Peter Huber Kältemaschinenbau Grande Fleur w Dynamisches
Peter Huber Kältemaschinenbau Grande Fleur w Dynamisches — Temperature control units.
The Grande Fleur w Dynamisches is a water-cooled temperature control unit in the Unistate Bis 45C series, rated for -40 to 200 °C with a variable-speed pressure pump and Pilot ONE controller. Its water-cooled condenser using natural refrigerant R-1270 (GWP 0) distinguishes it from air-cooled siblings, suiting indoor installations with a 5–20 °C cooling water supply.
Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Grande Fleur w Dynamisches Current | — | 0, 01 | 208-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -40–200 |
| Grande Fleur Dynamisches | — | — | — | — | — | — |
| Grande Fleur w-eo Dynamisches | — | — | — | — | — | — |
| Petite Fleur-eo Dynamisches | 58 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Petite Fleur w Dynamisches | — | 0, 01 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Unistat 410w Dynamisches | 61 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -45–250 |
| Unistat 425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat 430w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P405w Dynamisches | — | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | -45–250 |
| Unistat P425 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430w Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
Full specifications
| Control output | POKO ONE |
| Expansion volume | 3 l |
| Pump connection | M24x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -40...200 °C |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-1270 (A3, H220) (1st stage) |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 36 л/час (at 15°C, flow -20°C) at 15°C, flow 0°C: 49 l/h · at 15°C, flow 15°C: 51 l/h |
| Cooling water differential pressure | 2 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled refrigeration circuit with R-1270 (A3, H220), GWP 0; first-stage refrigerant only.
- Variable-speed pressure pump with M24x1.5 male connection; pump speed adjustable via controller.
- Pilot ONE controller with POKO ONE control output; display resolution 0.01 K; external Pt100 sensor input.
- Thermally passive expansion tank (3 l) prevents HTF vapour formation and moisture absorption.
- Safety class III/FL with adjustable overtemperature protection per DIN 12876.
- Sibling models: Grande Fleur Dynamisches (air-cooled base), Grande Fleur-eo Dynamisches, Grande Fleur w-eo Dynamisches, Petite Fleur Dynamisches (58 dB(A)). The 'w' suffix denotes water cooling; 'eo' variants differ in configuration.
Behavior under load and media
- Temperature range -40 to 200 °C; cooling capacity with ethanol falls from 0.6 kW at 0 °C to 0.04 kW at -40 °C, so process load must be matched to actual operating temperature.
- Cooling water consumption at 15 °C supply: 49 l/h at 0 °C flow, 51 l/h at 15 °C flow, 36 l/h at -20 °C flow.
- Maximum permissible kinematic viscosity of heat transfer fluid: 50 mm²/s, limiting fluid selection at low temperatures.
- Cooling water circuit materials: copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM — only compatible water may be used.
Installation, connection and operation
- Cooling water connection G1/2 male; requires minimum differential pressure 2 bar and maximum pressure 6 bar.
- Cooling water intake temperature 5–20 °C; insulate pipes if water temperature can fall below dew point.
- Pump connections use bore shape Y (60°) per DIN 3863; ball socket fittings per DIN 3863 with sleeve nuts per DIN 3870 required.
- Power supply 208–240 V 1~/2~ 50/60 Hz; single/two-phase devices delivered with country-specific plug (specify at order). Voltage tolerance ±5%, frequency ±2%, not simultaneously in opposite directions.
- IP20 protection; ambient temperature 5–40 °C; dry indoor location required.
- Interfaces: Ethernet, USB (Host and Device), RS232.
Integration & Operating Notes
Site water supply must provide at least 2 bar differential pressure and must never exceed 6 bar at the unit inlet. Exceeding the maximum risks damage to the cooling circuit; insufficient differential pressure prevents adequate heat rejection and can trip the unit.
The unit is designed for cooling water between 5 °C and 20 °C. If the water temperature can fall below the ambient dew point, pipes must be insulated to prevent condensation on the cooling water circuit.
Supply voltage tolerance is ±5% and frequency tolerance ±2%, but combinations such as -5% voltage with +2% frequency are not permitted. Verify site power quality against this envelope before commissioning.
Single/two-phase devices (100–240 V) are delivered with a power cable and country-specific plug. The plug type must be stated when ordering; otherwise the unit may arrive without a suitable connector for the site.
The unit is rated IP20 and must operate between 5 °C and 40 °C ambient temperature. It must be installed in a dry, protected indoor location; moisture or dust ingress can damage electronics and invalidate the protection rating.
At the lowest operating temperature, the heat transfer fluid's kinematic viscosity must remain below 50 mm²/s. This limits fluid selection for -40 °C duty and affects pump performance and cooling capacity.
With ethanol, cooling capacity falls from 0.6 kW at 0 °C to 0.04 kW at -40 °C. Process heat loads must be matched to the actual operating temperature; a load acceptable at mid-range may be unachievable at the low end.
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Aggressive or incompatible water can corrode or degrade these materials; verify water quality before connection.
The unit uses R-1270 (A3, H220) in the first stage. Safety measures for flammable refrigerants apply; handling and service must follow local regulations for A3 refrigerants.
Selection & Integration Q&A
Choose the 'w' water-cooled variant when a 5–20 °C cooling water supply with 2–6 bar differential is available and air-cooled heat rejection is impractical. The Petite Fleur is smaller and rated 58 dB(A); select it only if its capacity fits. The 'eo' variants differ in configuration and should be compared on their own datasheets.
Provide cooling water at 5–20 °C with minimum 2 bar differential and maximum 6 bar pressure. Connect via G1/2 male fitting. Insulate pipes if water can fall below dew point. Water must be compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
Install indoors, dry, IP20, ambient 5–40 °C. Supply 208–240 V 1~/2~ 50/60 Hz with ±5% voltage and ±2% frequency, not opposing. Specify country-specific plug at order. Pump connections require DIN 3863 ball socket fittings and DIN 3870 sleeve nuts.
Base delivery includes power cable with country-specific plug only if specified. Order DIN 3863 ball socket fittings and DIN 3870 sleeve nuts for pump connections, plus G1/2 hoses for cooling water. Heat transfer fluid must have viscosity ≤50 mm²/s at lowest operating temperature. No maintenance interval is stated in the provided datasheet.
Dimensional Drawings & Technical Documentation
Dimensional drawings
Click drawing to inspect vector dimensions