Peter Huber Kältemaschinenbau CS 200-H by Van der Heijden
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CS 200-H by Van der Heijden

Peter Huber Kältemaschinenbau CS 200-H by Van der Heijden — Recirculating chillers with Noise level 78 dB.

The CS 200-H is a 37-litre-bath, air-cooled recirculating chiller with a 6 kW heater and a 105 l/min pump, operating from -20 to 85 °C on a 400 V three-phase supply. It is the largest of the CS series, distinguished by its 225–245 l loop volume and 78 dB(A) sound level. It serves as a central temperature-control unit for large fluid circuits requiring both heating and cooling.

Noise level
78 dB

Peter Huber Kältemaschinenbau Cs Chiller Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CMinimum fill volume, m³Power supplyPump connectionTemperature range, °C
CS 200-H by Van der Heijden Current 785–320.225400V 3~ 50HzG1 AG male-20–85
CS 100 by Van der Heijden 705–320.16400V 3~ 50HzG3/4 AG male-20–15
CS 100-H by Van der Heijden 715–320.175400V 3~ 50HzG3/4 AG male-20–85
CS 200 by Van der Heijden 785–320.225400V 3~ 50HzG1 AG male-20–15
CS 20-H by Van der Heijden 575–320.024230V 1~ 50HzG3/8 AG male-20–85
CS 25 by Van der Heijden 605–320.024230V 1~ 50HzG1/2 AG male-20–15
CS 50 by Van der Heijden 635–320.1400V 3~ 50HzG3/4 AG male-20–15
CS 75 by Van der Heijden 665–320.16400V 3~ 50HzG3/4 AG male-20–15
Huber CS 200-H 785–320.225460V 3~ 60HzG1 AG male-20–85
Huber CS 25 605–320.024208V 2~ 60HzG1/2 AG male-20–15
Huber CS 250HT OD 78-20–500.16460V 3~ 60HzG3/4 AG male10–22
Huber CS 330-H OD 78-20–500.16460V 3~ 60HzG1 AG male-20–85
Huber CS 550-H OD 79-20–500.16460V 3~ 60HzG1 AG male-20–85
Huber CS 550 OD 79-20–500.16460V 3~ 60HzG1 AG male-20–15
Huber LCS 80 77-20–500.04460V 3~ 60HzG1 AG male-45–15

Full specifications

Core Performance & Functional Specifications
Noise level 78 dB(A)
Bath volume 37 l
Heating capacity 6 kW
Maximum delivery flow rate 105 l/min (also 4, 7 bar pump pressure)
Minimum fill volume 225 l
Pump connection G1 AG male
Temperature range -20...85 °C
Dimensions, Weight & Mechanical Design
Weight 515 kg
Electrical, Power & Drive Specifications
Current consumption 23 A (max.)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Digital interface RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...32 °C
Protection rating IP52
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Operating behaviour

Installation and control

Integration & Operating Notes

Electrical
Site supply must tolerate coupled voltage/frequency deviations

The chiller accepts ±5 % voltage only when frequency stays within ±2 %, and the combination of -5 % voltage with +2 % frequency is explicitly forbidden. Verify the site supply quality against EN60034-1 before commissioning to avoid compressor or pump damage.

CS 200-H Immersion Circulator Datasheet, page 1
Electrical
Customer must provide and fit the three-phase plug

Because the unit draws 23 A (below 63 A), it ships with a cable but without a plug. The installing electrician must select and fit a plug rated for the 3×32 A fuse protection and the site's socket standard.

CS 200-H Immersion Circulator Datasheet, page 2
Process Limits
Loop volume must fall between 225 and 245 litres

The tank filling capacity is specified as a minimum of 225 l and a maximum of 245 l. Size the external circuit and expansion volume so the total fill lands in this window; underfilling risks pump cavitation and overheating, overfilling risks overflow.

CS 200-H Immersion Circulator Datasheet, page 1
Cooling & Media
Use 50:50 water–monoethylene glycol as heat-transfer fluid

The recommended thermofluid is a 50:50 water–monoethylene glycol mixture. This protects against freezing at the -20 °C lower limit and maintains heat-transfer performance; other fluids may change viscosity, pump load and cooling capacity.

CS 200-H Immersion Circulator Datasheet, page 2
Maintenance
Clean the air-cooled condenser periodically to hold capacity

The condenser is air-cooled with a high-efficiency fan. Dust and debris on the fins reduce heat rejection and cooling capacity; schedule cleaning according to site dust load and monitor capacity drift as an indicator.

CS 200-H Immersion Circulator Datasheet, page 1
Process Limits
Flow is balanced via the rear manual bypass valve

A manual bypass valve on the back of the chiller allows the delivery flow to be matched to the external circuit's resistance. Adjust it after installation to achieve the required pressure/flow point within the 105 l/min maximum and 4–7 bar pump pressure.

CS 200-H Immersion Circulator Datasheet, page 1
Electrical
Class A EMC equipment may need measures in residential areas

The unit is classified Class A, Group 1 under EN55011. It is intended for industrial environments; if installed near residential or sensitive electronics, additional EMC filtering or separation may be required.

CS 200-H Immersion Circulator Datasheet, page 2

Selection & Integration Q&A

When should I choose the CS 200-H rather than a smaller CS 100/125 variant?

Pick the CS 200-H when the circuit needs a 225–245 l fill volume, 6 kW heating and up to 105 l/min flow. The CS 100/125 siblings are smaller and quieter (70–75 dB(A)) but cannot serve large loops; choose them for lower volume and acoustic limits.

What mating parts and site services are required to connect the CS 200-H?

Provide G1 female fittings for the pump's G1 male thread, a 400 V 3~ 50 Hz supply protected at 3×32 A, and a customer-supplied plug (cable included, plug not). Fill with 50:50 water–monoethylene glycol within 225–245 l.

What ambient and duty limits apply to installing the CS 200-H?

Install in 5–32 °C ambient with free air flow to the air-cooled condenser; capacity is rated at 20 °C and derates above it. IP52 allows dripping water but not jets or immersion. Keep supply within ±5 % voltage only with ±2 % frequency.

What maintenance and accessories are needed beyond the base delivery?

Base delivery includes the cable but no plug; the customer supplies the plug and the 50:50 water–glycol fluid. Periodically clean the air-cooled condenser to sustain capacity. No other consumables or accessories are specified in the datasheet.