Peter Huber Kältemaschinenbau Unistat P430w Dynamisches
Peter Huber Kältemaschinenbau Unistat P430w Dynamisches — Temperature control units with Noise level 65 dB.
The Unistat P430w Dynamisches is a water-cooled, high-pressure-drop circulating temperature control unit with a magnetically coupled stainless-steel pump, covering -40 to 250 °C with 4 kW heating. Its 'P' designation targets externally closed loops with high pressure drop, where the pump delivers 30 l/min at 2.5 bar. It serves precise temperature control of reactors, calorimeters and pilot-scale process equipment.
Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P430w Dynamisches Current | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Grande Fleur Dynamisches | — | — | — | — | — | — |
| Grande Fleur w Dynamisches | — | 0, 01 | 208-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -40–200 |
| Grande Fleur w-eo Dynamisches | — | — | — | — | — | — |
| Petite Fleur-eo Dynamisches | 58 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Petite Fleur w Dynamisches | — | 0, 01 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Unistat 410w Dynamisches | 61 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -45–250 |
| Unistat 425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat 430w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P405w Dynamisches | — | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | -45–250 |
| Unistat P425 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
Full specifications
| Noise level | 65 dB(A) |
| Alarm output | optic, acoustic, relay |
| Control output | POKO ONE |
| Filling volume | 4 l (min.) |
| Heating capacity | 4 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -40...250 °C |
| Weight | 182 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The P430w is a free-standing, water-cooled Unistat with a magnetically coupled stainless-steel pump and a Pilot ONE controller; the 'P' suffix marks the high-pressure-drop variant, while 'w' denotes water cooling (air-cooled and CO2/GL variants exist in the series).
- Sibling models in the Unistat series (Grande Fleur variants) differ in configuration; the P430w is the modification specified for loops whose pressure drop would throttle a standard pump.
- Heating capacity is 4 kW; the working range is -40 to 250 °C with 0.01 K display resolution and Pt100 external sensor input.
Behavior under load and media
- Cooling power falls steeply at low temperature with ethanol: 3.5 kW at 0 °C but only 0.15 kW at -40 °C, so near-limit operation must be planned around this curve.
- Pump delivery is 97 l/min at zero head and 30 l/min at 2.5 bar; high-impedance circuits must be matched against this curve rather than the free-delivery figure.
- Heat transfer fluid kinematic viscosity must not exceed 50 mm²/s; the FL safety class requires a fire point of at least 65 °C, and ethanol use is permitted only under supervision.
Installation, control and operation
- The unit requires a 400 V 3~ 50 Hz supply with a 3x16 A fuse, is supplied with cable but without a plug, and has a leakage current above 3.5 mA; voltage and frequency tolerances of ±5% and ±2% must not combine unfavorably.
- Cooling water: G1/2 male connections, inlet temperature 5–20 °C, differential pressure at least 1 bar and not more than 6 bar; the circuit contains copper, 1.4401 stainless steel, MS, PA, PPE, PTFE and EPDM, so water chemistry must be compatible.
- The Pilot ONE controller provides Ethernet, USB host/device and RS232 interfaces, optic/acoustic/relay alarm output, adjustable pump speed, and setup of overtemperature protection, Delta T limiter and Process Safety mode.
- IP20 rating and 5–40 °C ambient limit restrict siting to dry, temperate indoor locations; EMC class A group 1 is intended for industrial environments.
Integration & Operating Notes
The electrical installation must be sized for a 400 V 3~ 50 Hz supply protected at 3x16 A. Voltage tolerance is ±5% and frequency ±2%, but the two must not combine in the unfavorable direction; verify site power quality before connection.
Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must supply and fit a suitable plug. Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must be verified by the installing electrician.
The cooling water circuit requires at least 1 bar differential pressure and must not exceed 6 bar, with inlet temperature between 5 and 20 °C. Out-of-range water temperature reduces cooling power and risks condensation, so pipes must be insulated where condensation can occur.
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s; selecting a fluid too viscous at low temperature will reduce flow and load the pump. For FL-rated fluid the fire point must be at least 65 °C, and ethanol is permitted only under supervised operation.
The unit must not be installed or started within an ATEX zone; explosion protection requires a separate Ex px cabinet and only the model/serial number on the Ex px rating plate is suitable. The refrigerant R-1270 (A3, H220) is flammable, so site ventilation and ignition-source control must be assessed.
Only transport the unit upright and emptied; horizontal transport damages the compressor. When moving between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults. Leveling feet must be activated and casters are not for transport.
In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles, but this mode requires the compressor to be set to Permanently ON. Configure the power-failure response (Off/Standby or ON) according to whether unattended restart is acceptable.
The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Hoses and connections must be checked periodically for leaks and quality and replaced as needed; refrigerant-circuit work requires approved refrigeration contractors.
Selection & Integration Q&A
Choose the P430w when the external loop has high pressure drop: its pump delivers 30 l/min at 2.5 bar, whereas a standard pump would be throttled. For low-impedance loops or air-cooled sites, a non-P or non-w variant of the series is the better fit.
Provide cooling water at 5–20 °C with 1–6 bar differential pressure through G1/2 male connections, and a 400 V 3~ 50 Hz supply fused at 3x16 A with a site-fitted plug. Verify RCD selection because leakage current exceeds 3.5 mA.
Install indoors at 5–40 °C ambient, IP20, outside any ATEX zone, upright and level with feet activated. Allow 2 h acclimation after cold/warm moves. The refrigerant is flammable R-1270, so ventilation and ignition-source control are required.
The three-phase plug, mating hoses/fittings for M30x1.5 pump connections and G1/2 cooling water, heat transfer fluid, and a drip tray are not included. Periodic hose inspection, cooling water shutoff at downtime, and refrigerant work by approved contractors are required.