Peter Huber Kältemaschinenbau KISS 212B
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS 212B

Peter Huber Kältemaschinenbau KISS 212B — Thermostatic baths.

The KISS 212B is a 10.5 l stainless-steel bath immersion circulator with a pressure/suction pump (14 l/min at 0.25 bar) and 1.5–2.1 kW heating, covering 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. Its non-variable pump and IP20 enclosure distinguish it from larger CC-series siblings, suiting closed-loop temperature control of external jacketed vessels and bath-mounted sample work.

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection

Model ControllerInterfacePump connectionTemperature range, °CWeight, kgMax delivery flow rate, liter/minute
KISS 212B Current KISSUSB (Device), RS232 InteM16x1 AG male25–2001114
CC-202C Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200922
CC-205B Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200927
CC-208B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–2001027
CC-212B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001122
CC-215B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–20012—
CC-220B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001427
CC-225B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001627
KISS 202C KISSUSB (Device), RS232 InteM16x1 AG male45–200814
KISS 205B Badumwälzthermostat mit Edelstahlbad KISSUSB (Device), RS232 InteM16x1 AG male45–200914
KISS 208B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001014
KISS 215B KISSUSB (Device), RS232 InteM16x1 AG male25–2001214
KISS 220B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001414
KISS 225B KISSUSB (Device), RS232 InteM16x1 AG male25–2001614

Full specifications

Core Performance & Functional Specifications
Alarm output optic, acoustic
Dimensions (w × D × H) 350 x 375 x 375 mm
Maximum delivery flow rate 14 l/min (pressure pump; 0, 25 bar)
Pump connection M16x1 AG male (Option, #19607)
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 25...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 05 K
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump; 0, 17 bar)
Bath opening WxD/BD mfr 290 x 152 / 150 mm (WxD/depth)
Bath volume mfr 10, 5 l
Filling volume min. mfr 5, 5 l (minimum)
Dimensions, Weight & Mechanical Design
Weight 11 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Installation, control and operation

Integration & Operating Notes

Electrical
Branch circuit needs 10–16 A fusing and UL 489 protection

The unit draws a maximum of 10 A and requires a minimum 10 A and maximum 16 A fuse. An external branch circuit protection device according to UL 489 is required for UL-compliant installations, so size the protective device and wiring accordingly.

Datasheet, page 1 and page 2, installation notes
Mounting
Install in a dry location; IP20 prohibits water ingress

The enclosure is rated IP20, so it must be installed in a dry location protected from water ingress and access to live parts. Ambient temperature must stay between 5 °C and 40 °C for proper operation; conditions outside this range can affect performance, safety, and warranty.

Datasheet, page 1, installation notes
Process Limits
Cooling method sets the achievable low-temperature limit

Without cooling the range is 25–200 °C; with water cooling it extends to 20–200 °C; with a refrigerator it reaches -30–200 °C. Select and connect the cooling utility or chiller before attempting sub-ambient setpoints, and account for the minimum filling of 5.5 l in the 10.5 l bath.

Datasheet, page 1, operating temperature range
Cooling & Media
Avoid acetone and polyglycols with the plastic pump

The plastic pump is not resistant to acetone or polyglycols (depending on the manufacturer). For freeze protection, use water with glycerin or ethylene glycol; selecting an incompatible thermofluid can cause chemical attack and pump failure.

Datasheet, page 2, operation note
Control & I/O
Autostart after power failure may restart heating and pumping

The controller has an autostart function for power failure, which may restart heating and pumping automatically. Include this behavior in the risk assessment and process safety concept, especially where unattended restart could be hazardous.

Datasheet, page 1, operation note
Electrical
Specify country-specific plug at order

Single- and two-phase devices (100 V to 240 V) ship with a power cable and country-specific plug that must be specified when ordering. Three-phase devices under 63 A ship with cable but no plug, so confirm the plug type for the destination country to avoid ordering delays.

Datasheet, page 2, installation note

Selection & Integration Q&A

When should I choose the KISS 212B rather than a CC-series sibling such as the CC-212B?

Choose the KISS 212B when you need a 10.5 l stainless-steel bath with a non-variable pressure/suction pump and 1.5–2.1 kW heating for 25–200 °C operation. The CC-series models are separate configurations; compare their bath volumes and pump/control features before selecting a larger or differently controlled unit.

What power supply and branch protection must be provided for the KISS 212B?

Provide a 200–240 V 1~/2~ 50/60 Hz supply with a 10 A minimum and 16 A maximum fuse, and an external branch circuit protection device according to UL 489. Voltage tolerance is ±10% only if frequency tolerance does not run in the opposite direction.

Can the KISS 212B be installed in a washdown area or outdoors?

No. The enclosure is IP20 and must be installed in a dry location protected from water ingress and access to live parts. Ambient temperature must remain between 5 °C and 40 °C for proper operation.

What accessories or consumables are not included in the base delivery and must be ordered separately?

The country-specific plug must be specified at order for single- and two-phase devices. A cooling utility or refrigerator is required for sub-ambient operation, and a compatible thermofluid (water with glycerin or ethylene glycol) must be supplied; acetone and polyglycols are not compatible with the plastic pump.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS 212B Immersion Circulator
Datasheet ·88,7 КБ ·2 pp.