Peter Huber Kältemaschinenbau CC-118A
Peter Huber Kältemaschinenbau CC-118A — Thermostatic baths.
The CC-118A is a polycarbonate-bath thermostat with an 18-litre working volume and a Pilot ONE controller, available in three cooling configurations that set the lower temperature limit: 25 °C without cooling, 20 °C with water cooling, and 15 °C with a refrigerator. Its variable-speed pressure/suction pump delivers up to 27 l/min at 0.7 bar, making it suited to closed-loop temperature control of external circuits in laboratory processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection
| Model | Controller | Interface | Max delivery flow rate, liter/minute | Pump connection | Weight, kg | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| CC-118A Current | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male (Option, # | 8 | 5–40 |
| CC-104A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-106A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 5 | 5–40 |
| CC-108A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 6 | 5–40 |
| CC-110A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-112A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 104A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male | 5 | 5–40 |
| KISS 106A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 5 | 5–40 |
| KISS 108A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 110A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 112A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 118A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
Full specifications
| Bath volume | 18 l |
| Dimensions (w × D × H) | 333 x 520 x 335 mm |
| Maximum delivery flow rate | 27 l/min (pressure pump; 0, 7 bar) suction pump; 0, 4 bar: 22 l/min |
| Pump connection | M16x1 AG male (Option, #19606) (Option, #19606) |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 25...100 °C with cooling unit: 15...100 °C · with water cooling: 20...100 °C |
| Temperature stability mfr | ±0, 02 K |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Bath opening WxD/BD mfr | 275 x 321 / 150 mm (WxD/depth triplet) |
| Filling volume min. mfr | 12, 5 l (minimum) |
| Weight | 8 kg |
| Current consumption | 10 A (max.) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Open polycarbonate bath with a bath bridge that includes a bore hole for an immersion cooler (TC45–TC100E), allowing sub-ambient operation without a built-in refrigerator.
- Bath opening 275 × 321 mm, depth 150 mm; total bath volume 18 l with a minimum filling capacity of 12.5 l.
- Within the CC-1xxA series (CC-104A, CC-106A, CC-108A, CC-110A), the CC-118A is the largest bath volume listed; the smaller siblings share the same controller and pump architecture but offer reduced working volumes.
- Safety class III / FL per DIN 12876 with adjustable overtemperature protection.
Thermal behaviour and media
- The achievable temperature range is configuration-dependent: 25…100 °C as supplied, 20…100 °C with water cooling, and 15…100 °C with a refrigerator. The lower limit must be matched to the selected cooling method at the ordering stage.
- Temperature stability is ±0.02 K; heating capacity is 1.5–2.1 kW.
- The pressure pump delivers up to 27 l/min at 0.7 bar; the suction pump up to 22 l/min at 0.4 bar. Pump speed is adjustable, so external circuit pressure drop can be accommodated.
- Polyglycol must not be used as heat-transfer fluid when the Huber pump adapter is fitted.
Installation, control and operation
- Power supply: 200–240 V, 1~ or 2~, 50/60 Hz; maximum current 10 A. Branch protection must be 10 A minimum and 16 A maximum, with external UL 489 branch circuit protection required.
- Single- and two-phase units are delivered with a power cable and country-specific plug, which must be specified at ordering.
- Voltage tolerance ±10% is permitted only if the frequency tolerance does not run in the opposite direction.
- Controller: Pilot ONE with Ethernet, USB (host and device) and RS232 interfaces; an external Pt100 sensor can be connected.
- Ambient temperature for correct operation: 5…40 °C. Protection rating IP20. Class A, Group 1 per EN55011.
- Pump connection: M16×1 male thread (option #19606).
Integration & Operating Notes
The datasheet specifies a minimum fuse of 10 A and a maximum fuse of 16 A. Installing a breaker outside this window either fails to protect the unit or permits currents above its rating; external UL 489 branch circuit protection is mandatory.
A ±10% voltage tolerance is allowed only when the frequency tolerance does not run in the opposite direction. Supply quality should be verified before commissioning to avoid operating outside the permitted envelope.
The bath reaches 25 °C without auxiliary cooling, 20 °C with water cooling, and 15 °C with a refrigerator. The process minimum must be defined before ordering, because the cooling configuration cannot be changed in the field without additional hardware.
Operating below 12.5 l risks pump cavitation and unstable temperature control. The working fluid charge must be at least this volume, and the bath bridge bore hole for an immersion cooler must remain accessible.
When the Huber pump adapter is installed, polyglycol must not be used as heat-transfer fluid. Fluid selection must account for this restriction to avoid pump damage or safety compromise.
Single- and two-phase units are supplied with a power cable and a country-specific plug. The plug type must be stated on the order; otherwise the unit may arrive with a plug unsuitable for the site.
The controller accepts an external Pt100 sensor, enabling temperature control at a point outside the bath. This is useful when the process temperature must be regulated at the external circuit rather than in the bath.
Selection & Integration Q&A
Choose the CC-118A when the application needs the largest bath volume in the series (18 l, minimum fill 12.5 l) or the 275 × 321 mm opening. The smaller siblings suit lower-volume work; all share the Pilot ONE controller and variable-speed pump.
External circuits must tolerate up to 27 l/min at 0.7 bar from the pressure pump or 22 l/min at 0.4 bar from the suction pump. The pump connection is M16×1 male (option #19606); polyglycol is not permitted with the Huber pump adapter.
Supply 200–240 V, 1~ or 2~, 50/60 Hz, max. 10 A, with branch protection between 10 A and 16 A and external UL 489 protection. Ambient must stay 5–40 °C; IP20 and Class A Group 1 per EN55011 apply.
The country-specific plug must be specified at ordering. The M16×1 pump connection is an option (#19606). An immersion cooler (TC45–TC100E) or refrigerator is required for operation below 25 °C and is not part of the base unit.