Peter Huber Kältemaschinenbau Unistat P405w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat P405w Dynamisches

Peter Huber Kältemaschinenbau Unistat P405w Dynamisches — Temperature control units.

The Unistat P405w Dynamisches is a water-cooled, high-pressure-drop circulating thermostat with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -45 to 250 °C with 3 kW heating. Its 'P' designation targets applications with high pressure drop, distinguishing it from the Grande Fleur siblings. It serves closed-loop temperature control of externally closed processes requiring elevated pump head.

Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionPower supplyPump connectionPump typeTemperature range, °C
Unistat P405w Dynamisches Current —0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st-45–250
Grande Fleur Dynamisches ——————
Grande Fleur w Dynamisches —0, 01208-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -40–200
Grande Fleur w-eo Dynamisches ——————
Petite Fleur-eo Dynamisches 580, 01 K220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Petite Fleur w Dynamisches —0, 01220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Unistat 410w Dynamisches 610, 01 K220-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -45–250
Unistat 425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat 430w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P425 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430w Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 3 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -45...250 °C
Electrical, Power & Drive Specifications
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Cooling & Media
Cooling water must be supplied and returned on site

The refrigeration machine is water-cooled, so the installation site must provide cooling water supply and return at G1/2 male connections. Differential pressure must be at least 1 bar and must not exceed 6 bar; intake temperature must lie between 5 and 20 °C, and pipes must be insulated below the condensation point.

Unistat P405w Technical Data, page 1–2
Electrical
Provide UL 489 branch protection and a site plug

Supply is 460 V 3~ 60 Hz with ±5 % voltage and simultaneous ±2 % frequency tolerance. External branch circuit protection according to UL 489 is mandatory. Because current is below 63 A, the unit arrives with cable but without plug, so the installer must provide a suitable plug or hardwired connection; connection must be made by an electrician.

Unistat P405w Technical Data, page 2; Unistat Operation Manual, page 37
Mounting
Transport upright, empty, and acclimate before power-on

Only transport the unit upright and emptied; horizontal transport damages the compressor. Casters are not for transport and carry only 25 % of total mass symmetrically. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to avoid condensation-induced electrical faults.

Unistat Operation Manual, page 22 and 29
Process Limits
Expansion vessel setting depends on open vs closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage to the application. For externally open bath applications the vessel must be locked to prevent fluid loss. The thermally passive tank otherwise prevents HTF vapour and moisture absorption.

Unistat Operation Manual, page 33; Unistat P405w Technical Data, page 1
Control & I/O
Process Safety needs the compressor set to Permanently ON

In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles. This mode requires the compressor to be set to Permanently ON; otherwise the protective cooling response is not available. Configure the power-failure behavior (Off/Standby vs ON) according to whether unattended restart is acceptable.

Unistat Operation Manual, page 19–21
Maintenance
Shut off cooling water even for overnight downtime

Close the cooling water supply even for short downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Periodically check hoses and hose connections for leaks and quality and replace as required. Refrigerant-circuit work must be performed by approved refrigeration/air-conditioning contractors.

Unistat Operation Manual, page 29; Unistat Operation Manual, Section 1.4.2

Selection & Integration Q&A

When should I choose the P405w Dynamisches rather than a Grande Fleur sibling?

Choose the P405w when the application has high pressure drop (the 'P' designation) and requires water cooling ('w'). Grande Fleur variants are different configurations of the series and should be selected when their cooling/connection configuration matches the duty; the P405w is the high-pressure-drop, water-cooled modification.

What must I provide for the cooling water and electrical interfaces before commissioning?

Provide cooling water supply/return at G1/2 male with 1–6 bar differential pressure and 5–20 °C intake, plus a 460 V 3~ 60 Hz supply with UL 489 branch protection and a site-supplied plug or hardwired connection, since the unit ships with cable but no plug.

What installation environment and duty-cycle limits apply?

Install indoors at 5–40 °C ambient, IP20, never in an ATEX zone. Transport upright and empty, acclimate about 2 hours after temperature change, and set the expansion vessel per open/closed application. Not permitted for medical-device or direct-foodstuff use.

What maintenance and consumables are not included in base delivery?

The base unit excludes the site plug, cooling water hoses and insulation, drip tray, and heat transfer fluid. Plan periodic hose/connection inspection and replacement, cooling water shutoff for overnight downtime, and refrigerant-circuit service only by approved contractors.