Peter Huber Kältemaschinenbau Grande Fleur w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Grande Fleur w Dynamisches

Peter Huber Kältemaschinenbau Grande Fleur w Dynamisches — Temperature control units.

The Grande Fleur w Dynamisches is a water-cooled temperature control unit in the Unistate Bis 45C series, rated for -40 to 200 °C with a variable-speed pressure pump and Pilot ONE controller. Its water-cooled condenser using natural refrigerant R-1270 (GWP 0) distinguishes it from air-cooled siblings, suiting indoor installations with a 5–20 °C cooling water supply.

Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionPower supplyPump connectionPump typeTemperature range, °C
Grande Fleur w Dynamisches Current —0, 01208-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -40–200
Grande Fleur Dynamisches ——————
Grande Fleur w-eo Dynamisches ——————
Petite Fleur-eo Dynamisches 580, 01 K220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Petite Fleur w Dynamisches —0, 01220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Unistat 410w Dynamisches 610, 01 K220-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -45–250
Unistat 425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat 430w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P405w Dynamisches —0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st-45–250
Unistat P425 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430w Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Expansion volume 3 l
Pump connection M24x1, 5 AG male
Pump speed adjustable Yes
Pump type Pressure pump, variable speed
External sensor connection Pt100
Temperature range -40...200 °C
Electrical, Power & Drive Specifications
Power supply 208-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-1270 (A3, H220) (1st stage)
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 36 л/час (at 15°C, flow -20°C)
at 15°C, flow 0°C: 49 l/h · at 15°C, flow 15°C: 51 l/h
Cooling water differential pressure 2 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and operation

Integration & Operating Notes

Cooling & Media
Cooling water pressure must stay within 2–6 bar differential

Site water supply must provide at least 2 bar differential pressure and must never exceed 6 bar at the unit inlet. Exceeding the maximum risks damage to the cooling circuit; insufficient differential pressure prevents adequate heat rejection and can trip the unit.

Datasheet, Specifications, page 2
Cooling & Media
Cooling water intake temperature limited to 5–20 °C

The unit is designed for cooling water between 5 °C and 20 °C. If the water temperature can fall below the ambient dew point, pipes must be insulated to prevent condensation on the cooling water circuit.

Datasheet, Specifications, page 2
Electrical
Voltage and frequency tolerances must not oppose each other

Supply voltage tolerance is ±5% and frequency tolerance ±2%, but combinations such as -5% voltage with +2% frequency are not permitted. Verify site power quality against this envelope before commissioning.

Datasheet, Specifications, page 2
Electrical
Country-specific plug must be specified at order

Single/two-phase devices (100–240 V) are delivered with a power cable and country-specific plug. The plug type must be stated when ordering; otherwise the unit may arrive without a suitable connector for the site.

Datasheet, Specifications, page 2
Mounting
Indoor dry location required; IP20 and 5–40 °C ambient

The unit is rated IP20 and must operate between 5 °C and 40 °C ambient temperature. It must be installed in a dry, protected indoor location; moisture or dust ingress can damage electronics and invalidate the protection rating.

Datasheet, Specifications, page 2
Process Limits
Heat transfer fluid viscosity must not exceed 50 mm²/s

At the lowest operating temperature, the heat transfer fluid's kinematic viscosity must remain below 50 mm²/s. This limits fluid selection for -40 °C duty and affects pump performance and cooling capacity.

Datasheet, Specifications, page 2
Process Limits
Cooling capacity drops sharply at low temperatures

With ethanol, cooling capacity falls from 0.6 kW at 0 °C to 0.04 kW at -40 °C. Process heat loads must be matched to the actual operating temperature; a load acceptable at mid-range may be unachievable at the low end.

Datasheet, Overview, page 1
Maintenance
Use only cooling water compatible with wetted materials

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Aggressive or incompatible water can corrode or degrade these materials; verify water quality before connection.

Datasheet, Specifications, page 2
Process Limits
Refrigerant R-1270 is flammable (A3) with GWP 0

The unit uses R-1270 (A3, H220) in the first stage. Safety measures for flammable refrigerants apply; handling and service must follow local regulations for A3 refrigerants.

Datasheet, Overview, page 1

Selection & Integration Q&A

When should I choose the Grande Fleur w Dynamisches over the air-cooled Grande Fleur Dynamisches or the Petite Fleur Dynamisches?

Choose the 'w' water-cooled variant when a 5–20 °C cooling water supply with 2–6 bar differential is available and air-cooled heat rejection is impractical. The Petite Fleur is smaller and rated 58 dB(A); select it only if its capacity fits. The 'eo' variants differ in configuration and should be compared on their own datasheets.

What cooling water supply and plumbing must be provided for this unit?

Provide cooling water at 5–20 °C with minimum 2 bar differential and maximum 6 bar pressure. Connect via G1/2 male fitting. Insulate pipes if water can fall below dew point. Water must be compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.

What are the ambient and electrical installation limits for this unit?

Install indoors, dry, IP20, ambient 5–40 °C. Supply 208–240 V 1~/2~ 50/60 Hz with ±5% voltage and ±2% frequency, not opposing. Specify country-specific plug at order. Pump connections require DIN 3863 ball socket fittings and DIN 3870 sleeve nuts.

What maintenance and accessories are required beyond the base delivery?

Base delivery includes power cable with country-specific plug only if specified. Order DIN 3863 ball socket fittings and DIN 3870 sleeve nuts for pump connections, plus G1/2 hoses for cooling water. Heat transfer fluid must have viscosity ≤50 mm²/s at lowest operating temperature. No maintenance interval is stated in the provided datasheet.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
Grande Fleur
Datasheet ·118,4 КБ ·2 pp.
View PDF