BINDER BD 56
BINDER ·Made in Germany · In Active Production

BINDER BD 56

BINDER BD 56 — Incubators with Temperature 5…100 °C.

The BD 56 is a 57 L natural-convection incubator (BD Avantgarde.Line, Art. 9010-0323/0324) with a 5–100 °C working range and 0.1 K display resolution. It is the mid-size standard model between the BD 23 and BD 115, distinguished by class 3.1 overtemperature protection and 0.3 K uniformity. It is intended for incubation of cultures at 37 °C and warm storage of liquids.

Temperature
5…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER BD Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeating up time to 100 c, minHeight, cmInterior volume, LLoad per rack, kg
BD 56 Current 5–10056.57562.55730
BD 115 5–10060.59070.511230
BD 23 5–100525949.52012
BD 260 5–10076759425340
BD 400 5–10076.5120102.540035
BD 720 5–100879015973745

Full specifications

Core Performance & Functional Specifications
Temperature 5...100 °C
Heating-up time to 100 °C 75 min
Interior volume 57 L (BD056-230V / BD056UL-120V)
Load per rack 30 kg (BD056-230V / BD056UL-120V)
Nominal power 0.3 kW
Permitted load 70 kg (BD056-230V / BD056UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.3 K
Back wall clearance 160 mm (BD056-230V / BD056UL-120V)
Side wall clearance 100 mm (BD056-230V / BD056UL-120V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 37°C mfr 15 min
Dimensions, Weight & Mechanical Design
Depth 565 mm (BD056-230V / BD056UL-120V)
Height 625 mm (BD056-230V / BD056UL-120V)
Width 560 mm (BD056-230V / BD056UL-120V)
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behaviour under load and in operation

Installation, connection and operation

Integration & Operating Notes

Electrical
Upstream 6.3 A time-lag fuse is mandatory

The chamber has no internal miniature circuit breaker; protection is an external 6.3 A time-lag (T) 5×20 mm / 250 V fuse. Size the branch circuit and verify supply within ±10 % of 230 V (or 120 V for the UL variant) before first start-up.

Quick Start Guide BD/BF/ED/FD/FED, Electrical connection data, p.18
Mounting
Keep 160 mm rear and 100 mm side clearance

Natural convection depends on free air circulation around the housing. Installing closer than the specified clearances, or in a non-ventilated recess, risks overheating and invalid temperature uniformity; the manual explicitly forbids unventilated recesses.

Data Sheet BD 56; Quick Start Guide, Requirements for the location of installation, p.18
Process Limits
Load no more than half the usable volume

Samples must remain inside the usable volume with at least 10 % of interior dimensions as wall clearance. Overloading blocks the buoyancy-driven airflow, so the 0.3 K uniformity and 0.1 K fluctuation figures no longer apply to the load.

Quick Start Guide, Placing the charging material in the usable volume, p.19
Control & I/O
Safety controller must track the working set-point

The class 3.1 safety controller is an absolute maximum limit (10–100 °C range). Set it 2–5 °C above the process set-point and re-check after every set-point change; otherwise a legitimate set-point change can trip the overtemperature protection and cut heating.

Quick Start Guide, Set-point entry for temperature, p.20; Operating Manual, Temperature safety devices, p.29-30
Maintenance
Annual electrical inspection and BINDER-authorized service

Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts, including repeat electrical tests per EN 50678/VDE 0701 and EN 50699/VDE 0702. Replace the door gasket only when the chamber is cold to avoid damaging it.

Operating Manual, Section 10.1-10.2, p.36
Cooling & Media
Hot-air disinfection limited to 100 °C on BD

Unlike ED/FD ovens (190 °C sterilization), the BD 56 permits hot-air disinfection only at 100 °C. Remove all flammable goods first. Racks may be autoclaved; use neutral cleaner 1002-0016 and disinfectant 1002-0022, and never apply neutral cleaner to zinc-coated hinge parts or the rear wall.

Quick Start Guide, Cleaning and decontamination, p.23

Selection & Integration Q&A

When should I choose the BD 56 rather than the BD 23 or BD 115?

Pick the 56 when 57 L accommodates your rack load (≤30 kg/rack, 70 kg total) at 37 °C. The 23 is smaller; the 115 gives more volume but the same 5–100 °C range and 0.1 K resolution. All share natural convection, so size is the deciding factor.

What upstream electrical protection and supply does the BD 56 require?

A 230 V 1~ 50/60 Hz supply within ±10 % (or 120 V/60 Hz for BD056UL-120V) protected by an external 6.3 A time-lag 5×20 mm fuse. Nominal power is 0.30 kW; the chamber has no internal circuit breaker, so the fuse is the only overcurrent protection.

Can the BD 56 be installed in a small closed cabinet or recess?

No. The manual forbids unventilated recesses. Install indoors on a non-flammable surface with 160 mm rear and 100 mm side clearance, ambient 18–40 °C, ≤70 % r.h. non-condensing, altitude ≤2000 m, and IP 20 means no water exposure.

What maintenance and consumables are needed beyond the base delivery?

Annual maintenance by BINDER-authorized personnel plus repeat electrical tests (EN 50678/50699). Consumables: 6.3 A time-lag fuses, door gasket (replace cold), neutral cleaner 1002-0016 and disinfectant 1002-0022. APT-COM 4, Pt100 sensor and analog output are options, not included.