Peter Huber Kältemaschinenbau Huber CS 330-H OD
Peter Huber Kältemaschinenbau Huber CS 330-H OD — Recirculating chillers with Noise level 78 dB.
The Huber CS 330-H OD is a 37-litre-bath recirculating chiller with an 8 kW heater, rated for -20 to +85 °C and outdoor installation down to -20 °C ambient. It is distinguished within the Cs series by its 460 V 3-phase 60 Hz supply, 49 A max current and 125 l/min pump flow, serving as a high-capacity temperature control unit for industrial loops requiring a 160–175 l fluid inventory.
Peter Huber Kältemaschinenbau Cs Chiller Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Minimum fill volume, m³ | Power supply | Pump connection | Temperature range, °C |
|---|---|---|---|---|---|---|
| Huber CS 330-H OD Current | 78 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| CS 100 by Van der Heijden | 70 | 5–32 | 0.16 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| CS 100-H by Van der Heijden | 71 | 5–32 | 0.175 | 400V 3~ 50Hz | G3/4 AG male | -20–85 |
| CS 200 by Van der Heijden | 78 | 5–32 | 0.225 | 400V 3~ 50Hz | G1 AG male | -20–15 |
| CS 200-H by Van der Heijden | 78 | 5–32 | 0.225 | 400V 3~ 50Hz | G1 AG male | -20–85 |
| CS 20-H by Van der Heijden | 57 | 5–32 | 0.024 | 230V 1~ 50Hz | G3/8 AG male | -20–85 |
| CS 25 by Van der Heijden | 60 | 5–32 | 0.024 | 230V 1~ 50Hz | G1/2 AG male | -20–15 |
| CS 50 by Van der Heijden | 63 | 5–32 | 0.1 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| CS 75 by Van der Heijden | 66 | 5–32 | 0.16 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| Huber CS 200-H | 78 | 5–32 | 0.225 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| Huber CS 25 | 60 | 5–32 | 0.024 | 208V 2~ 60Hz | G1/2 AG male | -20–15 |
| Huber CS 250HT OD | 78 | -20–50 | 0.16 | 460V 3~ 60Hz | G3/4 AG male | 10–22 |
| Huber CS 550-H OD | 79 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| Huber CS 550 OD | 79 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–15 |
| Huber LCS 80 | 77 | -20–50 | 0.04 | 460V 3~ 60Hz | G1 AG male | -45–15 |
Full specifications
| Noise level | 78 dB(A) |
| Bath volume | 37 l |
| Heating capacity | 8 kW |
| Maximum delivery flow rate | 125 l/min (also 4, 7 bar pump pressure) |
| Minimum fill volume | 160 l |
| Pump connection | G1 AG male |
| Temperature range | -20...85 °C |
| Weight | 530 kg |
| Current consumption | 49 A (max.) |
| Power supply | 460V 3~ 60Hz |
| Digital interface | RS232 |
| Ambient temperature limit | -20...50 °C |
| Safety class | I / NFL |
Description
Construction and series positioning
The CS 330-H OD is a recirculating chiller with a 37 l internal bath and a minimum filling capacity of 160 l for the external loop. It is the largest and only outdoor-rated variant among the listed Cs siblings (CS 100, CS 100-H, CS 125, CS 125-H), which are indoor units with lower noise levels (70–75 dB(A)) and smaller capacities. The CS 330-H OD is supplied with a remote operating panel on a 20 m cable, allowing the control interface to be mounted away from the chiller. The electrical components are rated IP54, and the unit is EMC Class A Group 1 for industrial environments.
Working range and media
- Temperature range: -20 to +85 °C, with heating capacity of 8 kW.
- Ambient temperature limit: -20 to +50 °C, but cooling capacity is only guaranteed up to +35 °C and derates above that.
- Recommended thermofluid: 50:50 water–monoethylene glycol mixture, which sets freeze protection and heat transfer basis.
- Pump connection: G1 male, with maximum delivery flow rate of 125 l/min and maximum pump pressure of 4.7 bar.
- Refrigerant: R-449A (A1, H280), 9 kg charge, 12.57 t CO2 equivalent.
Installation, connection and control
- Power supply: 460 V 3-phase 60 Hz, 49 A max current, 3x60 A fusing. Three-phase devices below 63 A are supplied with cable but without a plug, so the site must provide the connection.
- Voltage and frequency tolerances are ±5 % and ±2 %, but not simultaneously in the unfavorable direction.
- Dimensions: 1400 x 1000 x 1800 mm; weight 530 kg. Floor loading and transport route must be verified.
- Outdoor installation is permitted down to -20 °C ambient, requiring weather exposure and frost protection planning for the loop.
- Interface: RS232 for communication.
- Safety class: I / NFL.
Integration & Operating Notes
The unit draws up to 49 A per phase and requires 3x60 A fusing. The site electrical feed, cable cross-section and breaker rating must be sized accordingly before delivery. Three-phase devices below 63 A are supplied with cable but without a plug, so the customer must provide the site-side connection.
Cooling capacity is only guaranteed up to +35 °C ambient temperature and reduces above that. Sizing must use the capacity at the actual site ambient temperature, especially for outdoor installations in summer or in hot plant rooms.
The chiller is suitable for outside mounting with a minimum environmental temperature of -20 °C. Site planning must account for weather exposure and frost protection of the loop, including the recommended 50:50 water–monoethylene glycol mixture.
The operating panel is delivered as a remote control connected via a 20 m cable. Cable route and panel mounting location must be planned separately from the chiller, and the RS232 interface is available for communication.
The pump connection is G1 male with a maximum delivery flow rate of 125 l/min and maximum pump pressure of 4.7 bar. The external circuit, including piping and fittings, must be sized for these limits and for the minimum filling capacity of 160 l.
The refrigerant is R-449A (A1, H280) with a 9 kg charge and 12.57 t CO2 equivalent. Refrigerant handling, leak detection and CO2-equivalent reporting obligations apply under F-gas regulations.
Selection & Integration Q&A
Choose the CS 330-H OD when you need a 37 l bath, 8 kW heating, 125 l/min flow and outdoor installation down to -20 °C ambient. The CS 100/125 siblings are indoor units with lower noise (70–75 dB(A)) and smaller capacities, suited to indoor plant rooms with lower heat loads.
It requires a 460 V 3-phase 60 Hz supply with 49 A max current and 3x60 A fusing. Voltage and frequency tolerances are ±5 % and ±2 %, but not simultaneously in the unfavorable direction. The unit is supplied with cable but without a plug, so the site must provide the connection.
The chiller is suitable for outdoor mounting down to -20 °C ambient, but cooling capacity is only guaranteed up to +35 °C and derates above that. Electrical components are IP54, and the unit is EMC Class A Group 1 for industrial environments. Frost protection of the loop is required.
The base delivery includes a remote operating panel on a 20 m cable but no plug. The recommended thermofluid is a 50:50 water–monoethylene glycol mixture, and the loop must be filled to at least 160 l. Refrigerant R-449A (9 kg, 12.57 t CO2e) requires F-gas leak checks and reporting.