Peter Huber Kältemaschinenbau Minichiller 280 OLÉ Kompakter
Peter Huber Kältemaschinenbau Minichiller 280 OLÉ Kompakter — Recirculating chillers with Noise level 53 dB.
The Minichiller 280 OLÉ Kompakter is a compact air-cooled recirculating chiller with a pressure/suction pump and OLED controller, rated -5 to 40 °C with 0.28 kW cooling at 15 °C. Its 80 °C return-line limit and 1.4 l minimum fill distinguish it from larger 1000-series siblings. It serves low-heat-load laboratory and industrial cooling loops.
Peter Huber Kältemaschinenbau Minichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Minichiller 280 OLÉ Kompakter Current | 53 | OLÉ | I / NFL | -5–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 1000 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H Kompakter | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w Kompakter | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200w-H Kompakter | — | Pilot ONE | III / FL | -20–100 | — | — |
| Minichiller 1200w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 300 OLÉ Kompakter | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300w-H OLÉ Kompakter | 51 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300w OLÉ Kompakter | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 600-H OLÉ Kompakter | 56 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 600 Kompakter | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 110-120V 1~ 60Hz |
| Minichiller 600 OLÉ Kompakter | 56 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 60Hz |
| Minichiller 600w OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 800 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 800w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
Full specifications
| Noise level | 53 dB(A) |
| Expansion volume | 2 l |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure- and suction pump |
| Temperature range | -5...40 °C |
| Temperature stability | 1 K |
| Weight | 23 kg |
| Power supply | 220-240V 1~/2~ 50/60Hz |
| Controller | OLÉ |
| Display resolution | 0, 1 K |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Maximum return flow temperature | 80 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Air-cooled refrigeration circuit using R-290 (A3, H220) with a 0.041 kg charge; plastic pressure/suction pump with M16x1 male connections.
- Safety class I / NFL per DIN 12876; IP20 enclosure; 23 kg; OLED display with 0.1 K resolution.
- Within the Minichiller series, the 280 OLÉ is the smallest-capacity variant; the 1000 and 1000-H/w-H siblings offer higher cooling capacity and, for the -H versions, different pump characteristics. Choose the 280 when the heat load is low and space is tight.
Behavior under load and media
- Cooling capacity is 0.28 kW at 15 °C and 0.2 kW at 0 °C, so the unit is limited to low-heat-load applications; capacity drops as ambient temperature rises above 5–40 °C.
- The pump delivers up to 14 l/min at 0.25 bar pressure and 10.5 l/min suction at 0.17 bar; pump speed is not adjustable.
- The plastic pump is not resistant to acetone or polyglycols; for freeze protection, mix water with glysantine or ethylene glycol. Minimum filling capacity is 1.4 l with a 2 l expansion volume.
Installation, control and operation
- Requires 220–240 V 1~/2~ 50/60 Hz supply with a 10–16 A fuse; voltage tolerance ±5% only with simultaneous frequency tolerance ±2%, and -5% voltage with +2% frequency is not allowed.
- Interfaces: USB (Device) and RS232; controller is OLÉ with OLED display.
- Do not install valves or quick-release couplings in feed/discharge lines; install a drip tray below the unit; protect electrical lines below the unit from liquid ingress; do not install in an ATEX zone.
Integration & Operating Notes
The chiller is rated for -5 to 40 °C operation but tolerates up to 80 °C in the return line. Exceeding this limit can damage the refrigeration circuit and pump. Verify the process return temperature before connecting high-temperature loops.
The unit accepts 220–240 V 1~/2~ 50/60 Hz with a 10–16 A fuse. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%; combining -5% voltage with +2% frequency is not allowed. Confirm the site supply stays within this envelope to avoid compressor damage.
A condensation water / thermal fluid drip tray must be installed below the unit. Electrical lines directly below the unit must be protected against liquid ingress because condensation can form at the cooling water connections. This prevents shock hazard and fluid damage.
The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). Using these media can cause pump failure and leakage. For freeze protection, mix water with glysantine or ethylene glycol instead.
If salts such as chlorides or bromide are added to the water circuit, anti-corrosion agents are mandatory. Only drinking water or filtered river/sea water may be used as cooling water; sea water must not be used. This protects the cooling water circuit from corrosion.
Selection & Integration Q&A
Choose the 280 OLÉ for low heat loads (0.28 kW at 15 °C) and tight spaces. The 1000-series offers higher cooling capacity; the -H variants have different pump characteristics for higher pressure drop. Select the 280 when its capacity and 14 l/min pump flow suffice.
The pump connection is M16x1 male. Included accessories are a NW12 hose connector and M16x1 sleeve nuts, so hoses must match this thread. Do not install valves or quick-release couplings in the feed/discharge lines to avoid overpressure damage.
Use drinking water or filtered river/sea water only; sea water is prohibited. Add anti-corrosion agents if salts are present. For freeze protection, mix water with glysantine or ethylene glycol. The base delivery includes hose connector and sleeve nuts; hoses and heat transfer fluid are not included.