevent corrosion of copper, stainless steel, and other materials. Only drinking water or filtered river/sea water may be used; sea water is prohibited.

Minichiller Operation Manual, page 28
Process Limits
Do not use acetone or polyglycols as heat transfer media

The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). Using these fluids will damage the pump and void reliable operation. For freeze protection below 0 °C, use water mixed with glysantine or ethylene glycol.

Minichiller 600w OLÉ Datasheet, page 2
Safety
Do not install in ATEX zone; refrigerant R-290 is flammable

The unit must not be installed in an ATEX zone. The refrigerant is R-290 (propane, A3 flammable), so the room volume and refrigerant charge must be checked to avoid exceeding the limit value per m³ of ambient air. Additional safety measures may be required.

Minichiller Operation Manual, pages 13 and 16

Selection & Integration Q&A

When should I choose the Minichiller 600w OLÉ Kompakter over other Minichiller models like the 1000 series?

Choose this water-cooled model when a cooling water supply (5–20 °C, 3–6 bar differential) is available and you need -20 to 40 °C control with a pressure/suction pump. The 1000 series offers higher cooling capacity but may be air-cooled; select based on heat load and utilities.

What are the requirements for the cooling water circuit and adjacent piping?

The cooling water must be drinking water or filtered river/sea water (no sea water), with inlet temperature 5–20 °C and differential pressure 3–6 bar. Connections are G1/2 male. Do not install valves or quick-release couplings in the feed/discharge lines to the application.

What are the ambient and installation limits for this chiller?

Install indoors, dry location, ambient 5–40 °C, IP20 (not water-resistant). A drip tray is required below the unit. Do not install in ATEX zones. Electrical lines below must be protected from liquid. Class A Group 1 equipment for industrial EMC environments.

What maintenance and accessories are required that are not included in the base delivery?

A drip tray for condensation/thermal fluid must be provided by the installer. Anti-corrosion agents are mandatory if salts are added to the water circuit. For freeze protection below 0 °C, mix water with glysantine or ethylene glycol. No specific maintenance intervals are given in the available documents.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Minichiller Operation Manual
Manual ·2,9 МБ ·84 pp.
View PDF
PDF
Minichiller 600w OLÉ
Datasheet ·104,6 КБ ·2 pp.
View PDF