Peter Huber Kältemaschinenbau Huber CS 250HT OD
Peter Huber Kältemaschinenbau Huber CS 250HT OD — Recirculating chillers with Noise level 78 dB.
The Huber CS 250HT OD is a 460 V three-phase recirculating chiller with a 10–22 °C working range, designed for outdoor mounting down to -20 °C ambient. Its 34 kW cooling capacity at 20 °C ambient derates to 23 kW at 50 °C, so sizing must use site ambient. It serves as a central cooling source for processes requiring a stable, pump-circulated water-glycol loop.
Peter Huber Kältemaschinenbau Cs Chiller Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Minimum fill volume, m³ | Power supply | Pump connection | Temperature range, °C |
|---|---|---|---|---|---|---|
| Huber CS 250HT OD Current | 78 | -20–50 | 0.16 | 460V 3~ 60Hz | G3/4 AG male | 10–22 |
| CS 100 by Van der Heijden | 70 | 5–32 | 0.16 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| CS 100-H by Van der Heijden | 71 | 5–32 | 0.175 | 400V 3~ 50Hz | G3/4 AG male | -20–85 |
| CS 200 by Van der Heijden | 78 | 5–32 | 0.225 | 400V 3~ 50Hz | G1 AG male | -20–15 |
| CS 200-H by Van der Heijden | 78 | 5–32 | 0.225 | 400V 3~ 50Hz | G1 AG male | -20–85 |
| CS 20-H by Van der Heijden | 57 | 5–32 | 0.024 | 230V 1~ 50Hz | G3/8 AG male | -20–85 |
| CS 25 by Van der Heijden | 60 | 5–32 | 0.024 | 230V 1~ 50Hz | G1/2 AG male | -20–15 |
| CS 50 by Van der Heijden | 63 | 5–32 | 0.1 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| CS 75 by Van der Heijden | 66 | 5–32 | 0.16 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| Huber CS 200-H | 78 | 5–32 | 0.225 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| Huber CS 25 | 60 | 5–32 | 0.024 | 208V 2~ 60Hz | G1/2 AG male | -20–15 |
| Huber CS 330-H OD | 78 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| Huber CS 550-H OD | 79 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| Huber CS 550 OD | 79 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–15 |
| Huber LCS 80 | 77 | -20–50 | 0.04 | 460V 3~ 60Hz | G1 AG male | -45–15 |
Full specifications
| Noise level | 78 dB(A) (sound pressure level) |
| Bath volume | 37 l |
| Minimum fill volume | 160 l |
| Pump connection | G3/4 AG male |
| Temperature range | 10...22 °C |
| Weight | 510 kg (net) |
| Current consumption | 27 A (max) |
| Power supply | 460V 3~ 60Hz |
| Digital interface | RS232 |
| Ambient temperature limit | -20...50 °C (permissible) |
| Safety class | I / NFL |
Description
Construction and positioning
- The CS 250HT OD is the largest sibling in the Cs Chiller series listed here (CS 100, CS 100-H, CS 125, CS 125-H), with a 37 l internal bath and a 160–175 l external circuit filling capacity.
- It is a three-phase, 460 V/60 Hz unit drawing up to 27 A, supplied with a cable but no plug; the operating panel is a remote control on a 20 m cable.
- Outdoor installation is permitted from -20 °C to +50 °C ambient, with electrical components rated IP54.
Behavior under load and media
- Cooling capacity is 34 kW at 20 °C process and 20 °C ambient, falling to 23 kW at 20 °C process and 50 °C ambient; selection must use the actual ambient at the site.
- The circulation pump delivers up to 50 l/min at a maximum pressure of 6.5 bar; flow can be balanced via a manual bypass valve on the rear.
- Recommended thermofluid is a 50:50 water–monoethylene glycol mixture; the open atmospheric tank has a digital level indicator and requires 160–175 l total fill.
- Refrigerant is R-449A (A1, H280), 10 kg charge (13.97 t CO2 equivalent).
Installation, control and operation
- Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%; the combination of -5% voltage and +2% frequency is not allowed.
- Fuse protection must be 3×35 A; the unit emits EN55011 Class A, Group 1 disturbances, so EMC precautions may be needed in sensitive areas.
- Interface is RS232; the remote panel can be mounted up to 20 m from the chiller.
Integration & Operating Notes
The ±5% voltage tolerance applies only with ±2% frequency tolerance, and -5% voltage with +2% frequency is explicitly not allowed. Verify site power quality before connection to avoid damage or malfunction.
Three-phase devices under 63 A are delivered with a cable but no plug, so the installer must supply a suitable plug or hard-wired connection rated for 27 A and protected by 3×35 A fuses.
The chiller has a minimum filling capacity of 160 l and a maximum of 175 l; the external loop plus internal volume must be sized within this range to ensure proper pump operation and temperature stability.
The recommended thermofluid is a 50:50 mixture of water and monoethylene glycol; using other fluids may impair heat transfer or freeze protection. The open tank has a digital level indicator for monitoring.
The unit is suitable for outside mounting with a minimum environmental temperature of -20 °C and maximum of +50 °C, and electrical components are IP54 rated. Ensure the installation location meets these limits.
The operating panel is delivered as a remote control connected via a 20 m cable, so it can be mounted up to 20 m away from the chiller; plan cable routing accordingly.
The atmospheric open tank has a digital level indicator; maintain the fluid level within the specified filling capacity to prevent pump cavitation or overflow.
Selection & Integration Q&A
Choose the CS 250HT OD when you need higher cooling capacity (34 kW at 20 °C ambient) and can accommodate a 160–175 l circuit volume. The CS 125 and CS 100 are smaller siblings with lower noise (75 and 70 dB(A)) but are not specified for the same capacity or outdoor range.
It requires 460 V 3~ 60 Hz, draws up to 27 A, and needs 3×35 A fuses. Voltage tolerance is ±5% only with ±2% frequency tolerance; -5% voltage with +2% frequency is not allowed. A cable is supplied but no plug.
Yes, it is suitable for outside mounting from -20 °C to +50 °C ambient, with electrical components rated IP54. Cooling capacity derates from 34 kW at 20 °C ambient to 23 kW at 50 °C ambient, so size for the actual site ambient.
The base delivery includes a cable but no plug; the installer must provide a suitable plug or hard-wired connection. The recommended thermofluid is a 50:50 water–monoethylene glycol mixture, which must be supplied separately. Monitor fluid level via the digital indicator and maintain 160–175 l fill.