Peter Huber Kältemaschinenbau KISS 220B
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS 220B

Peter Huber Kältemaschinenbau KISS 220B — Thermostatic baths.

The KISS 220B is a 17-litre stainless-steel bath immersion circulator with a fixed-speed pressure/suction pump, rated for 25–200 °C (20–200 °C with water cooling, −30–200 °C with refrigerator). Its 200–240 V 1~/2~ supply and 10 A draw distinguish it within the KISS series. It serves as a temperature-control source for external circuits and immersed samples in laboratory processes.

Direct drop-in replacements

Eppendorf
MixMate ®
53%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection

Model ControllerInterfacePump connectionTemperature range, °CWeight, kgMax delivery flow rate, liter/minute
KISS 220B Current KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001414
CC-202C Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200922
CC-205B Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200927
CC-208B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–2001027
CC-212B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001122
CC-215B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–20012—
CC-220B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001427
CC-225B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001627
KISS 202C KISSUSB (Device), RS232 InteM16x1 AG male45–200814
KISS 205B Badumwälzthermostat mit Edelstahlbad KISSUSB (Device), RS232 InteM16x1 AG male45–200914
KISS 208B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001014
KISS 212B KISSUSB (Device), RS232 InteM16x1 AG male25–2001114
KISS 215B KISSUSB (Device), RS232 InteM16x1 AG male25–2001214
KISS 225B KISSUSB (Device), RS232 InteM16x1 AG male25–2001614

Full specifications

Core Performance & Functional Specifications
Alarm output optic, acoustic
Bath volume 17 l
Dimensions (w × D × H) 350 x 555 x 375 mm
Filling volume 10 l (min)
Maximum delivery flow rate 14 l/min (pressure pump; 0, 25 bar)
Pump connection M16x1 AG male (Option, #19607) (Option, #19607)
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 25...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 05 K
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump; 0, 17 bar)
Bath opening WxD/BD mfr 290 x 329 / 150 mm (WxD/depth)
Dimensions, Weight & Mechanical Design
Weight 14 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The KISS 220B is an immersion circulator with a 17-litre stainless-steel bath (minimum filling 10 l) and a bath opening of 290 × 329 mm, 150 mm deep, with 205 mm opening height. It uses the KISS controller with OLED display and provides USB (Device) and RS232 interfaces. Within the series, this model is the 220B variant; sibling models CC-202C, CC-205B, CC-208B and CC-212B are separate configurations and their specifications do not apply here. The unit weighs 14 kg and is rated IP20, safety class III/FL per DIN 12876.

Thermal behaviour and media

Temperature range depends on the cooling method: 25–200 °C without active cooling, 20–200 °C with water cooling, and −30–200 °C with a refrigerator. Temperature stability is ±0.05 K. Heating capacity is 1.5–2.1 kW. The pump is a pressure and suction type with fixed speed (not adjustable): pressure side up to 14 l/min at 0.25 bar, suction side up to 10.5 l/min at 0.17 bar. The plastic pump is not resistant to acetone or polyglycols; for freeze protection, water should be mixed with glycerine or ethylene glycol.

Installation and control

Supply is 200–240 V, 1~ or 2~, 50/60 Hz, with maximum current 10 A. Branch circuit protection must be 10–16 A and external protection per UL 489 is required. Voltage and frequency tolerances are ±10% and must not run in opposite directions. Ambient operating temperature is 5–40 °C. An autostart function restarts the unit after power failure. An optional Pt100 sensor connection (#10688, factory-fitted only) displays but does not control process temperature. An M16×1 AG male pump connection is available as option #19607.

Integration & Operating Notes

Electrical
Supply must be 200–240 V with 10–16 A branch protection

The unit draws up to 10 A and requires external branch circuit protection per UL 489 rated between 10 A and 16 A. Verify that the available supply is within 200–240 V and 50/60 Hz with ±10% tolerance, and that voltage and frequency deviations do not occur in opposite directions, to avoid operating outside the specified envelope.

KISS 220B Immersion Circulator Datasheet, page 1–2
Process Limits
Cooling method sets the usable temperature range

Without active cooling the bath operates from 25 °C to 200 °C; water cooling extends the lower limit to 20 °C; a refrigerator extends it to −30 °C. Select the cooling configuration before specifying the process temperature, because the achievable range is not the same across configurations.

KISS 220B Immersion Circulator Datasheet, page 1
Cooling & Media
Avoid acetone and polyglycols with the plastic pump

The pump's plastic wetted parts are not resistant to acetone or polyglycols (depending on manufacturer). For freeze protection, mix water with glycerine or ethylene glycol rather than using pure water. Confirm fluid compatibility before filling to prevent pump damage and contamination.

KISS 220B Immersion Circulator Datasheet, page 2
Control & I/O
Autostart restarts the unit after mains failure

An autostart function automatically restarts the unit after a power failure. For unattended operation, assess whether this behaviour is acceptable for the process, or provide external sequencing to prevent unexpected restart of heating and circulation.

KISS 220B Immersion Circulator Datasheet, page 1
Control & I/O
Pt100 option displays but does not control temperature

The optional Pt100 sensor connection (#10688) is factory-fitted only and is intended to display an external process temperature; it does not provide closed-loop control. Order it with the unit if external temperature monitoring is required, since it cannot be added later in the field.

KISS 220B Immersion Circulator Datasheet, page 1

Selection & Integration Q&A

When should I choose the KISS 220B instead of a sibling model such as the CC-205B or CC-212B?

Choose the KISS 220B when you need a 17-litre bath with a 25–200 °C base range and fixed-speed pressure/suction pump. Sibling CC models are separate configurations; compare their published bath volumes and temperature ranges before selecting, as their specifications do not apply to the 220B.

What external circuit and power provisions are required to integrate the KISS 220B?

Provide a 200–240 V, 1~ or 2~, 50/60 Hz supply with 10–16 A branch protection per UL 489. The pressure pump delivers up to 14 l/min at 0.25 bar and suction up to 10.5 l/min at 0.17 bar; size external tubing and fittings to stay within these limits. An M16×1 AG male pump connection is available as option #19607.

What maintenance and accessories should be planned for the KISS 220B that are not in the base delivery?

The Pt100 sensor connection (#10688) and M16×1 AG male pump connection (#19607) are options ordered separately; the Pt100 is factory-fitted only. No maintenance schedule is given in the supplied datasheet; consult the operating manual for service intervals. Use glycerine or ethylene glycol for freeze protection.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS 220B Immersion Circulator Datasheet
Datasheet ·102,8 КБ ·2 pp.
View PDF