Peter Huber Kältemaschinenbau KISS 110A
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS 110A

Peter Huber Kältemaschinenbau KISS 110A — Thermostatic baths.

The KISS 110A is a 7.5-litre polycarbonate-bath immersion circulator with a fixed-speed pressure/suction pump and 1.5–2.1 kW heater, covering 25–100 °C without cooling, 20–100 °C with water cooling, or 15–100 °C with a refrigerator. Its Safety Class III/FL and ±0.05 K stability distinguish it within the KISS series for precise, fail-safe temperature control of external circuits and bath-mounted samples.

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minutePump connectionWeight, kgAmbient temperature limit, °C
KISS 110A Current KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
CC-104A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-106A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male55–40
CC-108A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #65–40
CC-110A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-112A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #85–40
CC-118A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male (Option, #85–40
KISS 104A KISSUSB (Device), RS232 Inte14M16x1 AG male55–40
KISS 106A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #55–40
KISS 108A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 112A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40
KISS 118A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40

Full specifications

Core Performance & Functional Specifications
Alarm output optic, acoustic
Dimensions (w × D × H) 147 x 507 x 330 mm
Maximum delivery flow rate 14 l/min (pressure pump; 0, 25 bar)
Pump connection M16x1 AG male (Option, #19606) (Option, #19606)
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 25...100 °C
with cooling unit: 15...100 °C · with water cooling: 20...100 °C
Temperature stability mfr ±0, 05 K
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump; 0, 17 bar)
Bath opening WxD/BD mfr 130 x 310 / 150 mm (WxD/depth)
Bath volume mfr 7, 5 l
Filling volume min. mfr 4, 4 l (minimum)
Dimensions, Weight & Mechanical Design
Weight 6 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz V; Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behaviour and media

Installation, connection and control

Integration & Operating Notes

Electrical
Branch circuit must be fused 10–16 A with UL 489 protection

The unit draws up to 10 A on a 200–240 V single/two-phase supply. Install external branch circuit protection according to UL 489 and size fusing between 10 A and 16 A; undersizing risks nuisance trips, oversizing risks inadequate fault clearing.

Datasheet, Specifications (installation notes
Process Limits
Cooling method sets the usable lower temperature limit

Without cooling the bath cannot go below 25 °C; water cooling extends the range to 20 °C, and a refrigerator to 15 °C. Select the cooling utility before specifying setpoints below 25 °C, and provide the corresponding water or refrigerant circuit.

Datasheet, Overview
Cooling & Media
Avoid acetone and polyglycols with the standard plastic pump

The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). For freeze protection near or below ambient, mix water with glysantine or ethylene glycol rather than using pure water, which can freeze and damage the circuit.

Datasheet, Operation
Mounting
Install in a dry location within 5–40 °C ambient

IP20 protection means the unit must be kept dry and protected from water ingress. Ambient temperature at the installation site must stay between 5 °C and 40 °C to avoid derating or damage; condensation and splashing must be prevented.

Datasheet, Installation
Control & I/O
Optional Pt100 input is display-only, not for closed-loop control

The factory-fitted Pt100 sensor connection (#10688) only displays process temperature and cannot be used to control the bath. Do not specify it expecting external-sensor closed-loop control; use the internal KISS controller for setpoint regulation.

Datasheet, Selection
Electrical
Specify country-specific plug at ordering

Single/two-phase devices are supplied with a power cable and a country-specific plug that must be stated when ordering. Voltage and frequency tolerances of ±10% are permitted only if they do not act in opposite directions.

Datasheet, Installation

Selection & Integration Q&A

When should I choose the KISS 110A rather than a CC-104A/106A/108A/110A sibling?

Choose the KISS 110A when you need a 7.5 l bath with 25–100 °C range (15–100 °C with refrigerator) and the fixed-speed pressure/suction pump. Siblings differ in configuration; confirm bath volume and cooling option against your sample load before selecting.

What electrical supply and protection must be provided for the KISS 110A?

Provide 200–240 V, 1~ or 2~, 50/60 Hz with a country-specific plug specified at ordering. Maximum current is 10 A; install branch-circuit fusing of 10–16 A and external UL 489 protection. Voltage/frequency tolerance ±10% only if not opposing.

What maintenance and accessories are not included in the base delivery?

The base unit does not include the optional Pt100 sensor connection (#10688, factory-fitted) or the M16×1 AG male pump connection (#19606). A country-specific plug must be specified at ordering. No maintenance schedule is given in the datasheet; follow the manual for service intervals.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS 110A Immersion Circulator
Datasheet ·108,0 КБ ·2 pp.
View PDF