Peter Huber Kältemaschinenbau Grande Fleur Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Grande Fleur Dynamisches

Peter Huber Kältemaschinenbau Grande Fleur Dynamisches — Temperature control units with Noise level 61 dB.

The Grande Fleur Dynamisches is an air-cooled, variable-speed pressure-pump circulator with a -40 to 200 °C working range and 0.6 kW cooling capacity down to 0 °C, falling to 0.04 kW at -40 °C. It is the base Dynamisches variant of the Grande Fleur series, distinguished from the -eo and w siblings by its configuration rather than by a different temperature envelope. It serves as a temperature-control source for external circuits requiring 47 l/min maximum delivery through an M24x1.5 male pump connection.

Noise level
61 dB

Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionPower supplyPump connectionPump typeTemperature range, °C
Grande Fleur Dynamisches Current ——————
Grande Fleur w Dynamisches —0, 01208-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -40–200
Grande Fleur w-eo Dynamisches ——————
Petite Fleur-eo Dynamisches 580, 01 K220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Petite Fleur w Dynamisches —0, 01220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Unistat 410w Dynamisches 610, 01 K220-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -45–250
Unistat 425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat 430w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P405w Dynamisches —0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st-45–250
Unistat P425 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430w Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250

Full specifications

Description

Construction and series positioning

Behavior across the working range

Installation, connection and control

Integration & Operating Notes

Process Limits
Cooling capacity at -40 °C is only 0.04 kW

Near the lower temperature limit the available cooling capacity falls from 0.6 kW at 0 °C to 0.2 kW at -30 °C and 0.04 kW at -40 °C. Low-temperature processes must be sized against these figures, not the nominal 0.6 kW, or the unit will be unable to pull down or hold setpoint under load.

Grande Fleur Specifications, page 1
Electrical
Do not combine -5% voltage with +2% frequency

Voltage tolerance is ±5% only when frequency tolerance is simultaneously ±2%, and the specific combination of -5% voltage and +2% frequency is not permitted. Supply quality must be checked at the installation point before commissioning, since an out-of-envelope supply invalidates the rating.

Grande Fleur Specifications, page 2
Electrical
Specify country-specific plug at order

Single- and two-phase units ship with a power cable and a country-specific plug that must be stated when ordering. The 208–240 V, 1~/2~, 50/60 Hz supply, 11.5 A maximum current and 16 A fuse define the site circuit; a missing plug specification delays commissioning.

Grande Fleur Specifications, page 2
Mounting
Air-cooled condenser needs ventilated, dry IP20 location

The refrigeration machine is air-cooled, so the installation site must provide airflow and clearance for heat rejection. The IP20 enclosure is not water-protected and is unsuitable for wet or washdown areas; ambient must remain within 5–40 °C, and cooling capacity may drop as ambient temperature rises.

Grande Fleur Specifications, page 2
Process Media
Heat-transfer fluid viscosity must stay at or below 50 mm²/s

The maximum permissible kinematic viscosity is 50 mm²/s. At low temperatures many fluids thicken beyond this, overloading the pump and reducing flow; fluid selection must be checked against the coldest operating point, not only at room temperature.

Grande Fleur Specifications, page 2
Process Media
Charge at least 1.5 l with 3 l expansion volume

Minimum filling capacity is 1.5 l and the expansion tank holds 3 l. The charge must cover the external circuit plus the unit's own volume while leaving headroom for thermal expansion across the -40 to 200 °C range, otherwise underfilling or overflow occurs.

Grande Fleur Specifications, page 2
Process Media
Mating hoses must match M24x1.5 male thread

The pump connection is M24x1.5 male, so flexible tempering hoses or adapters on the external circuit must carry the matching female thread. Mismatched fittings cannot be adapted at the port without a correct adapter, and the 47 l/min maximum delivery sets the hose and circuit pressure-drop budget.

Grande Fleur Specifications, page 2
Compliance
R-290 flammable refrigerant drives site safety planning

The first stage uses R-290 (A3, H220), a flammable natural refrigerant with GWP 0.02 and UN number 3358. Installation and safety planning must account for the A3 classification; the unit is also EN55011 Class A, Group 1, so EMC measures may be needed in sensitive or residential environments.

Grande Fleur Specifications, pages 1–2

Selection & Integration Q&A

When should I choose the Grande Fleur Dynamisches rather than the -eo, w or w-eo siblings, or the Petite Fleur?

Pick the base Dynamisches when the standard 47 l/min variable-speed pressure pump and the -40 to 200 °C range fit the duty. The -eo shares the 47 l/min flow; the w and w-eo are separate configurations to be evaluated on their own data. Choose the Petite Fleur Dynamisches only if 58 dB(A) acoustics outweigh the smaller platform.

What must the external circuit and its connections provide to mate with this unit?

The pump port is M24x1.5 male, so hoses or adapters need the matching female thread. The circuit plus unit must take at least 1.5 l charge with 3 l expansion volume, and the fluid must stay at or below 50 mm²/s viscosity at the coldest setpoint. Maximum delivery is 47 l/min at 0.9 bar pump pressure.

What site conditions and supply quality does the installation require?

Dry, protected location: IP20 and air-cooled condenser need ventilation, ambient 5–40 °C. Supply 208–240 V, 1~/2~, 50/60 Hz, 11.5 A max, 16 A fuse; ±5% voltage only with ±2% frequency, and -5% voltage with +2% frequency is not allowed. Specify the country-specific plug at order.

What is not included in base delivery and what must be planned for commissioning and service?

The power cable and country-specific plug are supplied only if the plug type is specified at order. Heat-transfer fluid is not part of the unit charge and must be selected within the 50 mm²/s viscosity limit. External Pt100 sensors and circuit hoses/adapters for the M24x1.5 port are separate items to be provided.

Dimensional Drawings & Technical Documentation