Peter Huber Kältemaschinenbau KISS 205B Badumwälzthermostat mit Edelstahlbad
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS 205B Badumwälzthermostat mit Edelstahlbad

Peter Huber Kältemaschinenbau KISS 205B Badumwälzthermostat mit Edelstahlbad — Thermostatic baths.

The KISS 205B is a 4.8-litre stainless-steel bath thermostat with an integrated pressure/suction pump and 1.5–2.1 kW heating, covering 45–200 °C uncooled, 20–200 °C with water cooling and -30–200 °C with a refrigerator. Its ±0.05 K stability and DIN 12876 III/FL safety classification distinguish it within the KISS series for precise external-circuit tempering.

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection

Model ControllerInterfacePump connectionTemperature range, °CWeight, kgMax delivery flow rate, liter/minute
KISS 205B Badumwälzthermostat mit Edelstahlbad Current KISSUSB (Device), RS232 InteM16x1 AG male45–200914
CC-202C Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200922
CC-205B Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200927
CC-208B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–2001027
CC-212B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001122
CC-215B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–20012—
CC-220B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001427
CC-225B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001627
KISS 202C KISSUSB (Device), RS232 InteM16x1 AG male45–200814
KISS 208B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001014
KISS 212B KISSUSB (Device), RS232 InteM16x1 AG male25–2001114
KISS 215B KISSUSB (Device), RS232 InteM16x1 AG male25–2001214
KISS 220B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001414
KISS 225B KISSUSB (Device), RS232 InteM16x1 AG male25–2001614

Full specifications

Core Performance & Functional Specifications
Alarm output optic, acoustic
Dimensions (w × D × H) 178 x 337 x 355 mm
Maximum delivery flow rate 14 l/min (pressure pump max.; 0, 25 bar)
Pump connection M16x1 AG male
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 45...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 05 K
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump max.; 0, 17 bar)
Bath opening WxD/BD mfr 105 x 90 / 150 mm (WxD/BD)
Bath volume mfr 4, 8 l
Filling volume min. mfr 2, 5 l (min.)
Dimensions, Weight & Mechanical Design
Weight 9 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behaviour and media

Installation, control and operation

Integration & Operating Notes

Electrical
Provide UL 489 branch protection and a 10–16 A fuse

The unit draws up to 10 A from a 200–240 V, 1~ or 2~, 50/60 Hz supply and requires an external fuse of at least 10 A and at most 16 A. An external branch circuit protection device according to UL 489 must be installed; omitting it leaves the installation non-compliant and unprotected against fault current.

Datasheet, Technical Specifications, page 2
Electrical
Keep voltage and frequency tolerances from opposing each other

Voltage may deviate ±10% only as long as the frequency tolerance does not run in the opposite direction; for example, -10% voltage combined with +3% frequency is not permitted. Verify the site supply envelope before connecting, because a combined deviation outside this rule can stress the electronics and pump motor.

Datasheet, Technical Specifications, page 2
Process Limits
Select heat-transfer fluid compatible with the plastic pump

The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). For freeze protection use water mixed with glysantine or ethylene glycol. Specifying an incompatible fluid risks pump damage and loss of circulation in the external circuit.

Datasheet, Technical Specifications, page 2
Process Limits
Cooling method sets the achievable low temperature

Without cooling the bath operates only from 45 °C upward; water cooling through the 3/8" coil lowers the limit to 20 °C, and a refrigerator is needed to reach -30 °C. Plan the cooling source before commissioning if sub-45 °C operation is required, since the uncooled unit cannot reach it.

Datasheet, Technical Specifications, page 1
Mounting
Respect space requirements and the 5–40 °C ambient window

The unit is IP20 and must be installed where ambient temperature stays between 5 °C and 40 °C. Space requirements per the operating conditions must be respected for operation and ventilation; installing in a confined or hot location can derate performance or trip protection.

Datasheet, Technical Specifications, page 2
Control & I/O
Pt100 process display must be ordered factory-fitted

The optional Pt100 sensor connection (#10688) displays, but does not control, an external process temperature and is available only factory fitted at additional cost. It cannot be retrofitted in the field, so include it in the original order if external monitoring is needed.

Datasheet, Technical Specifications, page 1
Control & I/O
Autostart restarts the unit automatically after power loss

The autostart function restarts the unit following a power failure. For unattended processes assess whether automatic restart is acceptable for the connected external circuit, and rely on the adjustable overtemperature protection per DIN 12876 (safety class III/FL) as the protective layer.

Datasheet, Technical Specifications, page 1

Selection & Integration Q&A

When should I choose the KISS 205B rather than a larger sibling such as the CC-208B or CC-212B?

Pick the 205B when a 4.8-litre stainless bath (2.5 l minimum fill) and 105 x 90 x 150 mm opening suit the workload. Larger siblings offer more bath volume; the 205B keeps the same 45–200 °C range and ±0.05 K stability in a compact 178 x 337 x 355 mm, 9 kg envelope.

What plumbing and cooling accessories are needed to connect an external circuit and reach 20 °C?

Use the M16x1 male pump connection with the included NW12 hose connector and sleeve nuts. For 20 °C operation add the 3/8" tap-water cooling coil; for -30 °C a refrigerator is required. The pump is not resistant to acetone or polyglycols, so specify water/glysantine or water/ethylene glycol.

What electrical supply and protection must be provided at the installation site?

Provide 200–240 V, 1~ or 2~, 50/60 Hz with max. 10 A draw, an external fuse of 10–16 A and UL 489 branch circuit protection. Voltage tolerance is ±10% only if frequency tolerance does not oppose it; ambient must stay 5–40 °C and the unit is IP20.

What is not included in base delivery and what must be ordered or planned separately?

The Pt100 sensor connection (#10688) is factory-fitted only, at extra cost, and displays rather than controls process temperature. Cooling water supply and refrigerator are external. Included accessories are hose connector NW12, blank plug, sleeve nuts, bath-bridge cover, cooling coil and pump adaptor.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS 205B Immersion Circulator
Datasheet ·94,6 КБ ·2 pp.
View PDF