Peter Huber Kältemaschinenbau Minichiller 1200w-H OLÉ Kompakter
Peter Huber Kältemaschinenbau Minichiller 1200w-H OLÉ Kompakter — Recirculating chillers.
Water-cooled recirculating chiller with a pressure/suction pump and a fixed-speed plastic pump head, rated for -20 to 100 °C process temperature. The 'w' suffix denotes water cooling (G1/2 male, 3–6 bar differential), distinguishing it from air-cooled siblings in the Minichiller 1000/1200 series. It serves as a compact secondary cooling source for laboratory and pilot circuits where facility water is available.
Peter Huber Kältemaschinenbau Minichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Minichiller 1200w-H OLÉ Kompakter Current | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H Kompakter | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w Kompakter | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200w-H Kompakter | — | Pilot ONE | III / FL | -20–100 | — | — |
| Minichiller 280 OLÉ Kompakter | 53 | OLÉ | I / NFL | -5–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300 OLÉ Kompakter | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300w-H OLÉ Kompakter | 51 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300w OLÉ Kompakter | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 600-H OLÉ Kompakter | 56 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 600 Kompakter | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 110-120V 1~ 60Hz |
| Minichiller 600 OLÉ Kompakter | 56 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 60Hz |
| Minichiller 600w OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 800 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 800w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
Full specifications
| Maximum delivery flow rate | 24 l/min (pressure pump max; 0, 7 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure- and suction pump |
| Temperature range | -20...100 °C |
| Weight | 37 kg |
| Current consumption | 12 A (max) |
| Fuse rating | 16 A |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Controller | OLÉ |
| Display resolution | 0, 1 K |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 24 л/час (at 15°C, flow -20°C) at 15°C, flow -10°C: 30 л/час · at 15°C, flow 0°C: 36 l/h · at 15°C, flow 15°C: 39 l/h |
| Cooling water differential pressure | 3 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The Minichiller 1200w-H OLÉ is the water-cooled ('w') variant of the 1200-size Minichiller; siblings such as the 1000-H, 1000-H OLÉ and 1000w-H differ in cooling capacity size and/or cooling medium.
- Cooling is via a facility-water circuit with G1/2 male connections; the process circuit uses an M16x1 male pump connection.
- The pump is a pressure-and-suction type with fixed speed (not adjustable), so flow/pressure matching must be done by circuit design, not by the controller.
- Refrigerant is R-290 (A3, H220), a flammable natural refrigerant; the manual requires checking charge versus room volume.
Behavior and media
- Process temperature range is -20 to 100 °C; display resolution is 0.1 K on an OLED display.
- Cooling water consumption depends on process temperature at 15 °C water inlet: 39 l/h at 15 °C, 36 l/h at 0 °C, 30 l/h at -10 °C, 24 l/h at -20 °C (values as printed; note the unit inconsistency in the source table).
- Cooling power falls and required water flow may rise as cooling water temperature increases; size the facility supply with margin.
- The plastic pump is not resistant to acetone or polyglycols; for freeze protection use glysantine or ethylene glycol mixtures.
- Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only compatible water may be used.
Installation, control and operation
- Electrical supply is 208–240 V, 1~/2~, 50/60 Hz; voltage tolerance ±5% only with simultaneous frequency tolerance ±2%, and -5% voltage with +2% frequency is not allowed.
- Max current 12 A, fuse 16 A; single/two-phase units ship with a power cable and country-specific plug that must be specified at ordering.
- Interfaces: USB (Device) and RS232; controller is 'OLÉ' with OLED display.
- Protection IP20, safety class III/FL; must be installed dry, not in an ATEX zone, with a drip tray below and protection of electrical lines below against liquid ingress.
- Do not install valves or quick-release couplings in feed/discharge lines to the application.
Integration & Operating Notes
Below 3 bar the water-cooled refrigeration circuit cannot operate correctly; above 6 bar the cooling water circuit is overpressured. Verify the facility supply at the G1/2 male connections under flow, not only static pressure.
A ±5% voltage tolerance is permitted only with a simultaneous ±2% frequency tolerance; combinations such as -5% voltage with +2% frequency are explicitly not allowed. Confirm the site supply against this combined envelope before commissioning.
Do not install valves or quick-release couplings in the feed/discharge lines from the unit to the application or from the application back to the unit; closing them can overpressure the application, especially glass apparatus.
The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer), so these must not be used as heat transfer fluids. For freeze protection, mix water with glysantine or ethylene glycol rather than using pure water.
The optional Pt100 sensor connection (#10519) displays process temperature but does not control it, and is available only factory-fitted at additional cost. Engineers needing external process temperature control must select a different configuration.
Single/two-phase devices (100 V to 240 V) are supplied with a power cable and country-specific plug that must be specified when ordering; failing to specify it can delay commissioning.
Moving a unit in operation risks burns, freezing and escaping thermal fluid. Shut down and allow temperature equalization before relocation.
Selection & Integration Q&A
Pick the 1200w-H OLÉ when facility cooling water (5–20 °C, 3–6 bar differential) is available and you need the larger 1200-size water-cooled unit. Choose an air-cooled 1000-H variant if no water supply exists; choose the 1000w-H if the smaller 1000-size capacity suffices.
Cooling water at 5–20 °C with 3–6 bar differential pressure at G1/2 male connections, sized with margin because flow demand rises as water temperature increases. Electrical: 208–240 V, 1~/2~, 50/60 Hz within the coupled ±5% voltage / ±2% frequency envelope, 16 A fuse, country-specific plug specified at order.
Dry location, IP20, ambient 5–40 °C, not in an ATEX zone. Install a drip tray below, protect electrical lines below against liquid ingress, and verify R-290 charge versus room volume. Do not move the unit while operating.
The country-specific plug must be specified at ordering; the Pt100 display option (#10519) is factory-fitted at extra cost. Plan anti-corrosion dosing if salts are added to the water circuit, and glysantine or ethylene glycol for freeze protection. No maintenance interval is stated in the supplied digests.