Peter Huber Kältemaschinenbau KISS 208B
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS 208B

Peter Huber Kältemaschinenbau KISS 208B — Thermostatic baths.

The KISS 208B is a 7.5 l stainless-steel bath immersion circulator with a non-variable pressure/suction pump, rated 25–200 °C on its own, 20–200 °C with water cooling and −30–200 °C with a refrigerator. Its 1.5–2.1 kW heater and ±0.05 K stability distinguish it from smaller CC-202C/205B siblings; it serves as the temperature-control source for external circuits and bath processes up to 200 °C.

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection

Model ControllerInterfacePump connectionTemperature range, °CWeight, kgMax delivery flow rate, liter/minute
KISS 208B Current KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001014
CC-202C Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200922
CC-205B Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200927
CC-208B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–2001027
CC-212B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001122
CC-215B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–20012—
CC-220B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001427
CC-225B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001627
KISS 202C KISSUSB (Device), RS232 InteM16x1 AG male45–200814
KISS 205B Badumwälzthermostat mit Edelstahlbad KISSUSB (Device), RS232 InteM16x1 AG male45–200914
KISS 212B KISSUSB (Device), RS232 InteM16x1 AG male25–2001114
KISS 215B KISSUSB (Device), RS232 InteM16x1 AG male25–2001214
KISS 220B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001414
KISS 225B KISSUSB (Device), RS232 InteM16x1 AG male25–2001614

Full specifications

Core Performance & Functional Specifications
Alarm output optic, acoustic
Dimensions (w × D × H) 290 x 350 x 375 mm
Filling volume 4 l (min)
Maximum delivery flow rate 14 l/min (pressure pump; 0, 25 bar)
Pump connection M16x1 AG male (Option, #19607) (Option, #19607)
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 25...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 05 K
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump; 0, 17 bar)
Bath opening WxD/BD mfr 230 x 127 / 150 mm (WxD/depth)
Bath volume mfr 7, 5 l
Dimensions, Weight & Mechanical Design
Weight 10 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max)
Power supply 200-240V 1~/2~ 50/60Hz V; Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Control, safety and electrical

Integration & Operating Notes

Electrical
Supply tolerance combinations can damage the unit

The datasheet explicitly forbids combined deviations such as −10% voltage with +3% frequency. Verify the site supply against the 200–240 V, 1~/2~, 50/60 Hz envelope before commissioning, because a supply that looks acceptable on each axis alone may still violate the allowed combination.

KISS 208B Immersion Circulator Datasheet, page 2
Electrical
External UL 489 branch protection is mandatory

The equipment does not carry its own branch-circuit protection; the installer must provide a UL 489 device rated between 10 A and 16 A. Size the circuit for the 10 A maximum current draw and confirm the protective device is coordinated with the 200–240 V single- or two-phase feed.

KISS 208B Immersion Circulator Datasheet, page 2
Process Limits
Cooling method fixes the achievable low temperature

The unit reaches only 25 °C without auxiliary cooling, 20 °C with water cooling and −30 °C with a refrigerator. Select the cooling configuration at order stage; a unit specified without cooling cannot be pushed below 25 °C later by controller settings alone.

KISS 208B Immersion Circulator Datasheet, page 1
Cooling & Media
Acetone and polyglycols attack the plastic pump

The pump is plastic and not resistant to acetone or polyglycols (manufacturer-dependent). Confirm fluid compatibility before filling; for sub-zero operation use a water/glysantine or water/ethylene glycol mixture rather than pure water to avoid freezing damage.

KISS 208B Immersion Circulator Datasheet, page 2
Control & I/O
Autostart after power failure may be unwanted

The autostart function restarts the unit automatically after a power failure. For processes where an unattended restart is hazardous or would spoil a sample, this behavior must be assessed and the process designed around it before relying on the KISS controller in an automated line.

KISS 208B Immersion Circulator Datasheet, page 1

Selection & Integration Q&A

When should I choose the KISS 208B rather than a smaller or larger sibling such as the CC-205B or CC-212B?

Pick the 208B when the process needs a 7.5 l bath with 1.5–2.1 kW heating and 14 l/min pump flow. The CC-205B is the smaller-bath step; the CC-212B is larger. If your vessel or external loop exceeds the 208B bath volume, step up rather than overfilling.

What must I provide on the electrical and hydraulic side to integrate this circulator into an external circuit?

Provide a 200–240 V, 1~ or 2~, 50/60 Hz feed with a 10–16 A UL 489 branch device, respecting the forbidden voltage/frequency tolerance combinations. On the hydraulic side, the non-variable pump gives 14 l/min at 0.25 bar pressure and 10.5 l/min at 0.17 bar suction; the M16x1 male pump connection is an option (#19607), so confirm fittings before piping.

What is not included in base delivery that I must plan for in maintenance and operation?

The M16x1 male pump connection is an option (#19607), not standard, and no cooling unit is included unless specified—water cooling or a refrigerator must be ordered to reach 20 °C or −30 °C. Plan for a freeze-protection glycol mixture and for periodic verification of the adjustable DIN 12876 overtemperature limit; no consumables or cables are listed in the datasheet.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS 208B Immersion Circulator Datasheet
Datasheet ·93,7 КБ ·2 pp.