Peter Huber Kältemaschinenbau Unistat 425w Dynamisches
Peter Huber Kältemaschinenbau Unistat 425w Dynamisches — Temperature control units with Noise level 61 dB.
The Unistat 425w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -40 to 250 °C with 2 kW heating. It is the dynamic (high-flow) modification of the 4xx series, delivering up to 91 l/min at 1.5 bar for externally closed circuits. It serves precise temperature control of reactors, distillation systems and other jacketed process equipment.
Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 425w Dynamisches Current | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Grande Fleur Dynamisches | — | — | — | — | — | — |
| Grande Fleur w Dynamisches | — | 0, 01 | 208-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -40–200 |
| Grande Fleur w-eo Dynamisches | — | — | — | — | — | — |
| Petite Fleur-eo Dynamisches | 58 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Petite Fleur w Dynamisches | — | 0, 01 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | Pressure pump, variable | -40–200 |
| Unistat 410w Dynamisches | 61 | 0, 01 K | 220-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | -45–250 |
| Unistat 430w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P405w Dynamisches | — | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | -45–250 |
| Unistat P425 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P425w Dynamisches | 61 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430 Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
| Unistat P430w Dynamisches | 65 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | -40–250 |
Full specifications
| Noise level | 61 dB(A) |
| Control output | POKO ONE |
| Heating capacity | 2 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -40...250 °C |
| Weight | 177 kg |
| Current consumption | 13 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 150 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing water-cooled unit (suffix 'w') with a magnetically coupled, stainless-steel pump; pump speed is adjustable.
- Dynamic ('Dynamisches') variant of the 4xx series, distinguished by high delivery (91 l/min max, 1.5 bar max) for circuits with low pressure drop.
- Sibling models in the series include the Grande Fleur variants; this model is the 425w dynamic configuration.
- Controller: Pilot ONE with POKO ONE control output; display resolution 0.01 K; external Pt100 sensor input.
- Safety class III / FL (flammable fluid rating per DIN 12876).
Working range and media behavior
- Temperature range -40 to 250 °C; heating capacity 2 kW.
- Cooling water circuit: G1/2 male connections, 1 bar min differential pressure, 6 bar max pressure, 150 l/h consumption at 15 °C cooling water and 0 °C flow.
- Cooling capacity decreases as cooling water temperature rises; required flow may increase accordingly.
- Refrigerant R-452A, charge 0.97 kg (2.1 t CO2 equivalent, GWP 2141).
- Minimum filling capacity 3.6 l; expansion tank 3.3 l.
- Materials in the cooling water circuit include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only compatible cooling water may be used.
Installation, control and operation
- Power supply 400 V 3~ 50 Hz, max current 13 A, fuse 3x16 A; delivered with cable but without plug (three-phase devices <63 A).
- Leakage current >3.5 mA; protective earthing and RCD selection must account for this.
- Voltage/frequency tolerances apply together: ±5% voltage with ±2% frequency; opposite-sign deviations are not allowed.
- IP20; ambient temperature 5–40 °C; must not be installed in an ATEX zone.
- Interfaces: Ethernet, USB (host and device), RS232.
- Pilot ONE offers setup of overtemperature protection, Delta T limiter, process safety mode, and automatic or program-based temperature control.
- For externally closed applications the expansion vessel must remain unlocked; for externally open (bath) applications it must be locked.
- Cooling water intake designed for 5–20 °C; lower temperatures require insulating the cooling water pipes.
Integration & Operating Notes
The equipment has a leakage current greater than 3.5 mA. Protective earthing and the selection of residual current protective devices must account for this; a standard 30 mA RCD may not be suitable without engineering review.
The unit must not be installed or started up within an ATEX zone; explosion protection requires a separate Ex px cabinet. The IP20 rating and ambient temperature range of 5–40 °C define the allowable installation location. A condensation water / heat transfer fluid drip tray must be installed below the unit; national legislation may prescribe it.
In Process Safety mode, when the overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. The power-interruption response can be configured as Off/Standby (default) or ON/Temperature control active; choose according to whether unattended restart is acceptable.
Selection & Integration Q&A
Choose the 425w Dynamisches for high-flow, low-pressure-drop externally closed circuits needing up to 91 l/min at 1.5 bar and -40 to 250 °C. Grande Fleur variants differ in configuration; confirm pump curve and cooling capacity against your circuit's pressure drop before selecting.
A 400 V 3~ 50 Hz supply with 3x16 A fusing; max current 13 A. The unit comes with cable but no plug, so a suitable plug must be provided. Leakage current >3.5 mA requires RCD and earthing design review. Voltage/frequency tolerances apply together (±5% / ±2%).
Install indoors, IP20, ambient 5–40 °C, not in an ATEX zone. Transport and store upright only; allow ~2 h acclimation after temperature changes before power-on. Leveling feet must be activated. A drip tray is required below the unit. Cooling water intake 5–20 °C; shut off cooling water during overnight downtimes.
Base delivery excludes the power plug, cooling water hoses, heat transfer fluid, and drip tray. Periodic tasks include cleaning the hat-type strainer, inspecting hoses and mechanical seal, and checking the emergency stop switch. Refrigerant work requires approved contractors. A gas detection sensor and processing unit are optional and installed by the operating company.