BINDER BF 720 — Incubators with Temperature 10…100 °C.

The BINDER BF 720 is a forced-convection incubator with a 734 L interior, adjustable fan speed in 10% steps, and a temperature range of 10–100 °C. As the largest BF series model, it offers 1.75 kW heating and 0.1 K fluctuation, distinguishing it from smaller siblings like the BF 400. It is intended for incubation of cultures at 37 °C and precise tempering of harmless materials in large-volume laboratory processes.

Temperature
10…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER BF Series — Model Comparison & Selection

Model Temperature, °CHeating up time to 100 c, minHeight, cmInterior volume, LLoad per rack, kgPermitted load, kg
BF 720 Current 10–1004515973445315
BF 115 8–1006572.511430150
BF 260 7–1003084.525740270
BF 400 5–10012102.54003590
BF 56 7–1003562.5593070

Full specifications

Core Performance & Functional Specifications
Temperature 10...100 °C
Heating-up time to 100 °C 45 min
Interior volume 734 L
Load per rack 45 kg (BF720-230V / BF720UL-240V)
Nominal power 1.75 kW
Permitted load 315 kg (BF720-230V / BF720UL-240V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.3 K
Back wall clearance 160 mm (BF720-230V / BF720UL-240V)
Side wall clearance 100 mm (BF720-230V / BF720UL-240V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 37°C mfr 4 min
Dimensions, Weight & Mechanical Design
Depth 870 mm (BF720-230V / BF720UL-240V)
Height 1590 mm (BF720-230V / BF720UL-240V)
Width 1165 mm (BF720-230V / BF720UL-240V)
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Phase (Nominal voltage) mfr 1

Description

Construction and positioning

Behavior under load and in operation

Installation, control, and operation

Integration & Operating Notes

Electrical
Supply voltage must stay within ±10% of nominal

The BF 720 requires a 230 V single-phase supply (or 240 V for UL variant) with a tolerance of ±10%. Exceeding this range can cause malfunction or damage to the controller and heating system. Verify the supply quality before connection and use the specified external fuse of 12.5 A time-lag (T) 5×20 mm.

Operating Manual, Requirements for the location of installation, p.18
Process Limits
Do not load more than half of the usable volume

Loading the chamber beyond 50% of its usable volume or placing samples closer than 10% of the interior dimensions to the walls will disrupt airflow and compromise temperature uniformity. This constraint is critical for achieving the specified 0.3 K uniformity.

Operating Manual, Placing the charging material, p.19
Control & I/O
Set safety controller 2–5 °C above working set-point

The safety controller acts as an absolute maximum temperature limit. It must be set 2–5 °C above the desired working temperature to prevent accidental overtemperature while avoiding nuisance trips. Re-adjust after any set-point change.

Operating Manual, Set-point entry for temperature, p.20
Maintenance
Annual maintenance and electrical safety testing required

The chamber must be maintained at least once a year by licensed electricians or BINDER-authorized personnel. An annual repeat electrical inspection according to EN 50678/VDE 0701 and EN 50699/VDE 0702 is also required. Failure to comply may void warranty.

Operating Manual, Maintenance intervals, p.69
Maintenance
Replace door gasket only when cold

Attempting to replace the door gasket while the chamber is hot can damage the gasket and sealing surfaces. Always allow the chamber to cool to ambient temperature before replacement.

Operating Manual, Maintenance intervals, p.69

Selection & Integration Q&A

When should I choose the BF 720 over smaller BF series models like the BF 400?

Choose the BF 720 when you need a large 734 L capacity for high-throughput incubation or tempering. Smaller models (BF 400, BF 260) offer lower minimum temperatures (5–7 °C) but less volume. The BF 720 is ideal for bulk processing where 10 °C minimum is sufficient.

Can the BF 720 be installed in a corner or recess?

No, the chamber must not be installed in unventilated recesses. Minimum clearances of 160 mm at the rear and 100 mm at the sides are required for adequate ventilation. Install indoors on a non-flammable surface with ambient temperature 18–40 °C and humidity ≤70% non-condensing.

What maintenance and accessories are required for the BF 720?

Annual maintenance by authorized personnel and electrical safety testing are mandatory. Recommended cleaning agents are neutral cleaner (art. no. 1002-0016) and disinfectant spray (1002-0022). Racks can be autoclaved. Optional Ethernet interface and APT-COM 4 software are available but not included in base delivery.