Peter Huber Kältemaschinenbau Minichiller 800 OLÉ Kompakter
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Minichiller 800 OLÉ Kompakter

Peter Huber Kältemaschinenbau Minichiller 800 OLÉ Kompakter — Recirculating chillers.

The Minichiller 800 OLÉ Kompakter is an air-cooled recirculating chiller with a pressure/suction pump and OLÉ controller, rated -20...40 °C with an 80 °C return-line limit. Its 0.3 kW at -20 °C and 24 l/min at 0.7 bar distinguish it from larger 1000-series siblings. It serves closed-loop cooling of laboratory and pilot processes.

Peter Huber Kältemaschinenbau Minichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Minichiller 800 OLÉ Kompakter Current —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000w-H Kompakter —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Minichiller 1000w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1000w Kompakter —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Minichiller 1200-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1200 OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 1200w-H Kompakter —Pilot ONEIII / FL-20–100——
Minichiller 1200w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 280 OLÉ Kompakter 53OLÉI / NFL-5–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300 OLÉ Kompakter 53OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300w-H OLÉ Kompakter 51OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 300w OLÉ Kompakter 53OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Minichiller 600-H OLÉ Kompakter 56OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 600 Kompakter —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D110-120V 1~ 60Hz
Minichiller 600 OLÉ Kompakter 56OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 60Hz
Minichiller 600w OLÉ Kompakter —OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Minichiller 800w-H OLÉ Kompakter —OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz

Full specifications

Core Performance & Functional Specifications
Maximum delivery flow rate 24 l/min (pressure pump; 0, 7 bar)
Pump connection M16x1 AG male
Pump speed adjustable no
Pump type Pressure- and suction pump
Temperature range -20...40 °C
Dimensions, Weight & Mechanical Design
Weight 37 kg
Electrical, Power & Drive Specifications
Current consumption 4 A (max)
Power supply 208-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller OLÉ
Display resolution 0, 1 K
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Maximum return flow temperature 80 °C
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Process Limits
Return-line temperature may reach 80 °C

Although the operating range is -20...40 °C, the return line tolerates up to 80 °C. Exceeding this limit risks damage to the evaporator and pump; ensure the application cannot return hotter fluid, especially during startup or batch dumps.

Minichiller 800 OLÉ Technical Data, Specifications
Cooling & Media
Plastic pump incompatible with acetone and polyglycols

The wetted plastic pump is not resistant to acetone or polyglycols. Using these as heat transfer media will destroy the pump and contaminate the circuit; select an alternative fluid or a different pump material.

Minichiller 800 OLÉ Technical Data, Specifications
Cooling & Media
Freeze protection required for sub-zero setpoints

For operation below 0 °C, mix water with glysantine or ethylene glycol; pure water will freeze in the evaporator and damage the refrigeration circuit. Confirm the mixture concentration against the lowest setpoint.

Minichiller 800 OLÉ Technical Data, Specifications
Mounting
Drip tray and protected cabling required below unit

A condensation water / thermal fluid drip tray must be installed below the chiller, and electrical lines directly below must be protected against liquid ingress, because condensation can form at the cooling water connections.

Minichiller Operation Manual, Section 4 Installation
Mounting
Not permitted in ATEX zones

The unit must not be installed or started in an ATEX zone; operation there can cause explosion. Additionally, the R-290 (A3, H220) refrigerant charge must be checked against room volume to avoid exceeding the limit value per m³ of ambient air.

Minichiller Operation Manual, Section 4 Installation
Electrical
Voltage and frequency tolerances must not combine unfavorably

Supply tolerances are ±5 % voltage and ±2 % frequency, but a -5 % voltage combined with +2 % frequency is not allowed. Verify the site supply before connection to avoid motor and pump damage.

Minichiller 800 OLÉ Technical Data, Specifications
Electrical
Cable supplied without plug for three-phase below 63 A

Three-phase devices with current consumption less than 63 A are supplied with cable but without a plug; the installer must provide the appropriate plug and connection. This model draws max 4 A at 208-240 V 1~/2~.

Minichiller 800 OLÉ Technical Data, Specifications
Process Limits
No valves or quick-release couplings in application lines

Do not install valves or quick-release couplings in the feed/discharge lines between the chiller and the application; closing them can overpressure the circuit and damage glass apparatus or the pump.

Minichiller Operation Manual, Section 5 Commissioning
Maintenance
Anti-corrosion agents mandatory if salts are added

If salts such as chlorides or bromide are added to the water circuit, anti-corrosion agents are mandatory to prevent corrosion damage. Only drinking water or filtered river/sea water may be used; sea water must not be used.

Minichiller Operation Manual, Section 6 Service and Maintenance
Maintenance
Minimum filling capacity 2.8 l with 2.2 l expansion tank

The minimum filling capacity is 2.8 l and the expansion tank is 2.2 l; the external loop volume must be considered when filling to ensure enough heat transfer fluid and expansion volume for safe operation.

Minichiller 800 OLÉ Technical Data, Specifications

Selection & Integration Q&A

When should I choose the Minichiller 800 OLÉ Kompakter instead of a 1000-series sibling?

Choose the 800 OLÉ when the heat load at the lowest setpoint is modest (0.3 kW at -20 °C) and space/weight (37 kg) matter. The 1000-series siblings offer larger cooling capacity; select them when the load exceeds the 800's capacity at your setpoint.

What adjacent system parts and connections are required for this chiller?

The pump connections are M16x1 male; hose connector NW12 and sleeve nuts are included. Provide 208-240 V 1~/2~ 50/60 Hz supply with min 10 A fuse; three-phase below 63 A comes with cable but no plug. Do not add valves or quick-release couplings in the application lines.

What mounting and environment limits apply to the Minichiller 800 OLÉ Kompakter?

Install between 5 and 40 °C ambient, not in an ATEX zone, with a drip tray below and protected cabling. Check R-290 charge against room volume. Air-cooled natural draft requires adequate airflow; capacity falls as ambient rises.

What maintenance and consumables are needed, and what is not included in base delivery?

Use drinking or filtered river water only; sea water is prohibited. If salts are added, anti-corrosion agents are mandatory. For sub-zero operation, mix water with glysantine or ethylene glycol. Minimum filling 2.8 l; expansion tank 2.2 l. A plug for three-phase supply is not included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Minichiller Operation Manual
Manual ·2,9 МБ ·84 pp.
View PDF
PDF
Minichiller 800 OLÉ Technical Data
Datasheet ·100,7 КБ ·2 pp.
View PDF