Peter Huber Kältemaschinenbau CS 75 by Van der Heijden
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CS 75 by Van der Heijden

Peter Huber Kältemaschinenbau CS 75 by Van der Heijden — Recirculating chillers with Noise level 66 dB.

The CS 75 is a 37-litre-bath recirculating chiller in the Cs series, operating from -20 to +15 °C on a 400 V three-phase supply. It is the smallest sibling, distinguished by a 70 l/min pump and G3/4 male connection. It serves low-temperature process cooling with a 50:50 water-glycol fluid.

Noise level
66 dB

Peter Huber Kältemaschinenbau Cs Chiller Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CMinimum fill volume, m³Power supplyPump connectionTemperature range, °C
CS 75 by Van der Heijden Current 665–320.16400V 3~ 50HzG3/4 AG male-20–15
CS 100 by Van der Heijden 705–320.16400V 3~ 50HzG3/4 AG male-20–15
CS 100-H by Van der Heijden 715–320.175400V 3~ 50HzG3/4 AG male-20–85
CS 200 by Van der Heijden 785–320.225400V 3~ 50HzG1 AG male-20–15
CS 200-H by Van der Heijden 785–320.225400V 3~ 50HzG1 AG male-20–85
CS 20-H by Van der Heijden 575–320.024230V 1~ 50HzG3/8 AG male-20–85
CS 25 by Van der Heijden 605–320.024230V 1~ 50HzG1/2 AG male-20–15
CS 50 by Van der Heijden 635–320.1400V 3~ 50HzG3/4 AG male-20–15
Huber CS 200-H 785–320.225460V 3~ 60HzG1 AG male-20–85
Huber CS 25 605–320.024208V 2~ 60HzG1/2 AG male-20–15
Huber CS 250HT OD 78-20–500.16460V 3~ 60HzG3/4 AG male10–22
Huber CS 330-H OD 78-20–500.16460V 3~ 60HzG1 AG male-20–85
Huber CS 550-H OD 79-20–500.16460V 3~ 60HzG1 AG male-20–85
Huber CS 550 OD 79-20–500.16460V 3~ 60HzG1 AG male-20–15
Huber LCS 80 77-20–500.04460V 3~ 60HzG1 AG male-45–15

Full specifications

Core Performance & Functional Specifications
Noise level 66 dB(A)
Bath volume 37 l
Maximum delivery flow rate 70 l/min (3, 7 bar)
Minimum fill volume 160 l
Pump connection G3/4 AG male
Temperature range -20...15 °C
Dimensions, Weight & Mechanical Design
Weight 265 kg
Electrical, Power & Drive Specifications
Current consumption 14 A
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Digital interface RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...32 °C
Protection rating IP52
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

The CS 75 is an air-cooled recirculating chiller with a 37 l integrated bath and a minimum system fill of 160 l. Within the Cs series it is the smallest model; the CS 100, CS 100-H, CS 125 and CS 125-H are larger siblings with higher noise levels (70–75 dB(A) vs 66 dB(A) here). The unit uses refrigerant R-449A (A1, H280) with a CO2 equivalent of 6.29 t. The pump delivers up to 70 l/min through a G3/4 male connection, and flow can be trimmed via a manual bypass valve on the rear panel.

Operating envelope and media

Electrical and control

Installation notes

Voltage and frequency tolerances must be respected simultaneously; combinations such as -5% voltage with +2% frequency are not allowed. Space requirements around the air-cooled condenser must be observed per the operating conditions.

Integration & Operating Notes

Electrical
Do not combine -5% voltage with +2% frequency

Voltage and frequency tolerances must be respected simultaneously. Some combinations, such as -5% voltage with +2% frequency, are explicitly not allowed. Verify the site supply or generator against the permissible envelope before connection to avoid operating outside the unit's design limits.

Technical Datasheet, page 1
Process Limits
Cooling capacity is rated at 20 °C room temperature

Output data are valid for a room temperature of 20 °C. If the ambient temperature rises, the cooling capacity may drop. Size the heat load and plant-room ventilation with this derating in mind, and keep the ambient within the 5–32 °C range.

Technical Datasheet, page 1
Cooling & Media
Use 50:50 water–monoethylene glycol for the -20 °C range

The recommended thermofluid is a 50:50 water–monoethylene glycol mixture, which is necessary to reach the lower end of the -20 to +15 °C range without freezing. The system fill must be at least 160 l; the integrated bath holds 37 l.

Technical Datasheet, page 1–2
Mounting
Respect condenser clearance and IP52 protection limits

Space requirements around the air-cooled unit must be observed per the operating conditions. The IP52 rating protects against dust and dripping water but not water jets or immersion, so washdown cleaning is not permitted. Ambient must stay between 5 and 32 °C.

Technical Datasheet, page 1–2
Control & I/O
RS232 interface and rear bypass valve for flow trimming

The unit provides an RS232 interface for remote monitoring or control. Flow rate can be adjusted via the manual bypass valve on the back of the chiller to match pump delivery to the application, avoiding excessive flow or pressure in the external circuit.

Technical Datasheet, page 1

Selection & Integration Q&A

When should I choose the CS 75 rather than a larger sibling such as the CS 100 or CS 125?

Pick the CS 75 for the lowest noise (66 dB(A)) and smallest footprint in the series when the heat load fits its capacity. The CS 100/125 and -H variants offer more capacity but run at 70–75 dB(A); choose them only if the CS 75's cooling output is insufficient.

What hydraulic and electrical interfaces must the adjacent system provide?

The external circuit must mate with a G3/4 male pump connection and accommodate up to 70 l/min. Electrically, provide a 400 V 3~ 50 Hz supply fused at 3×20 A with a site-fitted plug; the unit ships with cable but no plug. RS232 is available for control.

What ambient and installation limits apply to the CS 75?

Operate only between 5 and 32 °C ambient, with condenser clearance per the operating conditions. IP52 allows dust and dripping water but not jets or immersion. Cooling capacity is rated at 20 °C room temperature and derates as ambient rises.

What is not included in base delivery and what fluid must I supply?

The unit is delivered with a power cable but without a plug, so a suitable three-phase plug must be sourced. The heat-transfer fluid is not included; prepare a 50:50 water–monoethylene glycol mixture and allow for a minimum system fill of 160 l.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CS 75 Circulating Bath — Technical Datasheet
Datasheet ·198,0 КБ ·2 pp.
View PDF