BINDER KB 720
BINDER KB 720 — Incubators with Temperature 0…70 °C.
The BINDER KB 720 is a 727 L cooling incubator (0–70 °C) with 1.5 kW nominal power and 640 W max heat compensation, distinguished within the KB series by its large interior and 200 kg permitted load. It provides 0.1 K fluctuation and 0.2 K uniformity for temperature-controlled storage or incubation of bulk samples.
Direct drop-in replacements
BINDER KB Series — Model Comparison & Selection
Full specifications
| Temperature | 0...70 °C |
| Interior volume | 727 L |
| Load per rack | 40 kg |
| Max. heat compensation | 640 W |
| Nominal power | 1.5 kW |
| Permitted load | 200 kg (KB720-230V / KB720UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.2 K |
| Back wall clearance | 100 mm (KB720-230V / KB720UL-120V) |
| Side wall clearance | 180 mm (KB720-230V / KB720UL-120V) |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 37°C mfr | 3 min |
| Depth | 900 mm (KB720-230V / KB720UL-120V) |
| Height | 1951 mm (KB720-230V / KB720UL-120V) |
| Interior width | 1011 mm (KB720-230V / KB720UL-120V) |
| Width | 1285 mm (KB720-230V / KB720UL-120V) |
| Net weight of the unit (empty) mfr | 285 kg |
| Power frequency | 50 Hz |
| Rated voltage | 220...240 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and positioning
- The KB 720 is the largest standard model in the KB series (siblings: KB 130, KB 260, KB 470, KB 1060), offering 727 L interior volume and a 200 kg permitted load (40 kg per rack).
- Exterior dimensions (W×H×D): 1,285 × 1,951 × 900 mm; net weight 285 kg. It is a floor-standing, single-phase (220–240 V, 50 Hz) unit with a 16 A fuse and 1.5 kW nominal power.
- Unlike smaller KB models, the KB 720 accommodates high-volume loads and multiple racks, making it suitable for scale-up or bulk incubation.
Behavior under load
- Temperature control range is 0–70 °C with 0.1 K fluctuation and 0.2 K uniformity, ensuring stable conditions across the large interior.
- Maximum heat compensation of 640 W allows the unit to counteract heat generated by loads or accessories (e.g., LED lighting) without exceeding temperature limits.
- Recovery time after a 30 s door opening at 37 °C is 3 minutes, minimizing process interruption during access.
Installation and operation
- Requires 100 mm rear and 180 mm side wall clearance for ventilation; do not install in unventilated recesses.
- The unit is designed for indoor use; ambient conditions and electrical supply must match the specifications. No additional control interfaces are detailed in the provided documents.
Integration & Operating Notes
Insufficient clearance restricts heat dissipation and can cause overheating or reduced performance. Ensure the installation location allows these minimum distances to walls or adjacent equipment, and avoid recessed spaces without ventilation.
The unit draws up to 1.5 kW at 220–240 V, 50 Hz, and requires a 16 A fuse. Verify the branch circuit can handle the nominal current and that the supply voltage stays within the specified range to prevent damage or tripping.
Exceeding the permitted load may deform racks or affect temperature uniformity. Distribute samples evenly and use appropriate racks rated for the load. The maximum heat compensation of 640 W must also be considered when loading heat-generating items.
Selection & Integration Q&A
Choose the KB 720 when you need high capacity (727 L) and can handle a 200 kg load, such as bulk incubation or large sample batches. Smaller models like KB 470 (470 L) or KB 260 (260 L) are better for limited space or lower throughput, but offer the same temperature range and control precision.
Annual electrical safety inspections are recommended. Racks and shelves are not included and must be ordered separately; the unit supports up to 40 kg per rack. No other consumables are specified in the provided documents. Regular cleaning and temperature calibration may be required per laboratory standards.