Peter Huber Kältemaschinenbau CC-110A
Peter Huber Kältemaschinenbau CC-110A — Thermostatic baths.
The CC-110A is a 7.5 l polycarbonate-bath immersion circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated 20–100 °C with water cooling (25–100 °C without, 15–100 °C with refrigerator). It differs from smaller CC-104A–CC-108A siblings by bath volume and 1.5–2.1 kW heating capacity, and serves as a transparent, ±0.02 K-stable thermostating stage for external loops and bath-mounted samples up to 100 °C.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection
| Model | Controller | Interface | Max delivery flow rate, liter/minute | Pump connection | Weight, kg | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| CC-110A Current | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-104A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-106A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 5 | 5–40 |
| CC-108A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 6 | 5–40 |
| CC-112A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 8 | 5–40 |
| CC-118A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 104A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male | 5 | 5–40 |
| KISS 106A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 5 | 5–40 |
| KISS 108A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 110A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 112A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 118A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
Full specifications
| Dimensions (w × D × H) | 147 x 507 x 330 mm |
| Maximum delivery flow rate | 27 l/min (pressure pump; 0, 7 bar) suction pump; 0, 4 bar: 22 l/min |
| Pump connection | M16x1 AG male (Option, #19606) |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 25...100 °C with cooling unit: 15...100 °C · with water cooling: 20...100 °C |
| Temperature stability mfr | ±0, 02 K |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Bath opening WxD/BD mfr | 130 x 310 / 150 mm |
| Bath volume mfr | 7, 5 l |
| Filling volume min. mfr | 5, 2 l (min) |
| Weight | 6 kg |
| Current consumption | 10 A (max) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Transparent polycarbonate bath, 7.5 l volume, minimum filling 5.2 l; bath opening 130 x 310 mm, depth 150 mm.
- Immersion circulator with Pilot ONE controller, Pt100 external sensor input, and a variable-speed pressure/suction pump (27 l/min at 0.7 bar pressure side; 22 l/min at 0.4 bar suction side).
- Within the CC-1xxA series, the CC-110A sits between the CC-108A and CC-112A; the step-up from smaller siblings is bath volume and 1.5–2.1 kW heating capacity, while the 100 °C polycarbonate ceiling is common to the line.
- M16x1 AG male pump connection is an option (#19606); the base unit is not supplied with it.
Thermal behavior and media
- Temperature range is configuration-dependent: 25–100 °C without cooling, 20–100 °C with water cooling, 15–100 °C with a refrigerator. Stability is ±0.02 K.
- The polycarbonate bath caps operation at 100 °C; higher-temperature work is outside this modification's scope.
- With the Huber pump adapter fitted, polyglycol is not permissible as heat transfer fluid; fluid selection must follow the pump-connection configuration.
- Adjustable overtemperature protection to DIN 12876 protects bath and process; it must be set at commissioning.
Installation, control and operation
- Supply: 200–240 V, 1~ or 2~, 50/60 Hz, max. current 10 A; branch circuit fusing 10 A minimum, 16 A maximum. Voltage tolerance ±10%, provided frequency tolerance does not run in the opposite direction.
- External branch protection to UL 489 is required. Three-phase devices below 63 A are delivered with cable but without a plug.
- IP20 enclosure: dry indoor location only. Ambient 5–40 °C. EMC classification EN55011 Class A, Group 1 — industrial environment; may cause radio interference in residential areas.
- Interfaces: Ethernet, USB host and device, RS232; Pt100 external sensor connection. Pressure equipment category Art. 4.3 PED.
Integration & Operating Notes
The unit draws up to 10 A at 200–240 V; the upstream fuse must be at least 10 A and no more than 16 A. Undersized wiring or a fuse below 10 A will trip during heat-up, while a fuse above 16 A removes the intended protection.
Voltage may deviate ±10% only if the frequency tolerance does not run in the opposite direction. A supply that is simultaneously high-voltage and low-frequency (or vice versa) is outside the permissible envelope and can stress the compressor/electronics.
External branch circuit protection according to UL 489 is mandatory. Three-phase devices below 63 A arrive with cable but without a plug, so the customer must supply a suitable mating plug and complete the connection on site.
The achievable range depends on the cooling source: 25–100 °C without cooling, 20–100 °C with water cooling, 15–100 °C with a refrigerator. Selecting the unit without planning the cooling method limits the process to 25 °C and above.
The transparent bath is polycarbonate and rated to a maximum of 100 °C. Operation above this temperature risks deformation or failure of the bath; higher-temperature processes require a different bath material.
Bath volume is 7.5 l with a minimum filling capacity of 5.2 l. Underfilling can cause pump cavitation and unstable temperature control; overfilling risks overflow when the fluid expands at 100 °C.
When the optional Huber pump adapter is fitted, polyglycol is not permissible as heat transfer fluid. Fluid selection must therefore be decided together with the pump-connection option, not after commissioning.
The IP20 rating means no protection against water. The unit must be placed in a dry indoor location, and ambient temperature must stay between 5 °C and 40 °C; condensation or splash near the bath will enter the enclosure.
Because the cut-out is adjustable, its setting can drift or be changed by operators. Maintenance should include a functional check of the overtemperature protection in addition to fluid inspection, so that the safety chain remains valid over the unit's life.
Selection & Integration Q&A
Pick the CC-110A when the process needs a 7.5 l transparent bath and up to 2.1 kW heating with ±0.02 K stability. Smaller siblings suit lower volumes; the CC-112A is for larger baths. All share the 100 °C polycarbonate ceiling.
Without cooling the unit is limited to 25–100 °C; water cooling extends it to 20–100 °C and a refrigerator to 15–100 °C. The external loop must handle 27 l/min at 0.7 bar on the pressure side, and polyglycol is not allowed with the Huber pump adapter.
A 200–240 V, 1~ or 2~, 50/60 Hz supply with max. 10 A draw, fused between 10 A and 16 A, plus UL 489 branch protection. Voltage tolerance is ±10% only if frequency does not drift in the opposite direction; three-phase units arrive without a plug.
The base unit excludes the M16x1 AG pump connection (option #19606) and any mating plug for three-phase supply. Maintenance must include checking the adjustable DIN 12876 overtemperature cut-out and the heat transfer fluid; no cables or consumables are included.