Peter Huber Kältemaschinenbau Huber CS 330-H OD
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Huber CS 330-H OD

Peter Huber Kältemaschinenbau Huber CS 330-H OD — Recirculating chillers with Noise level 78 dB.

The Huber CS 330-H OD is a 37-litre-bath recirculating chiller with an 8 kW heater, rated for -20 to +85 °C and outdoor installation down to -20 °C ambient. It is distinguished within the Cs series by its 460 V 3-phase 60 Hz supply, 49 A max current and 125 l/min pump flow, serving as a high-capacity temperature control unit for industrial loops requiring a 160–175 l fluid inventory.

Noise level
78 dB

Peter Huber Kältemaschinenbau Cs Chiller Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CMinimum fill volume, m³Power supplyPump connectionTemperature range, °C
Huber CS 330-H OD Current 78-20–500.16460V 3~ 60HzG1 AG male-20–85
CS 100 by Van der Heijden 705–320.16400V 3~ 50HzG3/4 AG male-20–15
CS 100-H by Van der Heijden 715–320.175400V 3~ 50HzG3/4 AG male-20–85
CS 200 by Van der Heijden 785–320.225400V 3~ 50HzG1 AG male-20–15
CS 200-H by Van der Heijden 785–320.225400V 3~ 50HzG1 AG male-20–85
CS 20-H by Van der Heijden 575–320.024230V 1~ 50HzG3/8 AG male-20–85
CS 25 by Van der Heijden 605–320.024230V 1~ 50HzG1/2 AG male-20–15
CS 50 by Van der Heijden 635–320.1400V 3~ 50HzG3/4 AG male-20–15
CS 75 by Van der Heijden 665–320.16400V 3~ 50HzG3/4 AG male-20–15
Huber CS 200-H 785–320.225460V 3~ 60HzG1 AG male-20–85
Huber CS 25 605–320.024208V 2~ 60HzG1/2 AG male-20–15
Huber CS 250HT OD 78-20–500.16460V 3~ 60HzG3/4 AG male10–22
Huber CS 550-H OD 79-20–500.16460V 3~ 60HzG1 AG male-20–85
Huber CS 550 OD 79-20–500.16460V 3~ 60HzG1 AG male-20–15
Huber LCS 80 77-20–500.04460V 3~ 60HzG1 AG male-45–15

Full specifications

Core Performance & Functional Specifications
Noise level 78 dB(A)
Bath volume 37 l
Heating capacity 8 kW
Maximum delivery flow rate 125 l/min (also 4, 7 bar pump pressure)
Minimum fill volume 160 l
Pump connection G1 AG male
Temperature range -20...85 °C
Dimensions, Weight & Mechanical Design
Weight 530 kg
Electrical, Power & Drive Specifications
Current consumption 49 A (max.)
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Digital interface RS232
Operating Conditions, Environment & Protection
Ambient temperature limit -20...50 °C
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

The CS 330-H OD is a recirculating chiller with a 37 l internal bath and a minimum filling capacity of 160 l for the external loop. It is the largest and only outdoor-rated variant among the listed Cs siblings (CS 100, CS 100-H, CS 125, CS 125-H), which are indoor units with lower noise levels (70–75 dB(A)) and smaller capacities. The CS 330-H OD is supplied with a remote operating panel on a 20 m cable, allowing the control interface to be mounted away from the chiller. The electrical components are rated IP54, and the unit is EMC Class A Group 1 for industrial environments.

Working range and media

Installation, connection and control

Integration & Operating Notes

Electrical
Supply must be 460 V 3-phase 60 Hz with 3x60 A fusing

The unit draws up to 49 A per phase and requires 3x60 A fusing. The site electrical feed, cable cross-section and breaker rating must be sized accordingly before delivery. Three-phase devices below 63 A are supplied with cable but without a plug, so the customer must provide the site-side connection.

Huber CS 330-H OD Technical Data, page 2
Process Limits
Cooling capacity derates above +35 °C ambient

Cooling capacity is only guaranteed up to +35 °C ambient temperature and reduces above that. Sizing must use the capacity at the actual site ambient temperature, especially for outdoor installations in summer or in hot plant rooms.

Huber CS 330-H OD Technical Data, page 1
Mounting
Outdoor installation down to -20 °C requires frost protection

The chiller is suitable for outside mounting with a minimum environmental temperature of -20 °C. Site planning must account for weather exposure and frost protection of the loop, including the recommended 50:50 water–monoethylene glycol mixture.

Huber CS 330-H OD Technical Data, page 1
Control & I/O
Remote operating panel on 20 m cable needs routing

The operating panel is delivered as a remote control connected via a 20 m cable. Cable route and panel mounting location must be planned separately from the chiller, and the RS232 interface is available for communication.

Huber CS 330-H OD Technical Data, page 1
Process Limits
External loop must be sized for 125 l/min and 4.7 bar

The pump connection is G1 male with a maximum delivery flow rate of 125 l/min and maximum pump pressure of 4.7 bar. The external circuit, including piping and fittings, must be sized for these limits and for the minimum filling capacity of 160 l.

Huber CS 330-H OD Technical Data, page 2
Maintenance
Refrigerant R-449A requires F-gas reporting and leak checks

The refrigerant is R-449A (A1, H280) with a 9 kg charge and 12.57 t CO2 equivalent. Refrigerant handling, leak detection and CO2-equivalent reporting obligations apply under F-gas regulations.

Huber CS 330-H OD Technical Data, page 1

Selection & Integration Q&A

When should I choose the CS 330-H OD over the smaller CS 100 or CS 125 siblings?

Choose the CS 330-H OD when you need a 37 l bath, 8 kW heating, 125 l/min flow and outdoor installation down to -20 °C ambient. The CS 100/125 siblings are indoor units with lower noise (70–75 dB(A)) and smaller capacities, suited to indoor plant rooms with lower heat loads.

What are the requirements for the electrical supply and site connection of this chiller?

It requires a 460 V 3-phase 60 Hz supply with 49 A max current and 3x60 A fusing. Voltage and frequency tolerances are ±5 % and ±2 %, but not simultaneously in the unfavorable direction. The unit is supplied with cable but without a plug, so the site must provide the connection.

What are the mounting and ambient limits for outdoor installation?

The chiller is suitable for outdoor mounting down to -20 °C ambient, but cooling capacity is only guaranteed up to +35 °C and derates above that. Electrical components are IP54, and the unit is EMC Class A Group 1 for industrial environments. Frost protection of the loop is required.

What maintenance and accessories are needed beyond the base delivery?

The base delivery includes a remote operating panel on a 20 m cable but no plug. The recommended thermofluid is a 50:50 water–monoethylene glycol mixture, and the loop must be filled to at least 160 l. Refrigerant R-449A (9 kg, 12.57 t CO2e) requires F-gas leak checks and reporting.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Huber CS 330-H OD Technical Data
Datasheet ·114,7 КБ ·2 pp.
View PDF