Peter Huber Kältemaschinenbau KISS 118A
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS 118A

Peter Huber Kältemaschinenbau KISS 118A — Thermostatic baths.

The KISS 118A is an immersion circulator with an 18-litre bath and a pressure/suction pump, delivering 15–100 °C with a refrigerator, 20–100 °C with water cooling, or 25–100 °C uncooled. Its 0.05 K stability at 70 °C and DIN 12876 class III/FL protection distinguish it within the KISS series for precise thermostatic control.

Direct drop-in replacements

Eppendorf
MixMate ®
53%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minutePump connectionWeight, kgAmbient temperature limit, °C
KISS 118A Current KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40
CC-104A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-106A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male55–40
CC-108A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #65–40
CC-110A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-112A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #85–40
CC-118A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male (Option, #85–40
KISS 104A KISSUSB (Device), RS232 Inte14M16x1 AG male55–40
KISS 106A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #55–40
KISS 108A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 110A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 112A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40

Full specifications

Core Performance & Functional Specifications
Alarm output optic, acoustic
Bath volume 18 l
Dimensions (w × D × H) 333 x 520 x 335 mm
Filling volume 11 l (min)
Maximum delivery flow rate 14 l/min (pressure pump; 0, 25 bar)
Pump connection M16x1 AG male (Option, #19606) (Option, #19606)
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 25...100 °C
with cooling unit: 15...100 °C · with water cooling: 20...100 °C
Temperature stability mfr ±0, 05 K
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump; 0, 17 bar)
Bath opening WxD/BD mfr 275 x 321 / 150 mm (WxD/depth)
Dimensions, Weight & Mechanical Design
Weight 8 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max)
Power supply 200-240V 1~/2~ 50/60Hz V; Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The KISS 118A is a bath-mounted immersion circulator with an 18-litre bath volume (minimum filling 11 l) and a pressure/suction pump that is not speed-variable. It differs from the CC-104A through CC-110A siblings by its larger bath and integrated pump configuration. The unit uses a KISS controller with an OLED display and provides USB (Device) and RS232 interfaces. The bath opening measures 275 × 321 mm with 150 mm depth, and the overall case is 333 × 520 × 335 mm at 8 kg.

Thermal behaviour and media

Temperature range depends on the cooling method: 25–100 °C without cooling, 20–100 °C with water cooling, and 15–100 °C with a refrigerator. Stability is ±0.05 K at 70 °C. Heating capacity is 1.5–2.1 kW. The pressure pump delivers up to 14 l/min at 0.25 bar, while the suction pump reaches 10.5 l/min at 0.17 bar. The plastic pump is not resistant to acetone or polyglycols; for freeze protection, water should be mixed with glysantine or ethylene glycol.

Installation and control

Power supply is 200–240 V, 1~/2~, 50/60 Hz with a maximum current of 10 A. External branch circuit protection per UL 489 is required. Voltage and frequency tolerances are ±10% only if they do not run in opposite directions. Single- and two-phase units are delivered with a power cable and country-specific plug (must be specified at ordering). The unit is IP20 and must be operated between 5 °C and 40 °C ambient. An optional factory-fitted Pt100 sensor connection (#10688) can display, not control, process temperature. The autostart function restarts after power failure, and adjustable overtemperature protection per DIN 12876 with safety class III/FL is provided.

Integration & Operating Notes

Electrical
External UL 489 branch protection is mandatory

The datasheet requires branch circuit protection according to UL 489. The installer must provide a listed breaker sized for the 10 A maximum current; omitting it violates the installation instructions and may void compliance.

KISS 118A Immersion Circulator Datasheet, page 2
Electrical
Voltage and frequency tolerances must not oppose

A ±10% voltage tolerance is permitted only if the frequency tolerance does not run in the opposite direction. For example, -10% voltage with +3% frequency is not allowed. Verify the site supply envelope before connection.

KISS 118A Immersion Circulator Datasheet, page 2
Process Media
Plastic pump incompatible with acetone and polyglycols

The standard plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). Using these media risks pump damage; select an alternative fluid or consult the factory for a resistant pump option.

KISS 118A Immersion Circulator Datasheet, page 2
Process Media
Use glycol mix for freeze protection below ambient

For operation near or below 15 °C, water should be mixed with glysantine or ethylene glycol to prevent freezing. Pure water may freeze and damage the bath or pump if the unit is used with a refrigerator.

KISS 118A Immersion Circulator Datasheet, page 2
Control & I/O
Pt100 sensor option is factory-fit only

The optional Pt100 connection (#10688) displays process temperature but does not control it, and is only available factory fitted. It cannot be retrofitted in the field, so specify it at ordering if external monitoring is needed.

KISS 118A Immersion Circulator Datasheet, page 1
Operation
Autostart restarts after power failure

The autostart function automatically restarts the unit after a power failure. For processes where unattended restart is hazardous, this behaviour must be considered; the datasheet lists it as a standard function.

KISS 118A Immersion Circulator Datasheet, page 1

Selection & Integration Q&A

When should I choose the KISS 118A instead of a CC-104A or CC-110A sibling?

Choose the KISS 118A when an 18-litre bath and a pressure/suction pump are needed. The CC-104A through CC-110A siblings have different bath sizes; select them for smaller volumes or different pump configurations. The 118A offers 15–100 °C with a refrigerator and 0.05 K stability.

What power supply and branch protection does the KISS 118A require?

It requires 200–240 V, 1~/2~, 50/60 Hz with a maximum current of 10 A. External branch circuit protection per UL 489 is mandatory. Single- and two-phase units come with a power cable and country-specific plug, which must be specified at ordering.

What maintenance and accessories are not included with the base unit?

The base delivery includes the circulator and, for single/two-phase units, a power cable with country-specific plug. The Pt100 sensor connection (#10688) is an optional factory-fitted extra. No maintenance schedule is specified in the datasheet; consult the manual for service intervals.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS 118A Immersion Circulator Datasheet
Datasheet ·116,6 КБ ·2 pp.
View PDF