BINDER BF 56
BINDER BF 56 — Incubators with Temperature 7…100 °C.
The BINDER BF 56 is a forced-convection incubator with a 59 L interior, adjustable fan speed in 10 % steps, and a working range of +7 to 100 °C. Within the BF series it is the smallest chamber, distinguished by the tightest temperature uniformity (0.2 K) and fluctuation (0.1 K) for culture incubation at 37 °C. It serves as a benchtop incubation stage where airflow and temperature homogeneity are critical.
Direct drop-in replacements
BINDER BF Series — Model Comparison & Selection
Full specifications
| Temperature | 7...100 °C |
| Heating-up time to 100 °C | 35 min |
| Interior volume | 59 L (BF056-230V / BF056UL-120V) |
| Load per rack | 30 kg (BF056-230V / BF056UL-120V) |
| Nominal power | 0.4 kW |
| Permitted load | 70 kg (BF056-230V / BF056UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.2 K |
| Back wall clearance | 160 mm (BF056-230V / BF056UL-120V) |
| Side wall clearance | 100 mm (BF056-230V / BF056UL-120V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 37°C mfr | 3 min |
| Depth | 565 mm (BF056-230V / BF056UL-120V) |
| Height | 625 mm (BF056-230V / BF056UL-120V) |
| Width | 560 mm (BF056-230V / BF056UL-120V) |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The BF 56 uses forced convection with a variable-speed fan, unlike the natural-convection BD series. Fan speed is set in 10 % increments, allowing airflow to be tuned for sensitive samples.
- As the smallest BF model (59 L), it sits below the BF 115, BF 260, BF 400 and BF 720 in volume. Its temperature range (+7 to 100 °C) is comparable to the BF 260 but starts 1 K lower than the BF 115 and 2 K higher than the BF 400.
- The interior dimensions are 560 mm wide × 625 mm high × 565 mm deep (net), with a permitted load of 70 kg and a maximum of 30 kg per rack.
Behavior under load and in operation
- Temperature fluctuation is 0.1 K and uniformity 0.2 K, supported by the forced-air circulation. Heating from ambient to 100 °C takes 35 minutes.
- After a 30-second door opening at 37 °C, the chamber recovers its set temperature in 3 minutes.
- The controller displays temperature with 0.1 °C resolution. The safety controller set-point range is 10 °C to 100 °C and should be set 2–5 °C above the working set-point.
- Loading must respect the usable volume: keep at least 10 % of interior dimensions as wall clearance and do not fill more than half the volume to maintain airflow and homogeneity.
Installation, connection and control
- The BF 56 is available in two voltage variants: BF056-230V (230 V 1~ 50/60 Hz, CEE 7/7 plug, 0.4 kW) and BF056UL-120V (120 V 1~ 60 Hz, NEMA 5-15 plug). The chamber fuse is external, 6.3 A time-lag (T) 5×20 mm / 250 V.
- Supply voltage tolerance is ±10 %; verify against the type plate before connection. The chamber is rated IP 20 and pollution degree 2 / over-voltage category II per IEC 61010-1.
- The R4 controller manages temperature and fan speed set-points, timer functions (Delayed Off, Temperature-dependent Delayed Off, Delayed On), ramp function, and optional USB data recording. Ethernet is optional for BF models.
- The overtemperature protective device class 1 trips at 120 °C; the safety controller (class 2 or 3.1) provides an adjustable absolute limit. After a trip, disconnect power for at least 10 seconds and let the chamber cool before reset.
Integration & Operating Notes
The chamber requires unobstructed airflow for heat dissipation and safe operation. Installing it closer than the specified wall clearances or in an unventilated recess will cause overheating and may invalidate performance specifications. Ensure the surface is non-flammable and the room is indoors.
The chamber does not include an internal fuse for the 230 V variant; a 6.3 A time-lag (T) 5×20 mm / 250 V fuse must be provided in the supply circuit. Verify that the supply voltage matches the type plate within ±10 % and that the circuit can deliver the 0.4 kW nominal power without excessive voltage drop.
The safety controller acts as an absolute maximum temperature limit. After any change to the temperature set-point, re-adjust the safety controller set-point to 2–5 °C above the new value. The permissible range is 10 °C to 100 °C; setting it too close may cause nuisance trips, too high may fail to protect samples.
Samples must remain within the usable volume with at least 10 % of interior dimensions as clearance from walls. Overloading or placing samples too close to the walls obstructs airflow, leading to temperature stratification and non-uniform results. The maximum load per rack is 30 kg and total permitted load is 70 kg.
The chamber must be maintained at least once a year by licensed electricians or BINDER-authorized personnel. An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required, as well as a test before initial startup and after any repair. Warranty is void if maintenance is performed by unauthorized personnel.
The BF 56 can be hot-air disinfected at 100 °C / 212 °F. All flammable goods must be removed before disinfection. Racks are removable and can be autoclaved. Use neutral cleaning agent (art. no. 1002-0016) or disinfectant spray (art. no. 1002-0022); do not apply neutral cleaner to zinc-coated hinge parts or rear wall.
Selection & Integration Q&A
Choose the BF 56 for small workloads up to 59 L where bench space is limited and precise temperature uniformity (0.2 K) is needed. The BF 115 offers +8 to 100 °C and larger volume; the BF 260 has the same +7 to 100 °C range but more capacity. Select the BF 56 when sample throughput fits its 59 L and 70 kg load limit.
The BF056-230V requires 230 V 1~ 50/60 Hz, 0.4 kW, with an external 6.3 A time-lag (T) 5×20 mm / 250 V fuse. The BF056UL-120V uses 120 V 1~ 60 Hz with a NEMA 5-15 plug. Supply voltage must be within ±10 % of nominal. The chamber has no internal fuse for the 230 V variant; the fuse must be provided in the installation.
Install indoors on a non-flammable surface with ambient temperature +18 to +40 °C, max 70 % r.h. non-condensing, and altitude up to 2000 m. Maintain 160 mm rear clearance and 100 mm side clearance. Do not install in unventilated recesses or explosive atmospheres. The chamber is IP 20 and must not be exposed to water.
Annual maintenance by authorized personnel and annual electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required. Recommended consumables: neutral cleaning agent art. no. 1002-0016 and disinfectant spray art. no. 1002-0022. Optional accessories include APT-COM 4 software, Pt100 object temperature sensor, analog output, and Ethernet interface (not standard on BF).