Peter Huber Kältemaschinenbau KISS 202C
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS 202C

Peter Huber Kältemaschinenbau KISS 202C — Thermostatic baths.

The KISS 202C is a 3.5 l stainless-steel bath immersion circulator with a non-variable pressure/suction pump (14 l/min, 0.25 bar) and 1.5–2.1 kW heating. Its 45–200 °C base range extends to 20–200 °C with water cooling or −30–200 °C with a refrigerator, positioning it as the high-temperature member of the KISS series for precise (±0.05 K) external loop tempering.

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection

Model ControllerInterfacePump connectionTemperature range, °CWeight, kgMax delivery flow rate, liter/minute
KISS 202C Current KISSUSB (Device), RS232 InteM16x1 AG male45–200814
CC-202C Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200922
CC-205B Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200927
CC-208B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–2001027
CC-212B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001122
CC-215B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–20012—
CC-220B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001427
CC-225B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001627
KISS 205B Badumwälzthermostat mit Edelstahlbad KISSUSB (Device), RS232 InteM16x1 AG male45–200914
KISS 208B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001014
KISS 212B KISSUSB (Device), RS232 InteM16x1 AG male25–2001114
KISS 215B KISSUSB (Device), RS232 InteM16x1 AG male25–2001214
KISS 220B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001414
KISS 225B KISSUSB (Device), RS232 InteM16x1 AG male25–2001614

Full specifications

Core Performance & Functional Specifications
Filling volume 2 l
Maximum delivery flow rate 14 l/min (pressure pump; 0, 25 bar)
Pump connection M16x1 AG male
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 45...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 05 K
Alarm message mfr optic, acoustic
Dimensions (W x D x H) mfr 178 x 260 x 355 mm
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump; 0, 17 bar)
Bath opening WxD/BD mfr ø = 25 / 150
Bath volume mfr 3, 5 l
Dimensions, Weight & Mechanical Design
Weight 8 kg
Electrical, Power & Drive Specifications
Current consumption 10 A
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Installation, control and operation

Integration & Operating Notes

Electrical
Provide a 10 A branch with UL 489 breaker

The unit draws up to 10 A on a 200–240 V 1~/2~ 50/60 Hz supply and requires a minimum 10 A fuse plus external branch circuit protection according to UL 489; the panel must include a listed breaker, and the country-specific plug must be stated at order time.

KISS 202C Immersion Circulator Datasheet, Specifications / Installation
Process Limits
Cooling method sets the achievable low temperature

Without auxiliary cooling the unit operates only from 45 °C upward; reaching 20 °C requires a water-cooling loop and reaching −30 °C requires a refrigerator, so the required process low point dictates which cooling configuration must be specified.

KISS 202C Immersion Circulator Datasheet, Overview
Cooling & Media
Avoid acetone and polyglycols with the plastic pump

The non-variable plastic pump is not resistant to acetone or polyglycols (depending on manufacturer), so these heat transfer fluids must be excluded; for freeze protection water should be mixed with glysantine or ethylene glycol rather than used pure.

KISS 202C Immersion Circulator Datasheet, Operation
Control & I/O
Decide on autostart before unattended operation

The autostart function restarts the unit automatically after a power failure; for sensitive or unattended processes this behavior may be unacceptable and must be evaluated during commissioning alongside the adjustable overtemperature cut-out per DIN 12876 (safety class III/FL).

KISS 202C Immersion Circulator Datasheet, Safety
Mounting
IP20 enclosure demands a dry, protected location

With IP20 protection the unit must be installed away from water spray in a dry location, and ambient temperature must remain between 5 °C and 40 °C; the EN55011 Class A Group 1 classification means it is intended for industrial, not residential, electromagnetic environments.

KISS 202C Immersion Circulator Datasheet, Installation / Compliance

Selection & Integration Q&A

When should I choose the KISS 202C rather than a larger sibling such as the CC-208B or CC-212B?

Pick the 202C when a 3.5 l bath, 2 l filling and 25 mm opening suit the samples and 200 °C is the ceiling; move to a larger sibling when immersion depth or sample volume exceeds the 25 mm opening.

What must the external circuit and power installation provide for this unit?

The loop must be drivable by 14 l/min at 0.25 bar through an M16×1 male connection; the supply is 200–240 V 1~/2~ 50/60 Hz with a 10 A fuse and UL 489 branch protection, and the plug type must be ordered.

What environment and duty limits apply to installation?

Install in a dry, spray-free location at 5–40 °C ambient; IP20 and EN55011 Class A Group 1 restrict it to industrial settings, and the ±10% voltage tolerance applies only if frequency does not deviate oppositely.

What is not included in base delivery that I must plan for?

The country-specific power plug must be specified at order; water cooling or a refrigerator is required to go below 45 °C; and heat transfer fluid (water with glysantine or ethylene glycol for freeze protection) is a separate consumable.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS 202C Immersion Circulator Datasheet
Datasheet ·96,8 КБ ·2 pp.
View PDF