Peter Huber Kältemaschinenbau CC-104A
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-104A

Peter Huber Kältemaschinenbau CC-104A — Thermostatic baths.

The CC-104A is a 3-litre immersion circulator with a transparent polycarbonate bath, rated for 15–100 °C with a refrigerator, 20–100 °C with water cooling, or 25–100 °C uncooled. Its variable-speed pressure/suction pump delivers up to 27 l/min at 0.7 bar, and the Pilot ONE controller holds stability to 0.02 K. It serves as a compact heating/circulation source for external loops and small immersed samples.

Direct drop-in replacements

Eppendorf
MixMate ®
47%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minutePump connectionWeight, kgAmbient temperature limit, °C
CC-104A Current Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-106A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male55–40
CC-108A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #65–40
CC-110A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-112A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #85–40
CC-118A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male (Option, #85–40
KISS 104A KISSUSB (Device), RS232 Inte14M16x1 AG male55–40
KISS 106A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #55–40
KISS 108A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 110A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 112A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40
KISS 118A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40

Full specifications

Core Performance & Functional Specifications
Bath volume 3 l
Dimensions (w × D × H) 147 x 235 x 330 mm
Filling volume 2 l (minimum)
Heating capacity 1.5 - 2.1 kW
Maximum delivery flow rate 27 l/min (pressure pump; also 0, 7 bar)
suction pump; also 0, 4 bar: 22 l/min
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 25...100 °C
with cooling unit: 15...100 °C · with water cooling: 20...100 °C
Temperature stability 0.02 K
Bath opening WxD/BD mfr ø = 25 / 150 (opening WxD/depth combined)
Dimensions, Weight & Mechanical Design
Weight 6 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Installation, control and operation

Integration & Operating Notes

Electrical
Branch circuit needs UL 489 breaker and 10–16 A fuse

The installer must provide external branch circuit protection listed to UL 489; the circuit must be rated between the 10 A minimum and 16 A maximum fuse. A circuit outside this window either trips on inrush or fails to protect the 10 A maximum current draw.

Datasheet, Technical Specifications / Installation
Electrical
Avoid −10 % voltage combined with +3 % frequency

Voltage and frequency each tolerate ±10 %, but the combination of a 10 % undervoltage with a 3 % overfrequency is explicitly not allowed. Site power quality should be checked before commissioning, since this corner of the envelope can overheat the compressor or pump motor.

Datasheet, Installation
Process Limits
Cooling source fixes the lower end of the range

Without cooling the unit cannot go below 25 °C; water cooling extends it to 20 °C and a refrigerator to 15 °C. The setpoint specification must therefore be matched to the installed cooling method, and the 100 °C ceiling is set by the polycarbonate bath, not the controller.

Datasheet, Product Overview / Technical Specifications
Control & I/O
External Pt100 allows control at the load, not the bath

The Pilot ONE controller accepts an external Pt100 sensor, so the control point can be placed at the sample or in the external loop instead of the bath. This matters when the 2 l filling volume and small bath would otherwise dominate the measured temperature.

Datasheet, Technical Specifications
Maintenance
Register the unit to obtain the 4-year warranty

The extended 4-year warranty is granted only on registration; without it the standard coverage applies. Registration should be completed at commissioning so the warranty period is not lost.

Datasheet, Product Overview

Selection & Integration Q&A

When should I choose the CC-104A rather than a larger sibling such as the CC-106A or CC-108A?

Pick the CC-104A when the load fits a 25 mm × 150 mm opening and a 2 l filling volume, or when the external loop is short. Larger siblings add bath volume for bigger immersed samples or more loop inventory; they do not change the 100 °C polycarbonate ceiling.

What must the external circuit and its fittings provide to work with this unit's pump?

The loop must be sized against 27 l/min at 0.7 bar on the pressure side and 22 l/min at 0.4 bar on the suction side, with M16x1 male pump connections requiring matching hose adapters. Excessive loop pressure drop will reduce circulation below these maxima.

What site conditions and electrical protection does the installation require?

A 200–240 V, 1~ or 2~, 50/60 Hz supply with a 10–16 A fuse and external UL 489 branch protection, in a dry location at 5–40 °C ambient. Avoid −10 % voltage combined with +3 % frequency; specify the country-specific plug at ordering.

What is not included in base delivery that I must plan for?

The cooling source (water cooling or refrigerator) needed below 25 °C, hose adapters for the M16x1 pump connections, and the external Pt100 sensor if control at the load is required. The power cable and country plug are included only if specified at ordering.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-104A Immersion Circulator
Datasheet ·134,2 КБ ·2 pp.
View PDF