Peter Huber Kältemaschinenbau CC-110A
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-110A

Peter Huber Kältemaschinenbau CC-110A — Thermostatic baths.

The CC-110A is a 7.5 l polycarbonate-bath immersion circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated 20–100 °C with water cooling (25–100 °C without, 15–100 °C with refrigerator). It differs from smaller CC-104A–CC-108A siblings by bath volume and 1.5–2.1 kW heating capacity, and serves as a transparent, ±0.02 K-stable thermostating stage for external loops and bath-mounted samples up to 100 °C.

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minutePump connectionWeight, kgAmbient temperature limit, °C
CC-110A Current Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-104A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-106A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male55–40
CC-108A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #65–40
CC-112A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #85–40
CC-118A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male (Option, #85–40
KISS 104A KISSUSB (Device), RS232 Inte14M16x1 AG male55–40
KISS 106A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #55–40
KISS 108A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 110A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 112A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40
KISS 118A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40

Full specifications

Core Performance & Functional Specifications
Dimensions (w × D × H) 147 x 507 x 330 mm
Maximum delivery flow rate 27 l/min (pressure pump; 0, 7 bar)
suction pump; 0, 4 bar: 22 l/min
Pump connection M16x1 AG male (Option, #19606)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 25...100 °C
with cooling unit: 15...100 °C · with water cooling: 20...100 °C
Temperature stability mfr ±0, 02 K
Heating capacity mfr 1, 5 - 2, 1 kW
Bath opening WxD/BD mfr 130 x 310 / 150 mm
Bath volume mfr 7, 5 l
Filling volume min. mfr 5, 2 l (min)
Dimensions, Weight & Mechanical Design
Weight 6 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Installation, control and operation

Integration & Operating Notes

Electrical
Size the branch circuit for 10 A load and 10–16 A fusing

The unit draws up to 10 A at 200–240 V; the upstream fuse must be at least 10 A and no more than 16 A. Undersized wiring or a fuse below 10 A will trip during heat-up, while a fuse above 16 A removes the intended protection.

CC-110A Technical Data Sheet, Specifications
Electrical
Verify supply quality: ±10% voltage, no opposing frequency drift

Voltage may deviate ±10% only if the frequency tolerance does not run in the opposite direction. A supply that is simultaneously high-voltage and low-frequency (or vice versa) is outside the permissible envelope and can stress the compressor/electronics.

CC-110A Technical Data Sheet, Installation
Electrical
Provide UL 489 branch protection and a plug for three-phase units

External branch circuit protection according to UL 489 is mandatory. Three-phase devices below 63 A arrive with cable but without a plug, so the customer must supply a suitable mating plug and complete the connection on site.

CC-110A Technical Data Sheet, Installation
Process Limits
Match cooling method to the required low temperature

The achievable range depends on the cooling source: 25–100 °C without cooling, 20–100 °C with water cooling, 15–100 °C with a refrigerator. Selecting the unit without planning the cooling method limits the process to 25 °C and above.

CC-110A Technical Data Sheet, Overview
Process Limits
Never exceed 100 °C in the polycarbonate bath

The transparent bath is polycarbonate and rated to a maximum of 100 °C. Operation above this temperature risks deformation or failure of the bath; higher-temperature processes require a different bath material.

CC-110A Technical Data Sheet, Specifications
Process Limits
Charge at least 5.2 l of heat transfer fluid

Bath volume is 7.5 l with a minimum filling capacity of 5.2 l. Underfilling can cause pump cavitation and unstable temperature control; overfilling risks overflow when the fluid expands at 100 °C.

CC-110A Technical Data Sheet, Specifications
Cooling & Media
Do not use polyglycol with the Huber pump adapter

When the optional Huber pump adapter is fitted, polyglycol is not permissible as heat transfer fluid. Fluid selection must therefore be decided together with the pump-connection option, not after commissioning.

CC-110A Technical Data Sheet, Operation
Mounting
Install indoors only; IP20 excludes water ingress

The IP20 rating means no protection against water. The unit must be placed in a dry indoor location, and ambient temperature must stay between 5 °C and 40 °C; condensation or splash near the bath will enter the enclosure.

CC-110A Technical Data Sheet, Specifications
Maintenance
Re-verify the overtemperature cut-out during each service

Because the cut-out is adjustable, its setting can drift or be changed by operators. Maintenance should include a functional check of the overtemperature protection in addition to fluid inspection, so that the safety chain remains valid over the unit's life.

CC-110A Technical Data Sheet, Specifications

Selection & Integration Q&A

When should I choose the CC-110A rather than the smaller CC-104A/CC-106A/CC-108A or the larger CC-112A?

Pick the CC-110A when the process needs a 7.5 l transparent bath and up to 2.1 kW heating with ±0.02 K stability. Smaller siblings suit lower volumes; the CC-112A is for larger baths. All share the 100 °C polycarbonate ceiling.

What external cooling and fluid circuit must be provided for the CC-110A to reach its lower temperatures?

Without cooling the unit is limited to 25–100 °C; water cooling extends it to 20–100 °C and a refrigerator to 15–100 °C. The external loop must handle 27 l/min at 0.7 bar on the pressure side, and polyglycol is not allowed with the Huber pump adapter.

What electrical supply and branch protection does the CC-110A require at the installation site?

A 200–240 V, 1~ or 2~, 50/60 Hz supply with max. 10 A draw, fused between 10 A and 16 A, plus UL 489 branch protection. Voltage tolerance is ±10% only if frequency does not drift in the opposite direction; three-phase units arrive without a plug.

What maintenance and accessory scope should be planned beyond the base delivery?

The base unit excludes the M16x1 AG pump connection (option #19606) and any mating plug for three-phase supply. Maintenance must include checking the adjustable DIN 12876 overtemperature cut-out and the heat transfer fluid; no cables or consumables are included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-110A Immersion Circulator — Technical Data Sheet
Datasheet ·107,2 КБ ·2 pp.
View PDF