Peter Huber Kältemaschinenbau Unistat P425w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat P425w Dynamisches

Peter Huber Kältemaschinenbau Unistat P425w Dynamisches — Temperature control units with Noise level 61 dB.

The Unistat P425w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -40 to 250 °C at 0.01 K display resolution. Its 2 kW heating capacity and 97 l/min pump delivery suit externally closed circuits with high pressure drop, while cooling power falls to 0.1 kW at -40 °C. It serves as a precise heating/cooling source for reactor jackets and laboratory process loops.

Noise level
61 dB

Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionPower supplyPump connectionPump typeTemperature range, °C
Unistat P425w Dynamisches Current 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Grande Fleur Dynamisches ——————
Grande Fleur w Dynamisches —0, 01208-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -40–200
Grande Fleur w-eo Dynamisches ——————
Petite Fleur-eo Dynamisches 580, 01 K220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Petite Fleur w Dynamisches —0, 01220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Unistat 410w Dynamisches 610, 01 K220-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -45–250
Unistat 425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat 430w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P405w Dynamisches —0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st-45–250
Unistat P425 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430w Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250

Full specifications

Core Performance & Functional Specifications
Noise level 61 dB(A)
Control output POKO ONE
Heating capacity 2 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -40...250 °C
Dimensions, Weight & Mechanical Design
Weight 178 kg
Electrical, Power & Drive Specifications
Current consumption 10 A
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 150 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working behavior

Installation and control

Integration & Operating Notes

Electrical
Size supply and RCD for >3.5 mA leakage current

The unit draws 10 A on a 400 V 3~ 50 Hz supply protected by 3x16 A, but leakage current exceeds 3.5 mA. Protective earth continuity and residual-current device selection must be engineered for this leakage level; voltage tolerance is ±5% only with ±2% frequency, and -5% voltage with +2% frequency is not permitted.

Unistat P425w Technical Data, page 2
Cooling & Media
Cooling water must stay within 1–6 bar and 5–20 °C intake

The cooling water circuit requires 150 l/h at 15 °C with 1 bar differential pressure and a maximum of 6 bar; intake water must remain between 5 °C and 20 °C. Wetted materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used. Insulate internal cooling-water pipes when water temperature falls below the condensation point.

Unistat P425w Technical Data, page 2
Mounting
Transport and store upright; acclimate 2 hours before power-on

Horizontal transport damages the compressor, and only an emptied unit may be moved. Casters are not for transport and carry only 25% of total mass symmetrically; use an industrial truck. Leveling feet must be activated and parking brakes engaged before commissioning. Allow about 2 hours acclimation when moving between cold and warm environments before switching on.

Unistat Operation Manual, Commissioning, page 22 and page 29
Process Limits
Expansion vessel must stay unlocked for closed applications

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage to the application. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. The unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet.

Unistat Operation Manual, Commissioning, page 33 and page 15
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles, but this mode requires the compressor to be set to Permanently ON. Power Failure Auto resumes temperature control automatically after an outage if it was active; choose the power-interruption behavior according to whether unattended restart is acceptable.

Unistat Operation Manual, Setup Mode, page 19–20
Maintenance
Shut off cooling water even for overnight downtime

Shut off the cooling water supply even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning contractors.

Unistat Operation Manual, Commissioning and Maintenance, page 29

Selection & Integration Q&A

When should I choose the P425w Dynamisches rather than a sibling Grande Fleur variant?

Choose this P variant when the external circuit has high pressure drop and needs up to 97 l/min at 3 bar; the w suffix means water cooling, so a facility water supply at 1–6 bar and 5–20 °C is required. Grande Fleur siblings differ in configuration and should be evaluated against their own datasheets.

What adjacent system parts and connections must be prepared for this unit?

The pump connection is M30x1.5 male using bore shape Y (60°) per DIN 3863 with ball socket and sleeve nut per DIN 3870; cooling water is G1/2 male. A 400 V 3~ 50 Hz supply with 3x16 A fuse is required, and an external Pt100 sensor can be connected. Matching hoses and fittings are not included and must be ordered.

What mounting and duty-cycle limits apply from the manual?

Install upright only, never in an ATEX zone, with ambient 5–40 °C and IP20 protection. Activate leveling feet and parking brakes; casters are not for transport. Allow 2 hours acclimation after temperature changes. For closed applications keep the expansion vessel unlocked; for open bath applications lock it.

What maintenance schedule and consumables are required beyond base delivery?

Periodically check hoses and connections for leaks and quality and replace as needed; shut off cooling water even overnight. Clean liquefier fins and the hat-type strainer/dirt trap; inspect and replace heat transfer fluid and rinse the circuit. Refrigerant-circuit work requires approved contractors. Hoses, fittings, fluid and a drip tray are not included.