Peter Huber Kältemaschinenbau Huber CS 200-H
Peter Huber Kältemaschinenbau Huber CS 200-H — Recirculating chillers with Noise level 78 dB.
The Huber CS 200-H is a refrigerated circulating bath with a 37 L internal bath and a 225–245 L system fill, operating from -20 to 85 °C. It is the largest of the Cs Chiller series, distinguished by its 460 V 3~ 60 Hz supply and 49 A draw. It serves as a central temperature-control unit for closed-loop processes requiring high flow (125 L/min at 4.7 bar).
Peter Huber Kältemaschinenbau Cs Chiller Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Minimum fill volume, m³ | Power supply | Pump connection | Temperature range, °C |
|---|---|---|---|---|---|---|
| Huber CS 200-H Current | 78 | 5–32 | 0.225 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| CS 100 by Van der Heijden | 70 | 5–32 | 0.16 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| CS 100-H by Van der Heijden | 71 | 5–32 | 0.175 | 400V 3~ 50Hz | G3/4 AG male | -20–85 |
| CS 200 by Van der Heijden | 78 | 5–32 | 0.225 | 400V 3~ 50Hz | G1 AG male | -20–15 |
| CS 200-H by Van der Heijden | 78 | 5–32 | 0.225 | 400V 3~ 50Hz | G1 AG male | -20–85 |
| CS 20-H by Van der Heijden | 57 | 5–32 | 0.024 | 230V 1~ 50Hz | G3/8 AG male | -20–85 |
| CS 25 by Van der Heijden | 60 | 5–32 | 0.024 | 230V 1~ 50Hz | G1/2 AG male | -20–15 |
| CS 50 by Van der Heijden | 63 | 5–32 | 0.1 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| CS 75 by Van der Heijden | 66 | 5–32 | 0.16 | 400V 3~ 50Hz | G3/4 AG male | -20–15 |
| Huber CS 25 | 60 | 5–32 | 0.024 | 208V 2~ 60Hz | G1/2 AG male | -20–15 |
| Huber CS 250HT OD | 78 | -20–50 | 0.16 | 460V 3~ 60Hz | G3/4 AG male | 10–22 |
| Huber CS 330-H OD | 78 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| Huber CS 550-H OD | 79 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–85 |
| Huber CS 550 OD | 79 | -20–50 | 0.16 | 460V 3~ 60Hz | G1 AG male | -20–15 |
| Huber LCS 80 | 77 | -20–50 | 0.04 | 460V 3~ 60Hz | G1 AG male | -45–15 |
Full specifications
| Noise level | 78 dB(A) |
| Bath volume | 37 l |
| Minimum fill volume | 225 l |
| Pump connection | G1 AG male |
| Temperature range | -20...85 °C |
| Weight | 535 kg |
| Current consumption | 49 A |
| Power supply | 460V 3~ 60Hz |
| Digital interface | RS232 |
| Ambient temperature limit | 5...32 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
The CS 200-H is a refrigerated circulating chiller with an integrated 37 L bath and a total system fill of 225–245 L. Within the Cs Chiller series, it is the highest-capacity model; siblings CS 100, CS 100-H, CS 125 and CS 125-H are smaller and quieter (70–75 dB(A) vs. 78 dB(A) for this unit). The -H suffix denotes the high-temperature variant, extending the upper limit to 85 °C. The unit is supplied with a 460 V 3~ 60 Hz power supply, draws 49 A, and requires a 3×60 A fuse. Protection rating is IP20, and the pump connection is G1 AG male.
Behavior under load and media
Cooling capacity is rated at 20 °C ambient and derates as ambient temperature rises; selection must account for actual room conditions. The circulation pump delivers up to 125 L/min at 4.7 bar maximum delivery pressure, and flow can be adjusted via a manual bypass valve on the backside. The recommended thermofluid is a 50:50 water–monoethylene glycol mixture, suitable for the full -20 to 85 °C range. The refrigerant is R-449A (A1, H280) with a CO2 equivalent of 16.76 t from 12 kg charge, triggering leak-check and reporting obligations.
Installation, control and operation
The unit must be installed in an ambient temperature of 5–32 °C with sufficient clearance for ventilation; space requirements are referenced online. Electrical supply tolerances are interdependent: a -5% voltage with +2% frequency is not allowed. Three-phase devices under 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site. The interface is RS232 for remote monitoring or control. The equipment emits Class A, Group 1 disturbances per EN55011, so additional EMC measures may be needed in residential or sensitive environments.
Integration & Operating Notes
The unit is delivered with a cable but without a plug. The installer must provide a suitable 460 V 3~ 60 Hz plug and ensure the 3×60 A fuse protection is in place.
Operating outside this range may reduce cooling capacity or cause malfunction. Room climate control or ventilation must be planned, especially since cooling capacity is rated at 20 °C ambient and drops at higher temperatures.
The recommended fluid is a 50:50 mixture of water and monoethylene glycol. Using other fluids may affect heat transfer, pump performance, and freeze protection across the -20 to 85 °C range.
The manual bypass valve on the backside allows matching circulation to the application. Adjust to meet process requirements while avoiding pump overload; ensure the connected circuit's pressure drop does not exceed the pump's 4.7 bar maximum delivery pressure.
The unit contains 12 kg of R-449A with a CO2 equivalent of 16.76 t. This triggers periodic leak-check and reporting obligations under applicable regulations. Personnel must be aware of the A1, H280 hazard and follow ventilation precautions.
The RS232 interface allows integration with external control or data-logging systems. Plan for cable routing and protocol compatibility with the supervisory system.
The equipment emits Class A, Group 1 disturbances per EN55011. In residential or sensitive environments, additional EMC measures may be necessary to avoid interference with other equipment.
Selection & Integration Q&A
Choose the CS 200-H when you need the largest bath (37 L) and system fill (225–245 L) in the series, or when the 85 °C upper limit and 460 V 3~ 60 Hz supply match your infrastructure. The CS 100/125 models are smaller and quieter (70–75 dB(A) vs. 78 dB(A)) but have lower capacity.
It requires a 460 V 3~ 60 Hz supply with a 3×60 A fuse and draws 49 A. The unit is delivered with cable but without a plug, so a suitable plug must be provided. Voltage and frequency tolerances are interdependent: -5% voltage with +2% frequency is not allowed.
Install in an ambient temperature of 5–32 °C with sufficient clearance for ventilation. Cooling capacity is rated at 20 °C ambient and derates at higher temperatures. Respect space requirements; operating conditions are available online. Class A EMC disturbances may require filtering in sensitive environments.
The base delivery includes the chiller with cable but no plug. The recommended thermofluid is a 50:50 water–monoethylene glycol mixture, which must be supplied separately. Refrigerant R-449A (12 kg, CO2 equivalent 16.76 t) requires periodic leak checks and reporting. No other consumables or accessories are specified in the datasheet.