Peter Huber Kältemaschinenbau CC-225B
Peter Huber Kältemaschinenbau CC-225B — Thermostatic baths.
The CC-225B is a 23.5-litre stainless-steel bath with an integrated variable-speed pressure/suction pump and Pilot ONE controller, rated for 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. It is the largest bath in the CC-20xB series, distinguished by its 17-litre minimum fill and 27 l/min pressure-pump delivery, and serves as a temperature-control source for external circuits and immersed samples.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection
| Model | Controller | Interface | Pump connection | Temperature range, °C | Weight, kg | Max delivery flow rate, liter/minute |
|---|---|---|---|---|---|---|
| CC-225B Current | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 16 | 27 |
| CC-202C | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 22 |
| CC-205B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 27 |
| CC-208B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 10 | 27 |
| CC-212B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 11 | 22 |
| CC-215B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 12 | — |
| CC-220B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 14 | 27 |
| KISS 202C | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 8 | 14 |
| KISS 205B Badumwälzthermostat mit Edelstahlbad | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 9 | 14 |
| KISS 208B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 10 | 14 |
| KISS 212B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 11 | 14 |
| KISS 215B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 12 | 14 |
| KISS 220B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 14 | 14 |
| KISS 225B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 16 | 14 |
Full specifications
| Filling volume | 17 l (min) |
| Maximum delivery flow rate | 27 l/min (pressure pump; also 0, 7 bar) suction pump; also 0, 4 bar: 22 l/min |
| Pump connection | M16x1 AG male (Option, #19607) (Option, #19607) |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 25...200 °C with cooling unit: -30...200 °C · with water cooling: 20...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 350 x 555 x 425 mm |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Bath opening WxD/BD mfr | 290 x 329 / 200 |
| Bath volume mfr | 23, 5 l |
| Weight | 16 kg |
| Current consumption | 10 A (max) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The CC-225B is an immersion circulator/heating bath built around a 23.5-litre stainless-steel bath (bath opening 290 × 329 mm, depth 200 mm) with a bath bridge that includes a hole for an immersion cooling probe (TC45–TC100E). Within the CC-20xB series it is the largest-capacity variant; the CC-202C, CC-205B, CC-208B and CC-212B are smaller siblings. The unit weighs 16 kg and measures 350 × 555 × 425 mm (W × D × H). Heating capacity is 1.5–2.1 kW, and temperature stability is ±0.02 K. The controller is the Pilot ONE, with Ethernet, USB (host and device) and RS232 interfaces and a Pt100 external sensor connection.
Thermal behavior and media
Achievable temperature depends on the cooling method: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. The variable-speed pressure and suction pump delivers up to 27 l/min at 0.7 bar (pressure side) and 22 l/min at 0.4 bar (suction side), bounding the external circuit resistance that can be served. Minimum filling is 17 l of the 23.5-litre bath volume. When a Huber pump adapter is used, polyglycol must not be employed as the heat transfer fluid. Adjustable overtemperature protection according to DIN 12876 is provided and must be set for the application.
Installation, electrical and control
Supply is 200–240 V, 1~ or 2~, 50/60 Hz, with maximum current 10 A; branch protection must be sized between 10 A and 16 A, and external branch circuit protection according to UL 489 is required. Voltage and frequency tolerances of ±10% are acceptable only if they do not act in opposite directions. Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering. The enclosure is IP20, so the unit must be kept dry and not exposed to water; ambient temperature must remain between 5 °C and 40 °C. The equipment is classified Class A, Group 1 to EN55011 and may cause interference in residential environments.
Integration & Operating Notes
The unit draws up to 10 A at 200–240 V; the site must provide external branch circuit protection according to UL 489 sized between 10 A and 16 A. Undersizing causes nuisance trips at full heating load, while oversizing leaves the supply cable unprotected.
Voltage and frequency tolerances of ±10% are only acceptable if they do not act in opposite directions. A supply where voltage and frequency deviate in opposite senses exceeds the permitted envelope and must be corrected before commissioning.
Without cooling the bath operates from 25 °C; water cooling extends the lower limit to 20 °C; only a refrigerator reaches -30 °C. The process temperature must be chosen together with the cooling configuration, and an immersion cooler TC45–TC100E can be fitted in the bath-bridge hole.
When a Huber pump adapter is fitted, polyglycol is not permissible as the heat transfer fluid. Fluid selection must be fixed before ordering the adapter, since the wrong combination cannot be corrected by a setting.
The bath holds 23.5 l but requires a minimum filling of 17 l. Operating below this level risks pump cavitation and invalid temperature control; the heat transfer fluid charge must be planned for the full external loop plus this minimum.
The IP20 rating excludes water exposure, so the unit must be installed in a dry, protected location. Ambient temperature must stay between 5 °C and 40 °C; outside this band the controller and pump may not meet their specified performance.
Single- and two-phase devices are delivered with a power cable and a country-specific plug that must be stated on the order. Three-phase devices below 63 A arrive with cable but without plug, so termination hardware must be planned by the site.
The unit provides adjustable overtemperature protection according to DIN 12876. The cut-out must be configured for the fluid and process so that it protects both the heat transfer fluid and the immersed sample.
Selection & Integration Q&A
Pick the CC-225B when the process needs the 23.5-litre bath and 17-litre minimum fill, or the 27 l/min pressure-pump delivery for a large external loop. Smaller siblings suit lower volume demands; the CC-225B is the largest bath in the series.
The external loop must be servable by 27 l/min at 0.7 bar (pressure) and 22 l/min at 0.4 bar (suction). If a Huber pump adapter is used, polyglycol is not permissible as the heat transfer fluid.
Supply 200–240 V, 1~ or 2~, 50/60 Hz, max 10 A, with a UL 489 breaker of 10–16 A. Ambient must stay 5–40 °C, the IP20 enclosure must be kept dry, and ±10% voltage/frequency tolerances must not act in opposite directions.
Without cooling the bath starts at 25 °C; water cooling gives 20 °C, and a refrigerator is required for -30 °C. The bath bridge has a hole for an immersion cooler TC45–TC100E. The country-specific plug must be specified at ordering.