BINDER KB 130
BINDER KB 130 — Incubators with Temperature 0…70 °C.
The BINDER KB 130 is a 127 L compressor-cooled incubator with a 0–70 °C working range and RD4 controller, positioned between the KB 65 and KB 260 in the KB series. Its 270 W heat compensation and 0.1 K fluctuation make it suited for long-term storage at 4 °C and incubation at 37 °C. The 130 size requires the flexible tilt protection kit during installation, as do the KB 65 and KB 470.
Direct drop-in replacements
BINDER KB Series — Model Comparison & Selection
Full specifications
| Temperature | 0...70 °C |
| Interior volume | 127 L |
| Load per rack | 30 kg |
| Max. heat compensation | 270 W |
| Nominal power | 1.2 kW |
| Permitted load | 150 kg (KB130-230V / KB130UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.2 K |
| Back wall clearance | 100 mm (KB130-230V / KB130UL-120V) |
| Side wall clearance | 100 mm (KB130-230V / KB130UL-120V) |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 37°C mfr | 3 min |
| Depth | 710 mm (KB130-230V / KB130UL-120V) |
| Height | 1005 mm (KB130-230V / KB130UL-120V) |
| Interior width | 656 mm (KB130-230V / KB130UL-120V) |
| Width | 900 mm (KB130-230V / KB130UL-120V) |
| Net weight of the unit (empty) mfr | 124 kg |
| Power frequency | 50 Hz |
| Rated voltage | 220...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
The KB 130 (article 9020-0472) is a floor-standing cooling incubator with 127 L interior volume, 656 mm interior width, and 124 kg empty weight. It uses compressor technology with R600A refrigerant and the RD4 chamber controller. Within the KB series, the 130 size sits between the KB 65 and KB 260; unlike sizes 260 and larger, it does not require a wall spacer, but it does require the flexible tilt protection kit (shared with KB 65 and KB 470). The chamber is IP 20 and must be installed splash-proof.
Behavior under load and in different modes
The intended temperature application range is 4 °C to 37 °C for cultivation of microorganisms, though the full operating range is 0–70 °C. Temperature fluctuation is 0.1 K and uniformity 0.2 K. Maximum heat compensation is 270 W, and recovery time after a 30 s door opening at 37 °C is 3 minutes. The chamber supports hot-air disinfection, automatic setpoint switchover by date or cyclic control, and defrosting during refrigerating operation with anti-condensation protection. The temperature safety device is adjustable per DIN 12880:2007 class 2/3.3, configurable as a temperature limiter or temperature protection.
Installation, connection and control
Electrical supply is 220–240 V, 1~, 50 Hz, nominal power 1.2 kW, unit fuse 16.0 A. The RD4 controller provides Ethernet and USB interfaces; APT-COM 4 software can network up to 100 chambers. The rear connection panel includes interfaces for computer communication. Minimum wall clearances are 100 mm at the back and 100 mm at the side. The chamber must not be installed in unventilated recesses. The condenser is behind a tool-free removable machine room cover for cleaning.
Integration & Operating Notes
The flexible tilt protection kit must be mounted on KB/KB-UL 65, 130 and 470 during installation. Omitting it leaves the chamber without the prescribed tip-over safeguard, which is a safety-relevant deviation from the manual.
The chamber requires 100 mm wall clearance at the back and sidewise to ensure sufficient ventilation for heat dispersal. Installing it in an unventilated recess or closer than these distances risks overheating and chamber damage.
The KB 130 draws 1.2 kW nominal at 220–240 V, 1~, 50 Hz, with a 16.0 A unit fuse. The branch circuit and protective device must be coordinated with this nominal current; the UL variant (KB130UL-120V) uses a different supply and plug and is not interchangeable.
Materials containing acids or halides that may attack stainless steel, aluminum or copper components must not be loaded. Corrosive damage from such ingredients is excluded from BINDER liability, so process compatibility must be checked before use.
Regular maintenance is required at least once a year, and an annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended before initial startup and after any service. The controller shows 'Service due!' after one year and repeats it weekly until confirmed.
With increased dust in ambient air, the condenser behind the tool-free removable machine room cover must be cleaned by suction several times a year to maintain refrigerating performance. Refrigeration system work is restricted to personnel trained per EN 13313:2010 due to flammable R600A refrigerant.
Selection & Integration Q&A
Choose the KB 130 for 127 L interior volume when the KB 65 is too small and the KB 260 would require the wall spacer. The 130 shares the tilt protection kit requirement with the 65 and 470, but offers more usable volume than the 65 without the larger footprint of the 260.
The KB 130 needs 220–240 V, 1~, 50 Hz, 1.2 kW, with a 16.0 A unit fuse. The KB130UL-120V is a separate 120 V variant with NEMA 5-20 plug and different current; it cannot share the same branch circuit or plug configuration.
Maintain 100 mm clearance at the rear and sides, do not install in unventilated recesses, and mount the flexible tilt protection kit required for sizes 65, 130 and 470. The chamber is IP 20 and must be kept splash-proof.
Plan annual maintenance and electrical repeat inspection; the controller shows a service reminder after one year. Optional accessories include APT-COM 4 software, analog temperature outputs, zero-voltage alarm contacts, door lock, internal LED light, Pt 100 object sensor, and racks with U-rails or telescopic rails.