The CC-215B is a 15-litre stainless-steel bath circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated for 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. Its 1.5–2.1 kW heater and ±0.02 K stability distinguish it within the CC-2xxB series by volume and heating power. It serves as a temperature-control source for external circuits or direct bath immersion in laboratory processes.

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection

Model ControllerInterfacePump connectionTemperature range, °CWeight, kgMax delivery flow rate, liter/minute
CC-215B Current Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–20012—
CC-202C Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200922
CC-205B Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200927
CC-208B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–2001027
CC-212B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001122
CC-220B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001427
CC-225B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001627
KISS 202C KISSUSB (Device), RS232 InteM16x1 AG male45–200814
KISS 205B Badumwälzthermostat mit Edelstahlbad KISSUSB (Device), RS232 InteM16x1 AG male45–200914
KISS 208B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001014
KISS 212B KISSUSB (Device), RS232 InteM16x1 AG male25–2001114
KISS 215B KISSUSB (Device), RS232 InteM16x1 AG male25–2001214
KISS 220B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001414
KISS 225B KISSUSB (Device), RS232 InteM16x1 AG male25–2001614

Full specifications

Core Performance & Functional Specifications
Bath volume 15 l
Pump connection M16x1 AG male (Option, #19607)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 25...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 350 x 375 x 425 mm
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity pressure pump max. mfr 27 l/min; 0, 7 bar l/min; bar
Delivery capacity suction pump max. mfr 22 l/min; 0, 4 bar l/min; bar
Bath opening WxD/BD mfr 290 x 152 / 200 mm
Filling volume min. mfr 11, 3 l
Dimensions, Weight & Mechanical Design
Weight 12 kg
Electrical, Power & Drive Specifications
Current consumption 10 A
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behaviour and media

Control, installation and operation

Integration & Operating Notes

Electrical
Branch protection must be sized between 10 A and 16 A

The unit draws up to 10 A at 200–240 V; a fuse below 10 A will trip under full heating load, while a fuse above 16 A will not protect the supply cable. An external UL 489 branch circuit protection device is also required for compliance.

Datasheet, Specifications (Installation
Process Limits
Polyglycol is not permissible with a Huber pump adapter

When a Huber pump adapter is fitted, polyglycol must not be used as the heat transfer fluid. Select an alternative fluid compatible with the adapter materials and the required temperature range.

Datasheet, Operation (page 2
Cooling & Media
Sub-ambient operation requires an external cooling source

The base unit reaches 25–200 °C without cooling; water cooling extends the lower limit to 20 °C, and a refrigerator is needed for -30 °C. The bath bridge has a hole for a cooling probe to accept an immersion cooler such as TC45–TC100E.

Datasheet, Overview and Selection
Maintenance
Order the drain valve accessory if fluid draining is required

A drain valve with cap (#6839) is available as an accessory for emptying the bath; it is not part of the base delivery. Plan for it at order time to simplify fluid changes and maintenance.

Datasheet, Maintenance (page 2

Selection & Integration Q&A

When should I choose the CC-215B over the CC-212B or CC-205B?

Choose the CC-215B when a 15-litre bath and 1.5–2.1 kW heating are needed; the CC-212B and CC-205B offer smaller volumes. The CC-202C uses a different controller generation, so select it only if its control features match your requirements.

What power supply and branch protection does the CC-215B require?

It accepts 200–240 V single- or two-phase at 50/60 Hz and draws up to 10 A. Provide branch protection between 10 A and 16 A with an external UL 489 device, and specify the country-specific plug at order time.

Can I operate the CC-215B below 25 °C without adding cooling?

No. Without cooling the range is 25–200 °C; water cooling extends it to 20 °C, and a refrigerator is required for -30 °C. The bath bridge has a hole for an immersion cooler such as TC45–TC100E.

What maintenance accessories are not included in the base delivery?

The drain valve with cap (#6839) is an accessory for emptying the bath and is not included. The country-specific power plug must also be specified at order time; no cooling unit is part of the base model.