BINDER BF 720
BINDER BF 720 — Incubators with Temperature 10…100 °C.
The BINDER BF 720 is a forced-convection incubator with a 734 L interior, adjustable fan speed in 10% steps, and a temperature range of 10–100 °C. As the largest BF series model, it offers 1.75 kW heating and 0.1 K fluctuation, distinguishing it from smaller siblings like the BF 400. It is intended for incubation of cultures at 37 °C and precise tempering of harmless materials in large-volume laboratory processes.
Direct drop-in replacements
BINDER BF Series — Model Comparison & Selection
Full specifications
| Temperature | 10...100 °C |
| Heating-up time to 100 °C | 45 min |
| Interior volume | 734 L |
| Load per rack | 45 kg (BF720-230V / BF720UL-240V) |
| Nominal power | 1.75 kW |
| Permitted load | 315 kg (BF720-230V / BF720UL-240V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.3 K |
| Back wall clearance | 160 mm (BF720-230V / BF720UL-240V) |
| Side wall clearance | 100 mm (BF720-230V / BF720UL-240V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 37°C mfr | 4 min |
| Depth | 870 mm (BF720-230V / BF720UL-240V) |
| Height | 1590 mm (BF720-230V / BF720UL-240V) |
| Width | 1165 mm (BF720-230V / BF720UL-240V) |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and positioning
- The BF 720 is a floor-standing forced-convection incubator with a 734 L usable volume, net dimensions 1,165 mm wide × 1,590 mm high × 870 mm deep. It is the largest model in the BF series, which also includes sizes 56, 115, 260, and 400. Unlike smaller siblings, the BF 720 provides a temperature range starting at 10 °C (vs. 5–8 °C for others) and a nominal power of 1.75 kW. It is available in two voltage variants: BF720-230V (230 V, 50/60 Hz, CEE 7/7 plug) and BF720UL-240V (240 V, 60 Hz, NEMA 6-20 plug).
Behavior under load and in operation
- The chamber achieves a temperature uniformity of 0.3 K and fluctuation of 0.1 K, with a recovery time of 4 minutes after a 30-second door opening at 37 °C. Heating-up time to 100 °C is 45 minutes. The forced convection system allows fan speed adjustment in 10% increments, enabling optimization of airflow for different load patterns. The usable volume must not be loaded beyond half capacity, and samples require at least 10% wall clearance to maintain homogeneity.
Installation, control, and operation
- The BF 720 requires a single-phase 230 V supply (or 240 V for UL variant) with a 12.5 A time-lag fuse (5×20 mm, 250 V). It must be installed indoors on a non-flammable surface with 160 mm rear and 100 mm side clearance. The R4 controller provides a 0.1 °C display resolution and includes a safety controller that must be set 2–5 °C above the working set-point. The chamber supports timer functions (Delayed On/Off, temperature-dependent Delayed Off) and an optional Ethernet interface for network integration. Annual maintenance and electrical safety testing are mandatory.
Integration & Operating Notes
The BF 720 requires a 230 V single-phase supply (or 240 V for UL variant) with a tolerance of ±10%. Exceeding this range can cause malfunction or damage to the controller and heating system. Verify the supply quality before connection and use the specified external fuse of 12.5 A time-lag (T) 5×20 mm.
Loading the chamber beyond 50% of its usable volume or placing samples closer than 10% of the interior dimensions to the walls will disrupt airflow and compromise temperature uniformity. This constraint is critical for achieving the specified 0.3 K uniformity.
The safety controller acts as an absolute maximum temperature limit. It must be set 2–5 °C above the desired working temperature to prevent accidental overtemperature while avoiding nuisance trips. Re-adjust after any set-point change.
The chamber must be maintained at least once a year by licensed electricians or BINDER-authorized personnel. An annual repeat electrical inspection according to EN 50678/VDE 0701 and EN 50699/VDE 0702 is also required. Failure to comply may void warranty.
Attempting to replace the door gasket while the chamber is hot can damage the gasket and sealing surfaces. Always allow the chamber to cool to ambient temperature before replacement.
Selection & Integration Q&A
Choose the BF 720 when you need a large 734 L capacity for high-throughput incubation or tempering. Smaller models (BF 400, BF 260) offer lower minimum temperatures (5–7 °C) but less volume. The BF 720 is ideal for bulk processing where 10 °C minimum is sufficient.
No, the chamber must not be installed in unventilated recesses. Minimum clearances of 160 mm at the rear and 100 mm at the sides are required for adequate ventilation. Install indoors on a non-flammable surface with ambient temperature 18–40 °C and humidity ≤70% non-condensing.
Annual maintenance by authorized personnel and electrical safety testing are mandatory. Recommended cleaning agents are neutral cleaner (art. no. 1002-0016) and disinfectant spray (1002-0022). Racks can be autoclaved. Optional Ethernet interface and APT-COM 4 software are available but not included in base delivery.