Peter Huber Kältemaschinenbau CS 20-H by Van der Heijden
Peter Huber Kältemaschinenbau CS 20-H by Van der Heijden — Recirculating chillers with Noise level 57 dB.
The CS 20-H is a 37-litre air-cooled recirculating chiller with a 1 kW heater, covering -20 to 85 °C on a 230 V single-phase 50 Hz supply. It is the smallest modification in the Cs Chiller series, distinguished by its 24–27 l fill volume and 10 l/min pump. It serves as a temperature-control source for external circuits requiring both cooling and active heating.
Peter Huber Kältemaschinenbau Cs Chiller Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Minimum fill volume, m³ | Power supply | Pump connection | Temperature range, °C |
|---|---|---|---|---|---|---|
| CS 20-H by Van der Heijden Current | 57 | 5–32 | 0.024 | 230V 1~ 50Hz | G3/8 AG male | -20–85 |
| CS 100 by Van der Heijden | 70 | 5–32 | 0.16 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| CS 100-H by Van der Heijden | 71 | 5–32 | 0.175 | 400V 3~ 50Hz | G3/4 AG male | -20–85 |
| CS 200 by Van der Heijden | 78 | 5–32 | 0.225 | 400V 3~ 50Hz | G1 AG male | -20–15 |
| CS 200-H by Van der Heijden | 78 | 5–32 | 0.225 | 400V 3~ 50Hz | G1 AG male | -20–85 |
| CS 25 by Van der Heijden | 60 | 5–32 | 0.024 | 230V 1~ 50Hz | G1/2 AG male | -20–15 |
| CS 50 by Van der Heijden | 63 | 5–32 | 0.1 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| CS 75 by Van der Heijden | 66 | 5–32 | 0.16 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| Huber CS 200-H | 78 | 5–32 | 0.225 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| Huber CS 25 | 60 | 5–32 | 0.024 | 208V 2~ 60Hz | G1/2 AG male | -20–15 |
| Huber CS 250HT OD | 78 | -20–50 | 0.16 | 460V 3~ 60Hz | G3/4 AG male | 10–22 |
| Huber CS 330-H OD | 78 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| Huber CS 550-H OD | 79 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| Huber CS 550 OD | 79 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–15 |
| Huber LCS 80 | 77 | -20–50 | 0.04 | 460V 3~ 60Hz | G1 AG male | -45–15 |
Full specifications
| Noise level | 57 dB(A) |
| Bath volume | 37 l |
| Heating capacity | 1 kW |
| Minimum fill volume | 24 l |
| Pump connection | G3/8 AG male |
| Temperature range | -20...85 °C |
| Weight | 115 kg |
| Current consumption | 13 A |
| Fuse rating | 16 A |
| Power supply | 230V 1~ 50Hz |
| Digital interface | RS232 |
| Ambient temperature limit | 5...32 °C |
| Protection rating | IP52 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Air-cooled, CFC/HCFC-free refrigeration machine with integrated 1 kW heating and a 37 l bath; no water-cooling circuit is required.
- Within the Cs Chiller series, the CS 20-H is the smallest unit: siblings CS 100, CS 100-H, CS 125 and CS 125-H are larger and louder (70–75 dB(A) vs 57 dB(A) here).
- Pump connections are G3/8 AG male; the circulation pump delivers up to 10 l/min at a maximum pressure of 3.5 bar.
- Protection is IP52 and safety class I / NFL; the interface is RS232.
Behaviour across the working range
- Temperature range is -20 to 85 °C. The recommended heat transfer fluid is a 50:50 water–monoethylene glycol mixture, which provides freeze protection and suitable viscosity across this span.
- Cooling capacity is rated at 20 °C ambient and drops as ambient temperature rises, so capacity must be sized for actual room conditions.
- Flow rate is adjusted via a manual bypass valve on the back of the chiller, allowing the external loop to be matched without external throttling.
- The pump has integrated overtemperature protection; this supplements but does not replace fluid-level monitoring.
Installation, connection and operation
- Requires 230 V 1~ 50 Hz, maximum current 13 A, 16 A fuse. Voltage and frequency tolerances must be met simultaneously; -5 % voltage with +2 % frequency is not permitted.
- Single-phase devices are supplied with a power cable and country-specific plug that must be specified at order.
- Ambient temperature must remain between 5 and 32 °C. IP52 permits dust and dripping water but not jets or immersion.
- Fill volume is 24 l minimum to 27 l maximum; plan expansion space accordingly.
- Refrigerant is R-449A (A1, H280), charge 1.1 kg, 1.54 t CO2 equivalent. Emissions are EN55011 Class A, Group 1, intended for industrial electromagnetic environments.
Integration & Operating Notes
The unit draws up to 13 A on a 230 V single-phase 50 Hz supply and requires a 16 A fuse. The circuit must be dedicated and the supply must satisfy voltage and frequency tolerances simultaneously; a -5 % voltage combined with +2 % frequency is explicitly not allowed.
Cooling capacity is rated at 20 °C ambient and drops as room temperature rises. Installing above 32 °C or below 5 °C violates the stated ambient limits, and sizing without accounting for actual room conditions risks undersizing the chiller.
The recommended heat transfer fluid is a 50:50 water–monoethylene glycol mixture, which provides freeze protection and appropriate viscosity over the full -20 to 85 °C range. Other fluids may not provide adequate freeze or viscosity behaviour at the low end.
The tank accepts a minimum of 24 l and a maximum of 27 l. Underfilling risks pump dry running; overfilling leaves insufficient expansion space. Plan the fill volume during commissioning against this narrow 3 l window.
The flow rate is adjusted with a manual bypass valve on the backside of the chiller, so the external loop can be matched without external throttling. The unit provides an RS232 interface for remote monitoring or control.
Single-phase devices are supplied with a power cable and a country-specific plug that must be specified when ordering. Omitting this at order time delays commissioning because the plug may not match the site socket.
The feed pump has integrated overtemperature protection, but this is a safety layer and not a substitute for proper fluid-level monitoring. Operators should still verify fill level against the 24–27 l window to avoid dry running and pump damage.
The refrigerant is R-449A, classified A1 with GHS hazard H280 (pressurized gas). The charge is 1.1 kg with 1.54 t CO2 equivalent, which informs leak reporting and F-gas obligations for the site.
The IP52 rating protects against dust and dripping water but not against water jets or immersion. The unit should be placed away from washdown zones and splash sources to avoid ingress beyond the rated protection.
Selection & Integration Q&A
Pick the CS 20-H when the load fits its 37 l bath, 10 l/min pump and 1 kW heater, and when 57 dB(A) is required. The CS 100/125 and their -H variants are larger and louder (70–75 dB(A)); choose them only when higher capacity is needed.
A dedicated 230 V 1~ 50 Hz circuit with a 16 A fuse, since the unit draws up to 13 A. Voltage and frequency tolerances must be met simultaneously; -5 % voltage with +2 % frequency is not allowed. Specify the country-specific plug at order.
Ambient must stay between 5 and 32 °C; cooling capacity is rated at 20 °C and drops as room temperature rises. IP52 allows dust and dripping water but not jets or immersion, so avoid washdown zones. The air-cooled condenser needs adequate airflow.
Use a 50:50 water–monoethylene glycol mixture and fill between 24 l and 27 l. The power cable and country-specific plug must be specified at order. Flow is set via the rear manual bypass valve; RS232 is provided for remote control.