Peter Huber Kältemaschinenbau KISS 215B
Peter Huber Kältemaschinenbau KISS 215B — Thermostatic baths.
The KISS 215B is a 15-litre stainless-steel bath immersion circulator with a non-variable pressure/suction pump and KISS controller. Its usable range is configuration-dependent: 25–200 °C uncooled, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. It serves as the temperature-control core for external or internal circulation loops in laboratory processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection
| Model | Controller | Interface | Pump connection | Temperature range, °C | Weight, kg | Max delivery flow rate, liter/minute |
|---|---|---|---|---|---|---|
| KISS 215B Current | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 12 | 14 |
| CC-202C | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 22 |
| CC-205B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 27 |
| CC-208B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 10 | 27 |
| CC-212B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 11 | 22 |
| CC-215B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 12 | — |
| CC-220B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 14 | 27 |
| CC-225B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 16 | 27 |
| KISS 202C | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 8 | 14 |
| KISS 205B Badumwälzthermostat mit Edelstahlbad | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 9 | 14 |
| KISS 208B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 10 | 14 |
| KISS 212B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 11 | 14 |
| KISS 220B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 14 | 14 |
| KISS 225B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 16 | 14 |
Full specifications
| Alarm output | optic, acoustic |
| Bath volume | 15 l |
| Dimensions (w × D × H) | 350 x 375 x 425 mm |
| Filling volume | 10 l (min) |
| Maximum delivery flow rate | 14 l/min (pressure pump; 0, 25 bar) |
| Pump connection | M16x1 AG male (Option, #19607) |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Temperature range | 25...200 °C with cooling unit: -30...200 °C · with water cooling: 20...200 °C |
| Temperature stability mfr | ±0, 05 K |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min (suction pump; 0, 17 bar) |
| Bath opening WxD/BD mfr | 290 x 152 / 200 mm (WxD/depth) |
| Weight | 12 kg |
| Current consumption | 10 A (max) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The KISS 215B is a bath-mounted immersion circulator with a 15 l stainless-steel bath (minimum filling 10 l) and a bath opening of 290 x 152 mm, 200 mm deep.
- It uses a pressure and suction pump with fixed speed; maximum delivery is 14 l/min at 0.25 bar on the pressure side and 10.5 l/min at 0.17 bar on the suction side.
- Within the series, the 215B is the 15-litre variant; sibling models CC-202C, CC-205B, CC-208B and CC-212B are smaller-bath configurations. Choose the 215B when the working volume and immersion depth require the larger bath opening and 10 l minimum fill.
Thermal behavior and media
- Temperature stability is stated as ±0.05 K. Heating capacity is 1.5–2.1 kW.
- The achievable range is not a single figure: without cooling the unit covers 25–200 °C; with water cooling 20–200 °C; with a refrigerator -30–200 °C. The cooling method must be selected at the system level.
- The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer); these media require a more resistant plastic or must be avoided. For operation near or below 0 °C with water-based fluid, mix water with glysantine or ethylene glycol for freeze protection.
Installation, control and operation
- Supply is 200–240 V, 1~ or 2~, 50/60 Hz, with maximum current 10 A. The datasheet requires a minimum 10 A and maximum 16 A fuse, and an external branch-circuit protection device according to UL 489.
- Voltage and frequency tolerances are each ±10% but must not run in opposite directions.
- The controller is the KISS type with OLED display; interfaces are USB (Device) and RS232. Alarm output is optic and acoustic. Overtemperature protection is adjustable per DIN 12876; safety class is III / FL.
- The bath bridge has a hole for a cooling probe, e.g. for an immersion cooler TC45–TC100E. An optional factory-fitted Pt100 sensor connection (#10688) displays, but does not control, process temperature.
- The autostart function restarts the unit after a power failure; this may be undesirable for some processes. Ambient temperature must remain 5–40 °C. Protection rating is IP20. EMC classification is EN 55011 Class A, Group 1.
Integration & Operating Notes
The datasheet requires an external branch-circuit protection device according to UL 489. The supply must be 200–240 V, 1~ or 2~, 50/60 Hz, with a fuse rated between 10 A minimum and 16 A maximum; the maximum current draw is 10 A. Plan the panel and cabling around these limits before installation.
Both voltage and frequency may deviate by ±10%, but the datasheet prohibits a combination where the frequency tolerance runs in the opposite direction to the voltage tolerance. Verify the site supply or generator envelope against this rule; a supply that is, for example, -10% voltage with +3% frequency is not permitted.
The usable temperature range depends on the cooling utility: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. The bath bridge includes a hole for a cooling probe such as an immersion cooler TC45–TC100E. Specify the cooling configuration at the system level; the bath alone cannot reach sub-ambient setpoints.
The plastic pump is not resistant to acetone or polyglycols (depending on the manufacturer). Using these media with the standard pump risks damage. Either avoid these fluids or order a more resistant plastic. For water-based operation near or below 0 °C, mix water with glysantine or ethylene glycol for freeze protection.
The optional Pt100 sensor connection (#10688) is available only factory-fitted and cannot be retrofitted in the field. It displays, but does not control, process temperature. If process-temperature monitoring is required, include this option in the original order; it cannot be added later.
The autostart function restarts the unit automatically after a power failure. For processes where an unattended restart could be hazardous or could invalidate a run, this behavior must be considered at the system level. The datasheet notes it may be undesirable for some processes; plan the process logic accordingly.
Single- and two-phase devices are supplied with a power cable and a country-specific plug, which must be specified when ordering. Confirm the site receptacle type before placing the order to avoid a mismatch. The unit is rated IP20 and must be installed where ambient temperature remains between 5 °C and 40 °C.
Selection & Integration Q&A
Choose the 215B when the application needs the 15 l bath with a 10 l minimum fill and the 290 x 152 mm opening for larger immersion coolers or sample loads. Smaller siblings suit lower working volumes; the 215B trades footprint for bath capacity and immersion depth.
Provide a 200–240 V, 1~ or 2~, 50/60 Hz supply with a 10–16 A fuse and an external UL 489 branch-circuit protection device. For sub-ambient work, add water cooling or a refrigerator; the bath bridge accepts an immersion cooler TC45–TC100E. Specify the country-specific plug at order.
Ambient temperature must stay between 5 °C and 40 °C during operation. The unit is IP20 and EN 55011 Class A, Group 1, so additional EMC measures may be needed in residential or light-industrial environments. Voltage and frequency tolerances are ±10% each but must not oppose each other.
The base unit does not include the external UL 489 branch protection, the cooling probe or immersion cooler, or the factory-fitted Pt100 sensor connection (#10688), which cannot be retrofitted. The country-specific plug must be specified at order. For freeze protection below 0 °C, glysantine or ethylene glycol must be added to water.