Peter Huber Kältemaschinenbau TC45E
Peter Huber Kältemaschinenbau TC45E — Immersion coolers.
The TC45E is an air-cooled immersion cooler with a natural R-290 refrigerant circuit, rated for -45 to 100 °C and controlled by an MPC unit with 0.1 K display resolution. It delivers 0.24 kW at 0 °C but only 0.05 kW at -40 °C, so it suits low-load bath cooling and cold-trap duty rather than high-heat processes. No pump is fitted; circulation must come from the bath or an external loop.
Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Power supply | Refrigerant | Safety class | Temperature range, °C |
|---|---|---|---|---|---|---|
| TC45E Current | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| CT50 Single OLÉ | 59 | OLÉ | 220-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | III / FL | -50–50 |
| TC100 | 55 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100E-F | 55 | MPC | 110-120V 1~ 60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100-F | 58 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC45 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E-F | 54 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45-F | 59 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC50 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E-F | 60 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50-F | 60 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
Full specifications
| Pump speed adjustable | no |
| Pump type | no pump |
| External sensor connection | Pt100 |
| Temperature range | -45...100 °C |
| Temperature stability | 0.5 K |
| Alarm message mfr | optic, acoustic |
| Dimensions (W x D x H) mfr | 190 x 295 x 360 mm |
| Refrigeration machine mfr | air-cooled, natural refrigerant |
| Weight | 16 kg |
| Current consumption | 1.4 A (max.) |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Controller | MPC |
| Display resolution | 0.1 K |
| Temperature display | LED Display |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Refrigerant quantity | 0.041 kg (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 0.24 kW (0 °C) -20 °C: 0.18 kW · -30 °C: 0.1 kW · -40 °C: 0.05 kW |
| Safety class | I / NFL |
Description
Construction and series positioning
- The TC45E is a standalone immersion cooler: a refrigerated probe is placed directly into the bath or cold trap, and the refrigeration machine is air-cooled with R-290 (A3, H220) as the first-stage refrigerant (0.041 kg charge).
- Unlike the larger TC100/TC100E/TC100E-F siblings (55 dB(A)) and the CT50 Single OLÉ (59 dB(A)), the TC45E is the compact 16 kg unit at 190 x 295 x 360 mm; it has no pump and no adjustable pump speed, so it does not circulate the bath itself.
- Temperature is controlled by an MPC controller acting on a solenoid valve in the cooling circuit while the refrigeration unit runs continuously; the setpoint is shown on an LED display with 0.1 K resolution, and an external Pt100 sensor (included, connected by cable and plug) enables control at the point of use.
Behavior under load
- Cooling capacity is strongly temperature-dependent: 0.24 kW at 0 °C, 0.18 kW at -20 °C, 0.10 kW at -30 °C and 0.05 kW at -40 °C. Capacity is rated at 20 °C room temperature and falls as ambient rises, so the load must be sized at the actual operating point with margin for the site ambient.
- The working range is -45 to 100 °C with 0.5 K temperature stability; the continuous compressor plus solenoid-valve control means the unit modulates cooling rather than cycling the compressor.
Installation and operation
- Supply is 208-240 V, 1~ or 2~, 50/60 Hz, with maximum current consumption 1.4 A; the datasheet requires a minimum 10 A fuse and notes that voltage and frequency tolerances are coupled (±5 % voltage with ±2 % frequency, opposite-sign deviations not allowed).
- Ambient must remain between 5 and 40 °C; the unit is EMC class A, group 1 (EN 55011), intended for industrial environments. Single- and two-phase units ship with a power cable and country-specific plug that must be specified at ordering; space requirements around the air-cooled unit must be respected for heat rejection and service access.
Integration & Operating Notes
The unit draws up to 1.4 A on a 208-240 V, 1~ or 2~, 50/60 Hz supply and the datasheet calls for a minimum 10 A fuse. Voltage and frequency tolerances are coupled: ±5 % voltage with ±2 % frequency is acceptable, but opposite-sign deviations are explicitly not allowed, so supply quality must be checked before commissioning.
The air-cooled refrigeration circuit rejects heat to the room, and capacity is rated at 20 °C room temperature and drops as ambient rises. Installing the unit in a cold room below 5 °C or a hot enclosure above 40 °C is outside the permitted ambient band, and space requirements around the unit must be respected for airflow and service access.
Cooling capacity falls from 0.24 kW at 0 °C to 0.05 kW at -40 °C, so a load acceptable at near-zero temperatures may be unreachable near -45 °C. The immersion probe must be matched to the bath volume and heat ingress, and the 0.5 K stability figure applies only when the load stays within the available capacity at that operating point.
The MPC controller reads an external Pt100 sensor connected by cable and plug, and a Pt100 sensor is supplied with the unit, so external temperature control is possible without extra accessories. The refrigeration unit runs continuously and the controller actuates a solenoid valve in the cooling circuit, which defines the control behavior and the wear pattern of the valve.
Single- and two-phase devices are delivered with a power cable and a country-specific plug that must be stated when ordering; three-phase devices below 63 A come with cable but without plug, so the customer must provide and fit one. Omitting the plug specification delays commissioning because the unit cannot be connected on arrival.
The first stage uses 0.041 kg of R-290 (A3, H220), a flammable natural refrigerant. Installation must account for room ventilation and ignition-source requirements applicable to A3 refrigerants; the unit is not certified for use in explosive atmospheres and must not be placed where refrigerant can accumulate in a confined space.
Selection & Integration Q&A
Pick the TC45E for compact, low-load bath or cold-trap duty down to -45 °C where no pump is needed. The TC100/TC100E/TC100E-F and CT50 Single OLÉ are larger siblings; choose them when higher cooling capacity or a different noise/configuration profile is required, since the TC45E is the 16 kg, 190 x 295 x 360 mm unit with 0.24 kW at 0 °C.
208-240 V, 1~ or 2~, 50/60 Hz, max 1.4 A, with a minimum 10 A fuse. Voltage and frequency tolerances are coupled (±5 % / ±2 %; opposite-sign deviations not allowed). Single- and two-phase units ship with cable and a country-specific plug that must be specified at ordering.
Ambient must stay between 5 and 40 °C; capacity is rated at 20 °C room temperature and falls as ambient rises, so avoid cold rooms and hot enclosures. The unit is EMC class A, group 1 (EN 55011), intended for industrial environments, and space requirements around the air-cooled unit must be respected for heat rejection and service access.
A Pt100 external sensor with cable and plug is included, enabling external temperature control out of the box. The country-specific power plug must be specified at ordering for single- and two-phase units; three-phase units below 63 A come with cable but no plug, which the customer must provide and fit. No pump is fitted, so bath circulation must come from elsewhere.