Peter Huber Kältemaschinenbau CC-220B
Peter Huber Kältemaschinenbau CC-220B — Thermostatic baths.
The CC-220B is a 17-litre stainless-steel bath with a Pilot ONE immersion circulator, rated for 25...200 °C without cooling, 20...200 °C with water cooling and -30...200 °C with a refrigerator. Its variable-speed pressure/suction pump delivers up to 27 l/min at 0.7 bar and 22 l/min at 0.4 bar. It serves as a precision thermostating source for external circuits and open-bath work at ±0.02 K stability.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection
| Model | Controller | Interface | Pump connection | Temperature range, °C | Weight, kg | Max delivery flow rate, liter/minute |
|---|---|---|---|---|---|---|
| CC-220B Current | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 14 | 27 |
| CC-202C | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 22 |
| CC-205B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 45–200 | 9 | 27 |
| CC-208B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 10 | 27 |
| CC-212B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 11 | 22 |
| CC-215B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male | 25–200 | 12 | — |
| CC-225B | Pilot ONE | Ethernet, USB (Host u. D | M16x1 AG male (Option, # | 25–200 | 16 | 27 |
| KISS 202C | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 8 | 14 |
| KISS 205B Badumwälzthermostat mit Edelstahlbad | KISS | USB (Device), RS232 Inte | M16x1 AG male | 45–200 | 9 | 14 |
| KISS 208B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 10 | 14 |
| KISS 212B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 11 | 14 |
| KISS 215B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 12 | 14 |
| KISS 220B | KISS | USB (Device), RS232 Inte | M16x1 AG male (Option, # | 25–200 | 14 | 14 |
| KISS 225B | KISS | USB (Device), RS232 Inte | M16x1 AG male | 25–200 | 16 | 14 |
Full specifications
| Bath volume | 17 l |
| Dimensions (w × D × H) | 350 x 555 x 375 mm |
| Filling volume | 12 l (minimum) |
| Maximum delivery flow rate | 27 l/min (pressure pump; 0, 7 bar) suction pump; 0, 4 bar: 22 l/min |
| Pump connection | M16x1 AG male (Option, #19607) (Option, #19607) |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 25...200 °C with cooling unit: -30...200 °C · with water cooling: 20...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Bath opening WxD/BD mfr | 290 x 329 / 150 mm (WxD/depth) |
| Weight | 14 kg |
| Current consumption | 10 A (max.) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Open stainless-steel bath, 17 l volume, 12 l minimum filling, bath opening 290 x 329 mm with 150 mm depth; external envelope 350 x 555 x 375 mm, 14 kg.
- Bath bridge carries a hole for a cooling probe (e.g. immersion cooler TC45–TC100E), so sub-ambient or near-ambient operation is added by accessory rather than built in.
- Within the CC-2xxB series the 220B is the largest bath of the group (siblings CC-202C, CC-205B, CC-208B, CC-212B); the larger volume buys thermal inertia and space for immersed samples or external-loop buffer, not higher pump head.
Thermal behaviour and media
- Base range 25...200 °C; water cooling extends the low end to 20 °C; a refrigerator extends it to -30 °C. The achievable low temperature is therefore a function of the cooling option fitted, not of the circulator alone.
- Heating capacity 1.5–2.1 kW and stability ±0.02 K; the Pilot ONE controller accepts an external Pt100 sensor for control at the external load rather than in the bath.
- Pump is a variable-speed pressure-and-suction type: 27 l/min at 0.7 bar pressure side, 22 l/min at 0.4 bar suction side. Pressure and suction figures belong to their respective sides and must not be interchanged when sizing an external circuit.
Installation, control and operation
- Supply 200–240 V, 1~ or 2~, 50/60 Hz, max. current 10 A; branch fusing min. 10 A, max. 16 A, with external branch protection to UL 489. Voltage tolerance ±10%, provided frequency tolerance does not run in the opposite direction.
- Single/two-phase units ship with power cable and country-specific plug, which must be specified at order. Class A, Group 1 per EN55011; IP20, so dry location only.
- Interfaces: Ethernet, USB host and device, RS232; external Pt100 sensor connection; safety class III/FL.
- Ambient 5...40 °C. Optional M16x1 male pump connection (#19607); with the Huber pump adapter, polyglycol is not permissible as heat transfer fluid.
Integration & Operating Notes
The datasheet specifies a minimum fuse of 10 A and a maximum of 16 A with a maximum current draw of 10 A, and requires external branch circuit protection to UL 489. Oversizing the breaker beyond 16 A leaves the unit without the protection the manufacturer assumes; undersizing causes nuisance trips at run-up.
Supply may deviate ±10% in voltage only as long as the frequency tolerance does not run in the opposite direction. A site with low voltage and high frequency simultaneously falls outside the permitted envelope even though each deviation alone would be acceptable, so site power quality should be measured before commissioning.
With the Huber pump adapter installed, polyglycol is not permissible as heat transfer fluid. Fluid selection must therefore be fixed together with the pump-connection decision (optional M16x1 male, #19607), because changing one later invalidates the other and may require draining and flushing the bath and external loop.
Degree of protection is IP20, so the unit has no water-ingress protection and must be sited dry, away from dripping or splashing liquid and from filling or draining operations of adjacent equipment. Ambient must stay within 5...40 °C; the 14 kg mass and 350 x 555 x 375 mm footprint define the bench space and lifting plan.
A drain valve with cap (#6839) is an accessory for emptying the bath; without it, fluid removal at change intervals means lifting or siphoning a 12–17 l charge. Decide on the drain valve at order time, since retrofitting after commissioning requires draining the bath anyway.
Selection & Integration Q&A
Pick the 220B when the 17 l volume and 290 x 329 mm opening are needed for immersed samples or as buffer for an external loop; the smaller CC-202C/205B/208B/212B siblings suit smaller loads. Pump head and temperature range are not the differentiator, so size on volume and opening first.
A 200–240 V, 1~ or 2~, 50/60 Hz supply with a 10–16 A branch fuse and external protection to UL 489, within ±10% voltage and frequency tolerances that do not oppose. Specify the country-specific plug at order, since single/two-phase units ship with cable and plug only on request.
Yes. The Pilot ONE controller accepts an external Pt100 sensor, so the control point can be placed at the load. The variable-speed pressure/suction pump then has to be set against the external circuit's pressure drop, using 27 l/min at 0.7 bar on the pressure side and 22 l/min at 0.4 bar on the suction side as the limits.
The M16x1 male pump connection (#19607), the drain valve with cap (#6839) and any immersion cooler (TC45–TC100E) for sub-ambient work are options, as is the refrigerator for -30 °C. Heat transfer fluid is not supplied; if the Huber pump adapter is used, polyglycol is excluded.