Peter Huber Kältemaschinenbau KISS 118A
Peter Huber Kältemaschinenbau KISS 118A — Thermostatic baths.
The KISS 118A is an immersion circulator with an 18-litre bath and a pressure/suction pump, delivering 15–100 °C with a refrigerator, 20–100 °C with water cooling, or 25–100 °C uncooled. Its 0.05 K stability at 70 °C and DIN 12876 class III/FL protection distinguish it within the KISS series for precise thermostatic control.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection
| Model | Controller | Interface | Max delivery flow rate, liter/minute | Pump connection | Weight, kg | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| KISS 118A Current | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
| CC-104A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-106A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 5 | 5–40 |
| CC-108A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 6 | 5–40 |
| CC-110A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-112A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 8 | 5–40 |
| CC-118A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 104A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male | 5 | 5–40 |
| KISS 106A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 5 | 5–40 |
| KISS 108A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 110A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 112A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
Full specifications
| Alarm output | optic, acoustic |
| Bath volume | 18 l |
| Dimensions (w × D × H) | 333 x 520 x 335 mm |
| Filling volume | 11 l (min) |
| Maximum delivery flow rate | 14 l/min (pressure pump; 0, 25 bar) |
| Pump connection | M16x1 AG male (Option, #19606) (Option, #19606) |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Temperature range | 25...100 °C with cooling unit: 15...100 °C · with water cooling: 20...100 °C |
| Temperature stability mfr | ±0, 05 K |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min (suction pump; 0, 17 bar) |
| Bath opening WxD/BD mfr | 275 x 321 / 150 mm (WxD/depth) |
| Weight | 8 kg |
| Current consumption | 10 A (max) |
| Power supply | 200-240V 1~/2~ 50/60Hz V; Hz |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The KISS 118A is a bath-mounted immersion circulator with an 18-litre bath volume (minimum filling 11 l) and a pressure/suction pump that is not speed-variable. It differs from the CC-104A through CC-110A siblings by its larger bath and integrated pump configuration. The unit uses a KISS controller with an OLED display and provides USB (Device) and RS232 interfaces. The bath opening measures 275 × 321 mm with 150 mm depth, and the overall case is 333 × 520 × 335 mm at 8 kg.
Thermal behaviour and media
Temperature range depends on the cooling method: 25–100 °C without cooling, 20–100 °C with water cooling, and 15–100 °C with a refrigerator. Stability is ±0.05 K at 70 °C. Heating capacity is 1.5–2.1 kW. The pressure pump delivers up to 14 l/min at 0.25 bar, while the suction pump reaches 10.5 l/min at 0.17 bar. The plastic pump is not resistant to acetone or polyglycols; for freeze protection, water should be mixed with glysantine or ethylene glycol.
Installation and control
Power supply is 200–240 V, 1~/2~, 50/60 Hz with a maximum current of 10 A. External branch circuit protection per UL 489 is required. Voltage and frequency tolerances are ±10% only if they do not run in opposite directions. Single- and two-phase units are delivered with a power cable and country-specific plug (must be specified at ordering). The unit is IP20 and must be operated between 5 °C and 40 °C ambient. An optional factory-fitted Pt100 sensor connection (#10688) can display, not control, process temperature. The autostart function restarts after power failure, and adjustable overtemperature protection per DIN 12876 with safety class III/FL is provided.
Integration & Operating Notes
The datasheet requires branch circuit protection according to UL 489. The installer must provide a listed breaker sized for the 10 A maximum current; omitting it violates the installation instructions and may void compliance.
A ±10% voltage tolerance is permitted only if the frequency tolerance does not run in the opposite direction. For example, -10% voltage with +3% frequency is not allowed. Verify the site supply envelope before connection.
The standard plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). Using these media risks pump damage; select an alternative fluid or consult the factory for a resistant pump option.
For operation near or below 15 °C, water should be mixed with glysantine or ethylene glycol to prevent freezing. Pure water may freeze and damage the bath or pump if the unit is used with a refrigerator.
The optional Pt100 connection (#10688) displays process temperature but does not control it, and is only available factory fitted. It cannot be retrofitted in the field, so specify it at ordering if external monitoring is needed.
The autostart function automatically restarts the unit after a power failure. For processes where unattended restart is hazardous, this behaviour must be considered; the datasheet lists it as a standard function.
Selection & Integration Q&A
Choose the KISS 118A when an 18-litre bath and a pressure/suction pump are needed. The CC-104A through CC-110A siblings have different bath sizes; select them for smaller volumes or different pump configurations. The 118A offers 15–100 °C with a refrigerator and 0.05 K stability.
It requires 200–240 V, 1~/2~, 50/60 Hz with a maximum current of 10 A. External branch circuit protection per UL 489 is mandatory. Single- and two-phase units come with a power cable and country-specific plug, which must be specified at ordering.
The base delivery includes the circulator and, for single/two-phase units, a power cable with country-specific plug. The Pt100 sensor connection (#10688) is an optional factory-fitted extra. No maintenance schedule is specified in the datasheet; consult the manual for service intervals.