Peter Huber Kältemaschinenbau KISS 225B
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS 225B

Peter Huber Kältemaschinenbau KISS 225B — Thermostatic baths.

The KISS 225B is a 23.5 l stainless-steel bath with an integrated KISS immersion circulator, rated 25–200 °C without cooling, 20–200 °C with water cooling and -30–200 °C with a refrigerator. Its non-variable pressure/suction pump delivers 14 l/min at 0.25 bar and 10.5 l/min at 0.17 bar, distinguishing it from smaller CC-series baths. It serves as a temperature-controlled bath for external circulation or direct immersion in laboratory processes.

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection

Model ControllerInterfacePump connectionTemperature range, °CWeight, kgMax delivery flow rate, liter/minute
KISS 225B Current KISSUSB (Device), RS232 InteM16x1 AG male25–2001614
CC-202C Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200922
CC-205B Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200927
CC-208B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–2001027
CC-212B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001122
CC-215B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–20012—
CC-220B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001427
CC-225B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001627
KISS 202C KISSUSB (Device), RS232 InteM16x1 AG male45–200814
KISS 205B Badumwälzthermostat mit Edelstahlbad KISSUSB (Device), RS232 InteM16x1 AG male45–200914
KISS 208B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001014
KISS 212B KISSUSB (Device), RS232 InteM16x1 AG male25–2001114
KISS 215B KISSUSB (Device), RS232 InteM16x1 AG male25–2001214
KISS 220B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001414

Full specifications

Core Performance & Functional Specifications
Filling volume 15 l
Maximum delivery flow rate 14 l/min (pressure pump; 0, 25 bar)
Pump connection M16x1 AG male (Option, #19607)
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 25...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 05 K
Alarm message mfr optic, acoustic
Dimensions (W x D x H) mfr 350 x 555 x 425 mm
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump; 0, 17 bar)
Bath opening WxD/BD mfr 290 x 329 / 200 mm
Bath volume mfr 23, 5 l
Dimensions, Weight & Mechanical Design
Weight 16 kg
Electrical, Power & Drive Specifications
Current consumption 10 A
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Installation, control and operation

Integration & Operating Notes

Electrical
Branch protection must be sized 10–16 A with UL 489 device

The unit draws up to 10 A at 200–240 V; the datasheet requires an external branch circuit protection device according to UL 489 and permits a fuse between 10 A and 16 A. Undersizing will trip on heat-up; oversizing leaves the supply cable unprotected.

Datasheet, Specifications (installation notes
Process Limits
Plastic pump incompatible with acetone and polyglycols

The standard plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). Using these media will degrade the pump; select an alternative fluid or consult the factory for a more resistant pump before filling the 23.5 l bath.

Datasheet, Specifications (operation notes
Cooling & Media
Freeze protection requires glycol mix, not pure water

When operating near or below ambient, water should be mixed with glysantine or ethylene glycol for freeze protection. Pure water can freeze and damage the bath or pump; the bath bridge also provides a hole for an immersion cooler TC45–TC100E if active cooling is needed.

Datasheet, Specifications (operation notes
Control & I/O
Autostart restarts after power failure — assess process risk

The autostart function automatically restarts the unit after a power failure. For unattended processes this may be desired, but where a restart could be hazardous or invalidate a test, the function must be disabled or the process interlocked externally.

Datasheet, Specifications (operation notes
Electrical
Country-specific plug must be specified at order

Single- and two-phase devices are delivered with a power cable and country-specific plug; the plug type must be stated when ordering. Missing this detail delays commissioning because the cable is not field-configurable to all plug standards.

Datasheet, Specifications (installation notes

Selection & Integration Q&A

When should I choose the KISS 225B rather than a smaller CC-series bath?

Choose the KISS 225B when the process needs a 23.5 l bath (15 l minimum fill) and up to 200 °C with 1.5–2.1 kW heating. Smaller CC models should be selected only when their own verified volume and temperature data match the application; the KISS 225B is the larger-bath option in this series.

What external cooling and fluid provisions are needed to reach sub-ambient temperatures?

Without cooling the range is 25–200 °C; with water cooling 20–200 °C; with a refrigerator -30–200 °C. The bath bridge has a hole for an immersion cooler TC45–TC100E (holder accessory). For freeze protection, mix water with glysantine or ethylene glycol, not pure water.

What are the electrical supply and branch protection requirements for installation?

Supply is 200–240 V 1~/2~ 50/60 Hz, max. 10 A. Provide an external branch circuit protection device per UL 489 with a fuse between 10 A and 16 A. Voltage and frequency may deviate ±10% only if not in opposite directions. Specify the country-specific plug at order.

What maintenance and accessory items are not included in the base delivery?

The base unit includes the bath and immersion circulator but not the country-specific plug unless specified, nor the immersion cooler TC45–TC100E or its holder. The plastic pump is not resistant to acetone or polyglycols; alternative pump material must be requested. No maintenance schedule is given in the supplied datasheet.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS 225B Immersion Circulator
Datasheet ·115,9 КБ ·2 pp.
View PDF