Peter Huber Kältemaschinenbau KISS 205B Badumwälzthermostat mit Edelstahlbad
Peter Huber Kältemaschinenbau KISS 205B Badumwälzthermostat mit Edelstahlbad — Thermostatic baths.
The KISS 205B is a 4.8-litre stainless-steel bath thermostat with an integrated pressure/suction pump and 1.5–2.1 kW heating, covering 45–200 °C uncooled, 20–200 °C with water cooling and -30–200 °C with a refrigerator. Its ±0.05 K stability and DIN 12876 III/FL safety classification distinguish it within the KISS series for precise external-circuit tempering.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection
| Model | Controller | Interface | Pump connection | Temperature range, °C | Weight, kg | Max delivery flow rate, liter/minute |
|---|---|---|---|---|---|---|
| KISS 205B Badumwälzthermostat mit Edelstahlbad Current | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 9 | 14 |
| CC-202C | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 22 |
| CC-205B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 27 |
| CC-208B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 10 | 27 |
| CC-212B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 11 | 22 |
| CC-215B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 12 | — |
| CC-220B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 14 | 27 |
| CC-225B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 16 | 27 |
| KISS 202C | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 8 | 14 |
| KISS 208B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 10 | 14 |
| KISS 212B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 11 | 14 |
| KISS 215B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 12 | 14 |
| KISS 220B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 14 | 14 |
| KISS 225B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 16 | 14 |
Full specifications
| Alarm output | optic, acoustic |
| Dimensions (w × D × H) | 178 x 337 x 355 mm |
| Maximum delivery flow rate | 14 l/min (pressure pump max.; 0, 25 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Temperature range | 45...200 °C with cooling unit: -30...200 °C · with water cooling: 20...200 °C |
| Temperature stability mfr | ±0, 05 K |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min (suction pump max.; 0, 17 bar) |
| Bath opening WxD/BD mfr | 105 x 90 / 150 mm (WxD/BD) |
| Bath volume mfr | 4, 8 l |
| Filling volume min. mfr | 2, 5 l (min.) |
| Weight | 9 kg |
| Current consumption | 10 A (max.) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Bath volume 4.8 l (minimum filling 2.5 l), bath opening 105 x 90 mm, depth 150 mm; stainless-steel bath with cover for the bath bridge included.
- KISS controller with OLED display; pressure and suction pump, speed not variable, M16x1 male pump connection.
- Within the series the 205B is the 4.8-litre stainless-bath modification; the CC-202C/205B/208B/212B siblings differ in size and configuration, so pump and bath figures here apply to this model only.
Thermal behaviour and media
- Uncooled operation spans 45–200 °C; adding the 3/8" tap-water cooling coil extends the low end to 20 °C, and a refrigerator extends it to -30 °C.
- Temperature stability is ±0.05 K; heating capacity 1.5–2.1 kW.
- Pump delivers up to 14 l/min at 0.25 bar pressure side and 10.5 l/min at 0.17 bar suction side.
- The plastic pump is not resistant to acetone or polyglycols; use water with glysantine or ethylene glycol for freeze protection.
Installation, control and operation
- Supply 200–240 V, 1~ or 2~, 50/60 Hz, max. 10 A, external fuse 10–16 A; UL 489 branch protection required.
- Interfaces: USB (Device) and RS232; optical and acoustic alarm output.
- Adjustable overtemperature protection per DIN 12876; autostart after power failure; IP20, ambient 5–40 °C.
- Optional factory-fitted Pt100 sensor connection (#10688) displays, but does not control, process temperature.
Integration & Operating Notes
The unit draws up to 10 A from a 200–240 V, 1~ or 2~, 50/60 Hz supply and requires an external fuse of at least 10 A and at most 16 A. An external branch circuit protection device according to UL 489 must be installed; omitting it leaves the installation non-compliant and unprotected against fault current.
Voltage may deviate ±10% only as long as the frequency tolerance does not run in the opposite direction; for example, -10% voltage combined with +3% frequency is not permitted. Verify the site supply envelope before connecting, because a combined deviation outside this rule can stress the electronics and pump motor.
The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). For freeze protection use water mixed with glysantine or ethylene glycol. Specifying an incompatible fluid risks pump damage and loss of circulation in the external circuit.
Without cooling the bath operates only from 45 °C upward; water cooling through the 3/8" coil lowers the limit to 20 °C, and a refrigerator is needed to reach -30 °C. Plan the cooling source before commissioning if sub-45 °C operation is required, since the uncooled unit cannot reach it.
The unit is IP20 and must be installed where ambient temperature stays between 5 °C and 40 °C. Space requirements per the operating conditions must be respected for operation and ventilation; installing in a confined or hot location can derate performance or trip protection.
The optional Pt100 sensor connection (#10688) displays, but does not control, an external process temperature and is available only factory fitted at additional cost. It cannot be retrofitted in the field, so include it in the original order if external monitoring is needed.
The autostart function restarts the unit following a power failure. For unattended processes assess whether automatic restart is acceptable for the connected external circuit, and rely on the adjustable overtemperature protection per DIN 12876 (safety class III/FL) as the protective layer.
Selection & Integration Q&A
Pick the 205B when a 4.8-litre stainless bath (2.5 l minimum fill) and 105 x 90 x 150 mm opening suit the workload. Larger siblings offer more bath volume; the 205B keeps the same 45–200 °C range and ±0.05 K stability in a compact 178 x 337 x 355 mm, 9 kg envelope.
Use the M16x1 male pump connection with the included NW12 hose connector and sleeve nuts. For 20 °C operation add the 3/8" tap-water cooling coil; for -30 °C a refrigerator is required. The pump is not resistant to acetone or polyglycols, so specify water/glysantine or water/ethylene glycol.
Provide 200–240 V, 1~ or 2~, 50/60 Hz with max. 10 A draw, an external fuse of 10–16 A and UL 489 branch circuit protection. Voltage tolerance is ±10% only if frequency tolerance does not oppose it; ambient must stay 5–40 °C and the unit is IP20.
The Pt100 sensor connection (#10688) is factory-fitted only, at extra cost, and displays rather than controls process temperature. Cooling water supply and refrigerator are external. Included accessories are hose connector NW12, blank plug, sleeve nuts, bath-bridge cover, cooling coil and pump adaptor.