Peter Huber Kältemaschinenbau Huber CS 25
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Huber CS 25

Peter Huber Kältemaschinenbau Huber CS 25 — Recirculating chillers with Noise level 60 dB.

The Huber CS 25 is a 37-litre recirculating chiller with a -20…15 °C working range, positioned as the smallest capacity unit in the Cs Chiller series. It uses R-449A refrigerant and a 50 l/min, 4.5 bar pump to serve process loops requiring moderate cooling near ambient. Its 208 V two-phase supply and IP20 rating suit indoor industrial installations.

Noise level
60 dB

Peter Huber Kältemaschinenbau Cs Chiller Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CMinimum fill volume, m³Power supplyPump connectionTemperature range, °C
Huber CS 25 Current 605–320.024208V 2~ 60HzG1/2 AG male-20–15
CS 100 by Van der Heijden 705–320.16400V 3~ 50HzG3/4 AG male-20–15
CS 100-H by Van der Heijden 715–320.175400V 3~ 50HzG3/4 AG male-20–85
CS 200 by Van der Heijden 785–320.225400V 3~ 50HzG1 AG male-20–15
CS 200-H by Van der Heijden 785–320.225400V 3~ 50HzG1 AG male-20–85
CS 20-H by Van der Heijden 575–320.024230V 1~ 50HzG3/8 AG male-20–85
CS 25 by Van der Heijden 605–320.024230V 1~ 50HzG1/2 AG male-20–15
CS 50 by Van der Heijden 635–320.1400V 3~ 50HzG3/4 AG male-20–15
CS 75 by Van der Heijden 665–320.16400V 3~ 50HzG3/4 AG male-20–15
Huber CS 200-H 785–320.225460V 3~ 60HzG1 AG male-20–85
Huber CS 250HT OD 78-20–500.16460V 3~ 60HzG3/4 AG male10–22
Huber CS 330-H OD 78-20–500.16460V 3~ 60HzG1 AG male-20–85
Huber CS 550-H OD 79-20–500.16460V 3~ 60HzG1 AG male-20–85
Huber CS 550 OD 79-20–500.16460V 3~ 60HzG1 AG male-20–15
Huber LCS 80 77-20–500.04460V 3~ 60HzG1 AG male-45–15

Full specifications

Core Performance & Functional Specifications
Noise level 60 dB(A)
Bath volume 37 l
Minimum fill volume 24 l
Pump connection G1/2 AG male
Temperature range -20...15 °C
Dimensions, Weight & Mechanical Design
Weight 121 kg
Electrical, Power & Drive Specifications
Power supply 208V 2~ 60Hz
Interfaces, Connections & Control
Digital interface RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...32 °C
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

The CS 25 is the lowest-capacity model in the Cs Chiller family; sibling units (CS 100, CS 100-H, CS 125, CS 125-H) step up in size and noise level (70–75 dB(A) versus 60 dB(A) here). The 37-litre bath and 24-litre minimum fill define a compact hydraulic circuit, while the 121 kg weight and IP20 enclosure indicate a floor-standing indoor unit. The pump connection is G1/2 AG male, and flow is trimmed by a manual bypass valve on the rear panel.

Working range and media

Installation and operation

Integration & Operating Notes

Electrical
Do not combine -5% voltage with +2% frequency

The supply tolerance envelope is ±5% voltage and ±2% frequency, but the corner case of -5% voltage together with +2% frequency is explicitly prohibited. Verify the site supply against both limits simultaneously before commissioning, not just each independently.

Huber CS 25 Technical Data, Specifications, page 1
Electrical
Provide a mating plug on site for the supplied cable

The unit is delivered with a power cable but without a plug because it falls under the 63 A three-phase threshold. The installer must procure and fit a plug rated for the 208 V 2~ 60 Hz, 12.5 A supply and protect the circuit with 2×20 A fuses.

Huber CS 25 Technical Data, page 2
Process Limits
Use 50:50 water–monoethylene glycol as the thermofluid

The recommended heat-transfer fluid is a 50:50 water–monoethylene glycol mixture. This concentration provides freeze protection down to the -20 °C lower limit and maintains heat transfer; substituting water alone risks freezing and poor performance at the cold end of the range.

Huber CS 25 Technical Data, page 2
Cooling & Media
Derate cooling capacity for ambient above 20 °C

Published cooling capacity is valid at 20 °C room temperature. If the installation space runs warmer (up to the 32 °C maximum), available cooling drops, so the loop heat load must be checked against the derated capacity rather than the catalogue value.

Huber CS 25 Technical Data, page 1
Maintenance
Record the 2 kg R-449A charge for leak checks

The refrigerant charge is 2 kg R-449A with a CO₂ equivalent of 2.79 t. This figure must be logged for periodic leak inspection and F-gas reporting obligations; R-449A is classified A1 with GHS hazard H280 (pressurized gas), so handling requires appropriate precautions.

Huber CS 25 Technical Data, page 1
Control & I/O
RS232 is the only listed remote interface

The datasheet lists RS232 as the interface. Integrators planning supervisory control should provide an RS232 port or converter on the host side; no Ethernet or fieldbus option is stated for this model, so do not assume one is present.

Huber CS 25 Technical Data, Specifications

Selection & Integration Q&A

When should I choose the CS 25 rather than a larger Cs Chiller sibling such as the CS 100 or CS 125?

Pick the CS 25 when the heat load fits its 37-litre bath and -20…15 °C range and low noise (60 dB(A)) matters. The CS 100/125 siblings are larger, noisier units (70–75 dB(A)) for higher duties; the CS 25 is the compact end of the series.

What must I provide on the process side to connect the CS 25 loop?

The pump connection is G1/2 AG male, so match it with G1/2 female fittings. Ensure the external circuit pressure drop is compatible with the pump's 50 l/min at 4.5 bar, and fill with at least 24 litres of 50:50 water–monoethylene glycol.

What is not included in the base delivery that I must plan for?

The unit ships with a power cable but no plug (three-phase below 63 A), so a site-supplied plug is needed. Thermofluid (50:50 water–monoethylene glycol) and process-side G1/2 fittings are also not part of the chiller delivery.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Huber CS 25 Technical Data
Datasheet ·106,2 КБ ·2 pp.
View PDF