Peter Huber Kältemaschinenbau Huber CS 550 OD
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Huber CS 550 OD

Peter Huber Kältemaschinenbau Huber CS 550 OD — Recirculating chillers with Noise level 79 dB.

The Huber CS 550 OD is a 460 V/60 Hz, three-phase outdoor-rated recirculating chiller with a -20…15 °C working range at 2 K stability. Its 160 l minimum fill and 125 l/min pump distinguish it from smaller Cs-series siblings, and it serves industrial process cooling where ambient conditions and refrigerant compliance are planned.

Noise level
79 dB

Peter Huber Kältemaschinenbau Cs Chiller Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CMinimum fill volume, m³Power supplyPump connectionTemperature range, °C
Huber CS 550 OD Current 79-20–500.16460V 3~ 60HzG1 AG male-20–15
CS 100 by Van der Heijden 705–320.16400V 3~ 50HzG3/4 AG male-20–15
CS 100-H by Van der Heijden 715–320.175400V 3~ 50HzG3/4 AG male-20–85
CS 200 by Van der Heijden 785–320.225400V 3~ 50HzG1 AG male-20–15
CS 200-H by Van der Heijden 785–320.225400V 3~ 50HzG1 AG male-20–85
CS 20-H by Van der Heijden 575–320.024230V 1~ 50HzG3/8 AG male-20–85
CS 25 by Van der Heijden 605–320.024230V 1~ 50HzG1/2 AG male-20–15
CS 50 by Van der Heijden 635–320.1400V 3~ 50HzG3/4 AG male-20–15
CS 75 by Van der Heijden 665–320.16400V 3~ 50HzG3/4 AG male-20–15
Huber CS 200-H 785–320.225460V 3~ 60HzG1 AG male-20–85
Huber CS 25 605–320.024208V 2~ 60HzG1/2 AG male-20–15
Huber CS 250HT OD 78-20–500.16460V 3~ 60HzG3/4 AG male10–22
Huber CS 330-H OD 78-20–500.16460V 3~ 60HzG1 AG male-20–85
Huber CS 550-H OD 79-20–500.16460V 3~ 60HzG1 AG male-20–85
Huber LCS 80 77-20–500.04460V 3~ 60HzG1 AG male-45–15

Full specifications

Core Performance & Functional Specifications
Noise level 79 dB(A) (sound pressure level)
Bath volume 37 l
Minimum fill volume 160 l
Pump connection G1 AG male
Temperature range -20...15 °C
Temperature stability 2 K
Dimensions, Weight & Mechanical Design
Weight 540 kg (net)
Electrical, Power & Drive Specifications
Current consumption 50 A (max.)
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Digital interface RS232
Materials, Media & Process Compatibility
Refrigerant R-449A (1st stage; A1, H280)
Refrigerant quantity 16 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit -20...50 °C (permissible)
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Working range and load behavior

Installation, control and operation

Integration & Operating Notes

Electrical
Site must provide plug and disconnect for 50 A three-phase feed

The chiller is supplied with cable but without a plug because its current consumption is below 63 A. The customer must provide a suitable plug and disconnecting means rated for the 460 V/60 Hz three-phase supply and the 50 A maximum current, and verify that the combined voltage/frequency tolerance rule is met at the point of connection.

Huber CS 550 OD Circulating Bath Datasheet, page 2, installation notes
Process Limits
External loop volume must reach 160 l minimum before operation

Although the bath itself is 37 l, the minimum filling capacity for the system is 160 l, with a maximum of 195 l and a warning threshold at 175 l. The external circuit and expansion volume must be sized within these limits; operating below the minimum risks pump cavitation and poor temperature control, while exceeding the maximum leaves insufficient expansion space.

Huber CS 550 OD Circulating Bath Datasheet, page 2, performance data
Cooling & Media
Use 50:50 water–monoethylene glycol to match freeze protection

The recommended thermofluid is a 50:50 water–monoethylene glycol mixture. This ratio defines the freeze protection and heat-transfer basis for the -20…15 °C range; using plain water or a different concentration can cause freezing, corrosion or reduced heat transfer at low temperatures.

Huber CS 550 OD Circulating Bath Datasheet, page 2, operation
Mounting
Outdoor installation must respect -20 °C ambient and IP54 limits

The unit is suitable for outside mounting down to -20 °C ambient, and its electrical components are IP54-rated. Installation planning must account for low ambient temperatures and possible freezing of heat-transfer fluid, and must not expose the electrical enclosure to conditions beyond IP54 (dust and water ingress).

Huber CS 550 OD Circulating Bath Datasheet, page 1, installation
Control & I/O
Remote panel cable length constrains control-room placement

The operating panel is connected via a 20 m cable, so the control point can be located away from the chiller but only within that cable reach. Plan cable routing and panel mounting accordingly; the RS232 interface provides serial communication for integration with a higher-level controller.

Huber CS 550 OD Circulating Bath Datasheet, page 1, installation
Maintenance
Refrigerant charge triggers leak-check and reporting obligations

The unit contains 16 kg of R-449A with a CO2 equivalent of 22.35 t. This charge level typically triggers periodic refrigerant leak checks and record keeping in many jurisdictions. Service procedures must account for R-449A as a pressurized gas (H280), and leak response plans should be in place.

Huber CS 550 OD Circulating Bath Datasheet, page 1, safety

Selection & Integration Q&A

When should I choose the CS 550 OD rather than a smaller Cs-series sibling such as the CS 125?

Choose the CS 550 OD when the external loop requires a minimum fill of 160 l and pump delivery up to 125 l/min at 4.7 bar, with a -20…15 °C range. Smaller siblings like the CS 125 have lower capacity and noise (75 dB(A) vs 79 dB(A)) and suit lower-volume circuits.

What site power provisions are needed for the CS 550 OD, and what cable/plug is included?

The unit requires 460 V/60 Hz three-phase with 50 A maximum current. It is supplied with cable but without a plug because current is below 63 A, so the site must provide a suitable plug and disconnect. Verify the combined voltage/frequency tolerance rule at the connection point.

Can the CS 550 OD be installed outdoors, and what ambient limits apply?

Yes, it is suitable for outside mounting down to -20 °C ambient, with electrical components rated IP54. Cooling capacity drops as ambient temperature rises, so selection must use the actual site ambient. Freeze protection for the heat-transfer fluid must be planned for low ambient conditions.

What maintenance and refrigerant obligations come with the CS 550 OD, and what is not included in base delivery?

The 16 kg R-449A charge (22.35 t CO2 equivalent) typically triggers periodic leak checks and record keeping. The base delivery includes a 20 m remote panel cable but no plug; the customer must provide the plug and disconnect. Recommended thermofluid (50:50 water–MEG) is not included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Huber CS 550 OD Circulating Bath Datasheet
Datasheet ·114,6 КБ ·2 pp.
View PDF