Peter Huber Kältemaschinenbau CC-112A
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-112A

Peter Huber Kältemaschinenbau CC-112A — Thermostatic baths.

The CC-112A is a 12-litre polycarbonate-bath immersion circulator with a variable-speed pressure/suction pump and Pilot ONE controller. Its 25–100 °C base range extends to 20–100 °C with water cooling and 15–100 °C with a refrigerator, distinguishing it from smaller CC-104A–CC-110A siblings. It serves as a temperature-control source for external circuits and open bath work up to 100 °C.

Direct drop-in replacements

Eppendorf
MixMate ®
53%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minutePump connectionWeight, kgAmbient temperature limit, °C
CC-112A Current Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #85–40
CC-104A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-106A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male55–40
CC-108A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #65–40
CC-110A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-118A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male (Option, #85–40
KISS 104A KISSUSB (Device), RS232 Inte14M16x1 AG male55–40
KISS 106A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #55–40
KISS 108A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 110A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 112A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40
KISS 118A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40

Full specifications

Core Performance & Functional Specifications
Bath volume 12 l
Filling volume 8 l (min.)
Maximum delivery flow rate 27 l/min (pressure pump max.; also 0, 7 bar)
suction pump max.; also 0, 4 bar: 22 l/min
Pump connection M16x1 AG male (Option, #19606) (Option, #19606)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 25...100 °C
with cooling unit: 15...100 °C · with water cooling: 20...100 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 333 x 360 x 335 mm
Heating capacity mfr 1, 5 - 2, 1 kW
Bath opening WxD/BD mfr 275 x 161 / 150 mm
Dimensions, Weight & Mechanical Design
Weight 8 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz V, Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The CC-112A is the largest bath in the CC-10xA polycarbonate series, carrying a 12 l bath volume with a minimum filling capacity of 8 l. The bath opening measures 275 x 161 mm with a 150 mm depth, and the unit occupies 333 x 360 x 335 mm at 8 kg. Unlike the smaller CC-104A through CC-110A siblings, this modification provides the full 12 l working volume and the highest heating capacity of the group at 1.5–2.1 kW. The bath bridge includes a bore hole for a cooling probe, allowing an immersion cooler such as the TC45–TC100E to be added without changing the circulator.

Thermal behaviour and media

Installation and control

Power is supplied from 200–240 V, single- or two-phase, 50/60 Hz, with a maximum current of 10 A. The datasheet requires a branch fuse of at least 10 A and no more than 16 A, plus external branch circuit protection to UL 489. Voltage tolerance is ±10% provided the frequency tolerance does not run in the opposite direction. Single- and two-phase units ship with a power cable and country-specific plug that must be specified at order. Communication and control interfaces include Ethernet, USB host and device, and RS232. Ambient temperature must stay within 5–40 °C.

Integration & Operating Notes

Electrical
Branch protection must be sized between 10 A and 16 A

The datasheet specifies a minimum fuse of 10 A and a maximum fuse of 16 A for the 200–240 V supply. Undersizing will trip on the 10 A maximum current draw; oversizing removes the intended protection. External branch circuit protection to UL 489 is also required.

CC-112A Immersion Circulator Technical Data, page 1–2
Electrical
Voltage and frequency tolerances must not oppose each other

A ±10% voltage tolerance is permitted only when the frequency tolerance does not run in the opposite direction. Verify the site supply against both limits together, not independently, before commissioning.

CC-112A Immersion Circulator Technical Data, page 2
Process Limits
Minimum temperature depends on the cooling method chosen

The unit reaches 25 °C without auxiliary cooling, 20 °C with water cooling, and 15 °C only with a refrigerator. Select the cooling configuration before specifying the process setpoint; the bath bridge bore hole accepts an immersion cooler such as TC45–TC100E for intermediate needs.

CC-112A Immersion Circulator Technical Data, page 1
Cooling & Media
Polyglycol is prohibited with the Huber pump adapter

When a Huber pump adapter is fitted, polyglycol must not be used as the heat transfer fluid. Confirm fluid compatibility with the pump adapter before filling; the M16x1 male pump connection is an option (#19606).

CC-112A Immersion Circulator Technical Data, page 2
Mounting
IP20 enclosure requires a dry, splash-free location

The IP20 rating provides no protection against water ingress. Install the unit away from splashing water and maintain ambient temperature between 5 °C and 40 °C. Respect the manufacturer's space requirements for the operating conditions before final placement.

CC-112A Immersion Circulator Technical Data, page 1–2
Electrical
Country-specific plug must be ordered with the unit

Single- and two-phase devices in the 100–240 V class are delivered with a power cable and a country-specific plug that must be specified at the time of ordering. Omitting this detail delays commissioning because the plug type is not universal.

CC-112A Immersion Circulator Technical Data, page 2
Control & I/O
Class A EMC grouping may need shielding in sensitive areas

The equipment is classified Class A, Group 1 under EN55011. In residential or EMC-sensitive environments, additional filtering or separation from sensitive instrumentation may be needed to avoid disturbance.

CC-112A Immersion Circulator Technical Data, page 2

Selection & Integration Q&A

When should I choose the CC-112A rather than a smaller CC-104A–CC-110A sibling?

Pick the CC-112A when the application needs the full 12 l bath volume, an 8 l minimum fill, and the 1.5–2.1 kW heating capacity for faster recovery or larger immersed loads. Smaller siblings suit lower-volume work where the 275 x 161 mm opening and 12 l volume are unnecessary.

What supply and branch protection does the CC-112A require from the facility?

A 200–240 V single- or two-phase, 50/60 Hz supply with a maximum current of 10 A, protected by a branch fuse of at least 10 A and no more than 16 A, plus external UL 489 branch circuit protection. Voltage tolerance is ±10% only if frequency tolerance does not oppose it.

Can the CC-112A be installed in a washdown or humid laboratory area?

No. The enclosure is IP20 and not protected against water ingress. Install it in a dry location with ambient temperature between 5 °C and 40 °C, away from splashing water, and respect the manufacturer's space requirements for the operating conditions.

What is not included in the base delivery that I must plan for?

The country-specific plug must be specified at ordering; the M16x1 male pump connection is an option (#19606); and an immersion cooler such as TC45–TC100E is a separate accessory for the bath bridge bore hole. External branch circuit protection to UL 489 is also a site responsibility.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-112A Immersion Circulator Technical Data
Datasheet ·76,5 КБ ·2 pp.
View PDF