Peter Huber Kältemaschinenbau CC-205B
Peter Huber Kältemaschinenbau CC-205B — Thermostatic baths.
The CC-205B is a 4.8-litre stainless-steel bath circulator with a variable-speed pressure/suction pump (M16x1 male) and Pilot ONE controller. Its operating range is configuration-dependent: 45–200 °C uncooled, 20–200 °C with water cooling, or -30–200 °C with a refrigerator. It serves as a temperature-control source for external circuits and immersed samples.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection
| Model | Controller | Interface | Pump connection | Temperature range, °C | Weight, kg | Max delivery flow rate, liter/minute |
|---|---|---|---|---|---|---|
| CC-205B Current | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 27 |
| CC-202C | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 22 |
| CC-208B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 10 | 27 |
| CC-212B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 11 | 22 |
| CC-215B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 12 | — |
| CC-220B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 14 | 27 |
| CC-225B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 16 | 27 |
| KISS 202C | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 8 | 14 |
| KISS 205B Badumwälzthermostat mit Edelstahlbad | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 9 | 14 |
| KISS 208B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 10 | 14 |
| KISS 212B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 11 | 14 |
| KISS 215B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 12 | 14 |
| KISS 220B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 14 | 14 |
| KISS 225B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 16 | 14 |
Full specifications
| Filling volume | 3 l (minimum) |
| Maximum delivery flow rate | 27 l/min (pressure pump; also 0, 7 bar) suction pump; also 0, 4 bar: 22 l/min |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 45...200 °C with cooling unit: -30...200 °C · with water cooling: 20...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 178 x 337 x 355 mm |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Bath opening WxD/BD mfr | 105 x 90 / 150 mm |
| Bath volume mfr | 4, 8 l |
| Weight | 9 kg |
| Current consumption | 10 A (max.) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The CC-205B is the 4.8-litre bath opening (105 x 90 mm, 150 mm deep) member of the CC-20xB stainless-steel series, positioned between the smaller CC-202C and the larger CC-208B/CC-212B/CC-215B siblings. Heating capacity is voltage-dependent, from 1.5 kW at 200 V to 2.1 kW at 240 V, and temperature stability is stated as ±0.02 K. The enclosure is 178 x 337 x 355 mm and weighs 9 kg.
Thermal behaviour and media
- The achievable range is not a single figure: 45–200 °C without cooling, 20–200 °C with the 3/8" tap-water cooling coil, and -30–200 °C with a refrigerator. Selection must follow the cooling utility actually available on site.
- The variable-speed pressure/suction pump delivers up to 27 l/min at 0.7 bar (pressure side) and 22 l/min at 0.4 bar (suction side), so external-circuit pressure drop must be checked against these figures.
- Minimum filling volume is 3 l within the 4.8 l bath.
Control, connection and operation
The Pilot ONE controller provides Ethernet, USB host/device and RS232 interfaces, plus a Pt100 external sensor input. Pump speed is adjustable. The pump connection is M16x1 male. Overtemperature protection is adjustable per DIN 12876, safety class III/FL. The unit is rated IP20 and Class A, Group 1 to EN55011, so it is intended for non-residential EMC environments and must be kept from water spray and dust.
Integration & Operating Notes
The supply is 200–240 V 1~/2~ 50/60 Hz with a maximum current of 10 A. A minimum 10 A and maximum 16 A fuse is required, and external branch circuit protection to UL 489 is mandatory; undersizing the protection risks nuisance trips, oversizing removes the intended safeguard.
A ±10% voltage tolerance is permitted only if the frequency tolerance does not simultaneously run in the opposite direction. Combined worst-case deviations on an unstable grid can exceed the unit's design margin, so supply quality should be verified before commissioning.
Without cooling the range starts at 45 °C; with the 3/8" tap-water cooling coil it starts at 20 °C; only with a refrigerator does it reach -30 °C. The cooling utility must therefore be decided before specifying the unit, since it changes the usable process window.
The pressure pump provides up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar. A circuit whose pressure drop exceeds these values will reduce flow below the required rate, so the loop resistance should be calculated against these curves before installation.
Single/two-phase units are delivered with a power cable and a country-specific plug that has to be stated when ordering. Omitting this detail leaves the unit without a usable mains termination and forces a site retrofit.
The IP20 enclosure and the 5–40 °C ambient limit exclude wet, sprayed or dusty locations. Because the unit is Class A, Group 1 to EN55011, additional EMC measures may be needed if it is placed near sensitive residential or measurement electronics.
Overtemperature protection per DIN 12876 with safety class III/FL must be set to a value above the intended process temperature but below the limits of the bath fluid and immersed samples; an incorrect setting either trips the unit prematurely or removes the protective function.
Selection & Integration Q&A
Pick the CC-205B when a 4.8 l bath with a 105 x 90 mm opening and 3 l minimum fill suits the samples or immersion cooler. Larger siblings such as the CC-208B or CC-212B are for bigger bath volumes and openings; the CC-202C is the smaller alternative.
The pump connection is M16x1 male, so the loop plumbing must mate with that thread. Flow must stay within 27 l/min at 0.7 bar (pressure) and 22 l/min at 0.4 bar (suction), and a 3/8" tap-water coil connection is needed if water cooling is used.
Install in a dry room at 5–40 °C ambient, IP20, away from water spray and dust. Supply is 200–240 V 1~/2~ 50/60 Hz, max 10 A, with a 10–16 A fuse and UL 489 branch protection; voltage and frequency tolerances must not oppose each other.
The country-specific mains plug must be specified at ordering. A refrigerator is required to reach -30 °C and is not part of the uncooled configuration; the 3/8" tap-water cooling coil line must be provided on site. External Pt100 sensors and loop plumbing are separate items.