Peter Huber Kältemaschinenbau Unistat P430 Dynamisches
Peter Huber Kältemaschinenbau Unistat P430 Dynamisches — Temperature control units with Noise level 65 dB.
The Unistat P430 Dynamisches is a free-standing, air-cooled temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -40 to 250 °C with 4 kW heating. It is the dynamic-pressure variant (P) of the Unistat 4xx series, intended for externally closed circuits with high pressure drop, delivering 30 l/min at 2.5 bar.
Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P430 Dynamisches Current | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Grande Fleur Dynamisches | — | — | — | — | — | — |
| Grande Fleur w Dynamisches | — | 0, 01 | 208-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -40–200 |
| Grande Fleur w-eo Dynamisches | — | — | — | — | — | — |
| Petite Fleur-eo Dynamisches | 58 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Petite Fleur w Dynamisches | — | 0, 01 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Unistat 410w Dynamisches | 61 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -45–250 |
| Unistat 425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat 430w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P405w Dynamisches | — | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | -45–250 |
| Unistat P425 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430w Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
Full specifications
| Noise level | 65 dB(A) |
| Control output | POKO ONE |
| Filling volume | 4 l |
| Heating capacity | 4 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -40...250 °C |
| Weight | 197 kg |
| Current consumption | 15 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing Unistat 4xx platform (460×570×1629 mm, 197 kg) with IP20 protection, designed for dry indoor installation.
- The 'P' suffix denotes the high-pressure-drop pump variant; the 'Dynamisches' designation indicates suitability for dynamic temperature changes in externally closed applications.
- Sibling models in the series (Grande Fleur variants) differ in cooling/refrigerant configuration; this P430 uses R-1270 (A3, H220) refrigerant with 0.75 kg charge.
- Pump: magnetically coupled, stainless steel, speed-adjustable, with M30×1.5 male connections (Y-bore, ball-socket per DIN 3863/3870).
Thermal and hydraulic behavior
- Working range -40…250 °C; heating capacity 4 kW; cooling capacity falls from 3.5 kW at 0 °C to 0.15 kW at -40 °C with ethanol, so low-temperature duty must be sized against the curve.
- Pump delivers 97 l/min maximum but only 30 l/min at 2.5 bar; high-pressure-drop loops require checking the pump curve against the external circuit resistance.
- Minimum filling capacity 4 l; expansion tank 3.7 l — external loop volume must be checked against these limits.
- Control via Pilot ONE with POKO ONE output, 0.01 K display resolution, internal/process control modes and external Pt100 sensor input.
Installation and operation
- Requires 400 V 3~ 50 Hz supply with 3×16 A fuse; three-phase units below 63 A are delivered with cable but without plug.
- Leakage current >3.5 mA — RCD/protective earthing must be engineered accordingly.
- Ambient 5…40 °C; cooling capacity may drop above 20 °C room temperature.
- Adjustable overtemperature protection per DIN 12876, safety class III/FL; flammable refrigerant R-1270 requires site precautions.
- Interfaces: Ethernet, USB (Host and Device), RS232.
Integration & Operating Notes
The unit is factory-configured for three-phase 400 V 50 Hz and requires a 3×16 A fuse. Three-phase devices below 63 A are delivered with cable but without plug, so the customer must provide and fit a suitable plug. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations (e.g. -5% voltage with +2% frequency) are not allowed.
The datasheet states leakage current >3.5 mA. Site electrical design must account for this with a suitable residual current device or protective earthing arrangement; standard 30 mA RCDs may not be appropriate without further assessment.
The manual explicitly prohibits installation or start-up within an ATEX zone; explosion protection requires a separate Ex px cabinet and only the model/serial number stated on the Ex px cabinet rating plate is suitable. Site selection must exclude hazardous areas.
Only transport the unit upright and emptied; horizontal transport damages the compressor. Casters are not for transport and carry only 25% of total mass symmetrically — use an industrial truck. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults.
In Process Safety mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON to keep cooling active at the OT trigger. Configure accordingly for exothermic process protection.
The manual directs shutting off the cooling water supply even for shorter downtimes such as overnight, to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.
The responsible body must install a condensation water / heat transfer fluid drip tray below the temperature control unit; national legislation may prescribe it. Installation planning must include the drip tray and verification of local regulations.
Selection & Integration Q&A
Choose the P430 when your externally closed circuit has high pressure drop and you need dynamic temperature changes; the 'P' pump delivers 30 l/min at 2.5 bar. Sibling Grande Fleur variants differ in cooling/refrigerant configuration — select those only if their specific cooling or media configuration matches your process, not as a substitute for high-pressure capability.
Provide a 400 V 3~ 50 Hz supply with 3×16 A fuse and a suitable plug (cable supplied, plug not). Pump connections are M30×1.5 male with Y-bore and ball-socket per DIN 3863/3870 — order matching hoses and fittings. Ensure RCD/protective earthing handles >3.5 mA leakage. External Pt100 sensor and Ethernet/USB/RS232 interfaces are available for control integration.
Install indoors, dry, IP20, ambient 5…40 °C, never in an ATEX zone. Transport upright and empty; use an industrial truck, not casters. Allow 2 h acclimation after temperature changes. Leveling feet must be activated and parking brakes engaged. Cooling capacity drops above 20 °C ambient and at low temperatures; size against the curve.
Base delivery excludes plug, mating hoses/fittings, drip tray, and external Pt100 sensor. Periodically check hoses and connections for leaks and quality and replace as needed; shut off cooling water overnight. Clean liquefier fins and hat-type strainer/dirt trap; inspect mechanical seal and plug contacts. Heat transfer fluid must be inspected/replaced per manual; refrigerant work only by approved contractors.