Peter Huber Kältemaschinenbau Unistat P425w Dynamisches
Peter Huber Kältemaschinenbau Unistat P425w Dynamisches — Temperature control units with Noise level 61 dB.
The Unistat P425w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -40 to 250 °C at 0.01 K display resolution. Its 2 kW heating capacity and 97 l/min pump delivery suit externally closed circuits with high pressure drop, while cooling power falls to 0.1 kW at -40 °C. It serves as a precise heating/cooling source for reactor jackets and laboratory process loops.
Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P425w Dynamisches Current | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Grande Fleur Dynamisches | — | — | — | — | — | — |
| Grande Fleur w Dynamisches | — | 0, 01 | 208-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -40–200 |
| Grande Fleur w-eo Dynamisches | — | — | — | — | — | — |
| Petite Fleur-eo Dynamisches | 58 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Petite Fleur w Dynamisches | — | 0, 01 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Unistat 410w Dynamisches | 61 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -45–250 |
| Unistat 425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat 430w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P405w Dynamisches | — | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | -45–250 |
| Unistat P425 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430w Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
Full specifications
| Noise level | 61 dB(A) |
| Control output | POKO ONE |
| Heating capacity | 2 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -40...250 °C |
| Weight | 178 kg |
| Current consumption | 10 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 150 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing water-cooled cabinet (w suffix) measuring 460x554x1453 mm and weighing 178 kg; IP20, ambient 5–40 °C.
- Magnetically coupled stainless-steel pump, speed-adjustable, with M30x1.5 male pressure connection; cooling water G1/2 male.
- Within the Unistat series, the P designation targets applications with high pressure drop; the w suffix denotes water cooling rather than air cooling. Sibling Grande Fleur variants differ in configuration and are not interchangeable with this model's ratings.
Working behavior
- Operating range -40 to 250 °C with 0.01 K display resolution; heating capacity 2 kW.
- Pump delivers up to 97 l/min and 3 bar max, with 86 l/min at 0.5 bar; external circuit pressure drop must be calculated against this curve.
- Cooling capacity with ethanol drops to 0.1 kW at -40 °C, so low-temperature heat loads must be sized accordingly.
- Refrigerant R-452A, GWP 2141, 0.97 kg charge (2.1 t CO2e) — relevant for F-gas accounting.
Installation and control
- Requires 400 V 3~ 50 Hz supply with 3x16 A fuse; leakage current exceeds 3.5 mA, so PE and RCD selection must account for it.
- Cooling water: 1–6 bar differential pressure, 150 l/h at 15 °C, intake 5–20 °C; insulate internal cooling-water pipes when water is below the condensation point.
- Pilot ONE controller with POKO ONE control output; interfaces Ethernet, USB host/device, RS232; external Pt100 sensor connection.
- Safety class III/FL; only heat transfer fluids with fire point ≥65 °C for FL rating; ethanol requires supervised operation.
Integration & Operating Notes
The unit draws 10 A on a 400 V 3~ 50 Hz supply protected by 3x16 A, but leakage current exceeds 3.5 mA. Protective earth continuity and residual-current device selection must be engineered for this leakage level; voltage tolerance is ±5% only with ±2% frequency, and -5% voltage with +2% frequency is not permitted.
The cooling water circuit requires 150 l/h at 15 °C with 1 bar differential pressure and a maximum of 6 bar; intake water must remain between 5 °C and 20 °C. Wetted materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used. Insulate internal cooling-water pipes when water temperature falls below the condensation point.
Horizontal transport damages the compressor, and only an emptied unit may be moved. Casters are not for transport and carry only 25% of total mass symmetrically; use an industrial truck. Leveling feet must be activated and parking brakes engaged before commissioning. Allow about 2 hours acclimation when moving between cold and warm environments before switching on.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage to the application. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. The unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet.
In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles, but this mode requires the compressor to be set to Permanently ON. Power Failure Auto resumes temperature control automatically after an outage if it was active; choose the power-interruption behavior according to whether unattended restart is acceptable.
Shut off the cooling water supply even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning contractors.
Selection & Integration Q&A
Choose this P variant when the external circuit has high pressure drop and needs up to 97 l/min at 3 bar; the w suffix means water cooling, so a facility water supply at 1–6 bar and 5–20 °C is required. Grande Fleur siblings differ in configuration and should be evaluated against their own datasheets.
The pump connection is M30x1.5 male using bore shape Y (60°) per DIN 3863 with ball socket and sleeve nut per DIN 3870; cooling water is G1/2 male. A 400 V 3~ 50 Hz supply with 3x16 A fuse is required, and an external Pt100 sensor can be connected. Matching hoses and fittings are not included and must be ordered.
Install upright only, never in an ATEX zone, with ambient 5–40 °C and IP20 protection. Activate leveling feet and parking brakes; casters are not for transport. Allow 2 hours acclimation after temperature changes. For closed applications keep the expansion vessel unlocked; for open bath applications lock it.
Periodically check hoses and connections for leaks and quality and replace as needed; shut off cooling water even overnight. Clean liquefier fins and the hat-type strainer/dirt trap; inspect and replace heat transfer fluid and rinse the circuit. Refrigerant-circuit work requires approved contractors. Hoses, fittings, fluid and a drip tray are not included.