Peter Huber Kältemaschinenbau KISS 208B
Peter Huber Kältemaschinenbau KISS 208B — Thermostatic baths.
The KISS 208B is a 7.5 l stainless-steel bath immersion circulator with a non-variable pressure/suction pump, rated 25–200 °C on its own, 20–200 °C with water cooling and −30–200 °C with a refrigerator. Its 1.5–2.1 kW heater and ±0.05 K stability distinguish it from smaller CC-202C/205B siblings; it serves as the temperature-control source for external circuits and bath processes up to 200 °C.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection
| Model | Controller | Interface | Pump connection | Temperature range, °C | Weight, kg | Max delivery flow rate, liter/minute |
|---|---|---|---|---|---|---|
| KISS 208B Current | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 10 | 14 |
| CC-202C | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 22 |
| CC-205B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 27 |
| CC-208B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 10 | 27 |
| CC-212B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 11 | 22 |
| CC-215B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 12 | — |
| CC-220B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 14 | 27 |
| CC-225B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 16 | 27 |
| KISS 202C | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 8 | 14 |
| KISS 205B Badumwälzthermostat mit Edelstahlbad | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 9 | 14 |
| KISS 212B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 11 | 14 |
| KISS 215B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 12 | 14 |
| KISS 220B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 14 | 14 |
| KISS 225B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 16 | 14 |
Full specifications
| Alarm output | optic, acoustic |
| Dimensions (w × D × H) | 290 x 350 x 375 mm |
| Filling volume | 4 l (min) |
| Maximum delivery flow rate | 14 l/min (pressure pump; 0, 25 bar) |
| Pump connection | M16x1 AG male (Option, #19607) (Option, #19607) |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Temperature range | 25...200 °C with cooling unit: -30...200 °C · with water cooling: 20...200 °C |
| Temperature stability mfr | ±0, 05 K |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min (suction pump; 0, 17 bar) |
| Bath opening WxD/BD mfr | 230 x 127 / 150 mm (WxD/depth) |
| Bath volume mfr | 7, 5 l |
| Weight | 10 kg |
| Current consumption | 10 A (max) |
| Power supply | 200-240V 1~/2~ 50/60Hz V; Hz |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Immersion circulator with 7.5 l stainless-steel bath, 230 x 127 mm opening, 150 mm depth; minimum filling 4 l.
- Non-variable pressure and suction pump: 14 l/min at 0.25 bar pressure side, 10.5 l/min at 0.17 bar suction side.
- Within the series, the 208B is the 7.5 l / 2.1 kW step; the CC-202C and CC-205B are smaller-bath siblings, while CC-212B is the larger variant. Choose the 208B when the bath volume and heater headroom fit this middle size.
Thermal behavior and media
- Base range 25–200 °C; adding water cooling extends the low end to 20 °C, and a refrigerator to −30 °C. The cooling method must be selected at order stage because it defines the achievable range.
- Stability is ±0.05 K; heating capacity 1.5–2.1 kW. No digitized performance curves are supplied, so heat-up/cool-down rates cannot be quoted.
- The plastic pump is not resistant to acetone or polyglycols; for freeze protection, mix water with glysantine or ethylene glycol rather than using pure water.
Control, safety and electrical
- KISS controller with OLED display; USB (Device) and RS232 interfaces; optical and acoustic alarm output.
- Adjustable overtemperature protection per DIN 12876, safety class III/FL; autostart after power failure is available and must be assessed for the process.
- Supply 200–240 V, 1~ or 2~, 50/60 Hz, max. current 10 A; branch circuit fuse 10–16 A with external UL 489 protection. IP20, ambient 5–40 °C, Class A Group 1 per EN 55011.
Integration & Operating Notes
The datasheet explicitly forbids combined deviations such as −10% voltage with +3% frequency. Verify the site supply against the 200–240 V, 1~/2~, 50/60 Hz envelope before commissioning, because a supply that looks acceptable on each axis alone may still violate the allowed combination.
The equipment does not carry its own branch-circuit protection; the installer must provide a UL 489 device rated between 10 A and 16 A. Size the circuit for the 10 A maximum current draw and confirm the protective device is coordinated with the 200–240 V single- or two-phase feed.
The unit reaches only 25 °C without auxiliary cooling, 20 °C with water cooling and −30 °C with a refrigerator. Select the cooling configuration at order stage; a unit specified without cooling cannot be pushed below 25 °C later by controller settings alone.
The pump is plastic and not resistant to acetone or polyglycols (manufacturer-dependent). Confirm fluid compatibility before filling; for sub-zero operation use a water/glysantine or water/ethylene glycol mixture rather than pure water to avoid freezing damage.
The autostart function restarts the unit automatically after a power failure. For processes where an unattended restart is hazardous or would spoil a sample, this behavior must be assessed and the process designed around it before relying on the KISS controller in an automated line.
Selection & Integration Q&A
Pick the 208B when the process needs a 7.5 l bath with 1.5–2.1 kW heating and 14 l/min pump flow. The CC-205B is the smaller-bath step; the CC-212B is larger. If your vessel or external loop exceeds the 208B bath volume, step up rather than overfilling.
Provide a 200–240 V, 1~ or 2~, 50/60 Hz feed with a 10–16 A UL 489 branch device, respecting the forbidden voltage/frequency tolerance combinations. On the hydraulic side, the non-variable pump gives 14 l/min at 0.25 bar pressure and 10.5 l/min at 0.17 bar suction; the M16x1 male pump connection is an option (#19607), so confirm fittings before piping.
The M16x1 male pump connection is an option (#19607), not standard, and no cooling unit is included unless specified—water cooling or a refrigerator must be ordered to reach 20 °C or −30 °C. Plan for a freeze-protection glycol mixture and for periodic verification of the adjustable DIN 12876 overtemperature limit; no consumables or cables are listed in the datasheet.