BINDER BF 56
BINDER ·Made in Germany · In Active Production

BINDER BF 56

BINDER BF 56 — Incubators with Temperature 7…100 °C.

The BINDER BF 56 is a forced-convection incubator with a 59 L interior, adjustable fan speed in 10 % steps, and a working range of +7 to 100 °C. Within the BF series it is the smallest chamber, distinguished by the tightest temperature uniformity (0.2 K) and fluctuation (0.1 K) for culture incubation at 37 °C. It serves as a benchtop incubation stage where airflow and temperature homogeneity are critical.

Temperature
7…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER BF Series — Model Comparison & Selection

Model Temperature, °CHeating up time to 100 c, minHeight, cmInterior volume, LLoad per rack, kgPermitted load, kg
BF 56 Current 7–1003562.5593070
BF 115 8–1006572.511430150
BF 260 7–1003084.525740270
BF 400 5–10012102.54003590
BF 720 10–1004515973445315

Full specifications

Core Performance & Functional Specifications
Temperature 7...100 °C
Heating-up time to 100 °C 35 min
Interior volume 59 L (BF056-230V / BF056UL-120V)
Load per rack 30 kg (BF056-230V / BF056UL-120V)
Nominal power 0.4 kW
Permitted load 70 kg (BF056-230V / BF056UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Back wall clearance 160 mm (BF056-230V / BF056UL-120V)
Side wall clearance 100 mm (BF056-230V / BF056UL-120V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 37°C mfr 3 min
Dimensions, Weight & Mechanical Design
Depth 565 mm (BF056-230V / BF056UL-120V)
Height 625 mm (BF056-230V / BF056UL-120V)
Width 560 mm (BF056-230V / BF056UL-120V)
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in operation

Installation, connection and control

Integration & Operating Notes

Mounting
Maintain 160 mm rear and 100 mm side clearance

The chamber requires unobstructed airflow for heat dissipation and safe operation. Installing it closer than the specified wall clearances or in an unventilated recess will cause overheating and may invalidate performance specifications. Ensure the surface is non-flammable and the room is indoors.

Factory Specification Sheet; Operating Manual, Installation location p.20
Electrical
External 6.3 A time-lag fuse required upstream

The chamber does not include an internal fuse for the 230 V variant; a 6.3 A time-lag (T) 5×20 mm / 250 V fuse must be provided in the supply circuit. Verify that the supply voltage matches the type plate within ±10 % and that the circuit can deliver the 0.4 kW nominal power without excessive voltage drop.

Operating Manual, Electrical connection data p.18
Control & I/O
Set safety controller 2–5 °C above working set-point

The safety controller acts as an absolute maximum temperature limit. After any change to the temperature set-point, re-adjust the safety controller set-point to 2–5 °C above the new value. The permissible range is 10 °C to 100 °C; setting it too close may cause nuisance trips, too high may fail to protect samples.

Operating Manual, Set-point entry for temperature p.20; Setting the safety controller set-point p.22
Process Limits
Load usable volume to no more than half

Samples must remain within the usable volume with at least 10 % of interior dimensions as clearance from walls. Overloading or placing samples too close to the walls obstructs airflow, leading to temperature stratification and non-uniform results. The maximum load per rack is 30 kg and total permitted load is 70 kg.

Operating Manual, Placing the charging material p.19; Factory Specification Sheet
Maintenance
Annual maintenance and electrical safety test mandatory

The chamber must be maintained at least once a year by licensed electricians or BINDER-authorized personnel. An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required, as well as a test before initial startup and after any repair. Warranty is void if maintenance is performed by unauthorized personnel.

Operating Manual, Maintenance intervals p.69
Cooling & Media
Hot-air disinfection at 100 °C for BF series

The BF 56 can be hot-air disinfected at 100 °C / 212 °F. All flammable goods must be removed before disinfection. Racks are removable and can be autoclaved. Use neutral cleaning agent (art. no. 1002-0016) or disinfectant spray (art. no. 1002-0022); do not apply neutral cleaner to zinc-coated hinge parts or rear wall.

Operating Manual, Cleaning and decontamination p.23

Selection & Integration Q&A

When should I choose the BF 56 over the larger BF 115 or BF 260?

Choose the BF 56 for small workloads up to 59 L where bench space is limited and precise temperature uniformity (0.2 K) is needed. The BF 115 offers +8 to 100 °C and larger volume; the BF 260 has the same +7 to 100 °C range but more capacity. Select the BF 56 when sample throughput fits its 59 L and 70 kg load limit.

What are the power supply and fusing requirements for the BF 56?

The BF056-230V requires 230 V 1~ 50/60 Hz, 0.4 kW, with an external 6.3 A time-lag (T) 5×20 mm / 250 V fuse. The BF056UL-120V uses 120 V 1~ 60 Hz with a NEMA 5-15 plug. Supply voltage must be within ±10 % of nominal. The chamber has no internal fuse for the 230 V variant; the fuse must be provided in the installation.

What ambient conditions and clearances are required for installing the BF 56?

Install indoors on a non-flammable surface with ambient temperature +18 to +40 °C, max 70 % r.h. non-condensing, and altitude up to 2000 m. Maintain 160 mm rear clearance and 100 mm side clearance. Do not install in unventilated recesses or explosive atmospheres. The chamber is IP 20 and must not be exposed to water.

What maintenance and accessories are needed beyond the base delivery?

Annual maintenance by authorized personnel and annual electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required. Recommended consumables: neutral cleaning agent art. no. 1002-0016 and disinfectant spray art. no. 1002-0022. Optional accessories include APT-COM 4 software, Pt100 object temperature sensor, analog output, and Ethernet interface (not standard on BF).

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model BF 56 en
Datasheet ·706,9 КБ ·7 pp.
View PDF
PDF
QuickStartGuide BD BF ED FD FED E31 03 2023 dt en fr
Manual ·1,7 МБ ·35 pp.
View PDF
PDF
BD BF ED FD FED (E3.1) 07 2025 en
Manual ·18,7 МБ ·122 pp.
View PDF