Peter Huber Kältemaschinenbau KISS 215B
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS 215B

Peter Huber Kältemaschinenbau KISS 215B — Thermostatic baths.

The KISS 215B is a 15-litre stainless-steel bath immersion circulator with a non-variable pressure/suction pump and KISS controller. Its usable range is configuration-dependent: 25–200 °C uncooled, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. It serves as the temperature-control core for external or internal circulation loops in laboratory processes.

Direct drop-in replacements

Eppendorf
MixMate ®
53%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection

Model ControllerInterfacePump connectionTemperature range, °CWeight, kgMax delivery flow rate, liter/minute
KISS 215B Current KISSUSB (Device), RS232 InteM16x1 AG male25–2001214
CC-202C Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200922
CC-205B Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200927
CC-208B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–2001027
CC-212B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001122
CC-215B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–20012—
CC-220B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001427
CC-225B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001627
KISS 202C KISSUSB (Device), RS232 InteM16x1 AG male45–200814
KISS 205B Badumwälzthermostat mit Edelstahlbad KISSUSB (Device), RS232 InteM16x1 AG male45–200914
KISS 208B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001014
KISS 212B KISSUSB (Device), RS232 InteM16x1 AG male25–2001114
KISS 220B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001414
KISS 225B KISSUSB (Device), RS232 InteM16x1 AG male25–2001614

Full specifications

Core Performance & Functional Specifications
Alarm output optic, acoustic
Bath volume 15 l
Dimensions (w × D × H) 350 x 375 x 425 mm
Filling volume 10 l (min)
Maximum delivery flow rate 14 l/min (pressure pump; 0, 25 bar)
Pump connection M16x1 AG male (Option, #19607)
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 25...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 05 K
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump; 0, 17 bar)
Bath opening WxD/BD mfr 290 x 152 / 200 mm (WxD/depth)
Dimensions, Weight & Mechanical Design
Weight 12 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Installation, control and operation

Integration & Operating Notes

Electrical
Provide UL 489 branch protection; unit does not include it

The datasheet requires an external branch-circuit protection device according to UL 489. The supply must be 200–240 V, 1~ or 2~, 50/60 Hz, with a fuse rated between 10 A minimum and 16 A maximum; the maximum current draw is 10 A. Plan the panel and cabling around these limits before installation.

KISS 215B Immersion Circulator Datasheet, page 2
Electrical
Voltage and frequency tolerances must not oppose each other

Both voltage and frequency may deviate by ±10%, but the datasheet prohibits a combination where the frequency tolerance runs in the opposite direction to the voltage tolerance. Verify the site supply or generator envelope against this rule; a supply that is, for example, -10% voltage with +3% frequency is not permitted.

KISS 215B Immersion Circulator Datasheet, page 2
Process Limits
Select cooling method to reach the required setpoint

The usable temperature range depends on the cooling utility: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. The bath bridge includes a hole for a cooling probe such as an immersion cooler TC45–TC100E. Specify the cooling configuration at the system level; the bath alone cannot reach sub-ambient setpoints.

KISS 215B Immersion Circulator Datasheet, page 1
Cooling & Media
Avoid acetone and polyglycols with the standard plastic pump

The plastic pump is not resistant to acetone or polyglycols (depending on the manufacturer). Using these media with the standard pump risks damage. Either avoid these fluids or order a more resistant plastic. For water-based operation near or below 0 °C, mix water with glysantine or ethylene glycol for freeze protection.

KISS 215B Immersion Circulator Datasheet, page 2
Control & I/O
Pt100 option is factory-fitted and display-only

The optional Pt100 sensor connection (#10688) is available only factory-fitted and cannot be retrofitted in the field. It displays, but does not control, process temperature. If process-temperature monitoring is required, include this option in the original order; it cannot be added later.

KISS 215B Immersion Circulator Datasheet, page 1
Control & I/O
Autostart restarts after mains loss; assess process risk

The autostart function restarts the unit automatically after a power failure. For processes where an unattended restart could be hazardous or could invalidate a run, this behavior must be considered at the system level. The datasheet notes it may be undesirable for some processes; plan the process logic accordingly.

KISS 215B Immersion Circulator Datasheet, page 1
Mounting
Specify country-specific plug at order

Single- and two-phase devices are supplied with a power cable and a country-specific plug, which must be specified when ordering. Confirm the site receptacle type before placing the order to avoid a mismatch. The unit is rated IP20 and must be installed where ambient temperature remains between 5 °C and 40 °C.

KISS 215B Immersion Circulator Datasheet, page 2

Selection & Integration Q&A

When should I choose the KISS 215B rather than a smaller sibling such as the CC-205B or CC-208B?

Choose the 215B when the application needs the 15 l bath with a 10 l minimum fill and the 290 x 152 mm opening for larger immersion coolers or sample loads. Smaller siblings suit lower working volumes; the 215B trades footprint for bath capacity and immersion depth.

What adjacent system parts must be provided for the KISS 215B to operate?

Provide a 200–240 V, 1~ or 2~, 50/60 Hz supply with a 10–16 A fuse and an external UL 489 branch-circuit protection device. For sub-ambient work, add water cooling or a refrigerator; the bath bridge accepts an immersion cooler TC45–TC100E. Specify the country-specific plug at order.

What ambient and duty-cycle limits apply to the KISS 215B installation?

Ambient temperature must stay between 5 °C and 40 °C during operation. The unit is IP20 and EN 55011 Class A, Group 1, so additional EMC measures may be needed in residential or light-industrial environments. Voltage and frequency tolerances are ±10% each but must not oppose each other.

What maintenance and accessory items are not included in the base delivery of the KISS 215B?

The base unit does not include the external UL 489 branch protection, the cooling probe or immersion cooler, or the factory-fitted Pt100 sensor connection (#10688), which cannot be retrofitted. The country-specific plug must be specified at order. For freeze protection below 0 °C, glysantine or ethylene glycol must be added to water.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS 215B Immersion Circulator Datasheet
Datasheet ·97,3 КБ ·2 pp.